Validation of a leg-to-leg bioimpedance analysis system in assessing body composition in postmenopausal women

  • Paper
  • Published:
International Journal of Obesity Submit manuscript

Abstract

OBJECTIVES: To evaluate the validity of a leg-to-leg bioimpedance analysis (BIA) system in predicting body composition as measured by dual-energy X-ray absorptiometry (DXA) in postmenopausal women.

SUBJECTS AND METHODS: Body fat mass (FM), %Fat and fat free mass (FFM) were measured in 124 postmenopausal women (age: 51–63 y, body mass index (BMI): 17–38 kg/m2) first by the leg-to-leg BIA system, and then by DXA as reference method. Bland-Altman analysis was used to determine the bias and 95% limits of agreement between the two methods for the assessment of the individual. Precision error (CV%) of the BIA system was obtained by repeated measurements with intermediate repositioning.

RESULTS: The leg-to-leg BIA system had a high reproducibility with within-day CVs being 0.6% for FFM and 1.1% for FM, and between-day CVs about twice that. The impedance index (Ht2/Z) obtained by the leg-to-leg BIA was moderately correlated to FFM measured by DXA (r=0.66). A significant, systematic bias was observed between the two methods. The BIA system overestimated FM by a mean of 3.1 kg, and underestimated FFM by 2.7 kg. The analysis of 95% limits of agreement showed that for most individuals, %Fat estimated by the BIA might differ from that measured by DXA by 12% below to 45% above, indicating the lack of agreement between the two methods for the assessment of the individual.

CONCLUSIONS: The leg-to-leg BIA system can provide simple, rapid and highly reproducible measurements of body composition for groups, but it has limited accuracy for the assessment of the individual. Population-specific equations will be needed to improve its accuracy in estimating body composition in postmenopausal women.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to X Xie.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Xie, X., Kolthoff, N., Bärenholt, O. et al. Validation of a leg-to-leg bioimpedance analysis system in assessing body composition in postmenopausal women. Int J Obes 23, 1079–1084 (1999). https://doi.org/10.1038/sj.ijo.0801034

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1038/sj.ijo.0801034

  • Springer Nature Limited

Keywords

This article is cited by

Navigation