Abstract
Quantitative mechanical properties of biological tissues can be mapped using the shear wave elastography technique. This technology has demonstrated a great potential in various organs but shows a limit due to wave attenuation in biological tissues. An option to overcome the inherent loss in shear wave magnitude along the propagation pathway may be to stimulate tissues closer to regions of interest using alternative motion generation techniques. The present study investigated the feasibility of generating shear waves by applying a Lorentz force directly to tissue mimicking samples for magnetic resonance elastography applications. This was done by combining an electrical current with the strong magnetic field of a clinical MRI scanner. The Local Frequency Estimation method was used to assess the real value of the shear modulus of tested phantoms from Lorentz force induced motion. Finite elements modeling of reported experiments showed a consistent behavior but featured wavelengths larger than measured ones. Results suggest the feasibility of a magnetic resonance elastography technique based on the Lorentz force to produce an shear wave source.
Similar content being viewed by others
Introduction
Stiffness difference between healthy and abnormal biological tissues has long been put to use in the form of manual palpation to diagnose potential diseases. Although helpful to roughly estimate tissue stiffness, more precise and quantitative stiffness assessments can be reached using modern techniques such as shear wave elastography imaging1,2,3,4. This method relies on the propagation of shear waves of which the speed is proportional to the square root of the shear modulus of the medium5 in linearly elastic, homogenous, isotropic and infinite solids. Shear wave elastography requires to (1) induce shear waves within the sample, (2) track local displacements propagating as a shear wave, and (3) compute mechanical parameters from displacement images. Although some imaging techniques have been proposed to track shear waves, such as optical coherence tomography6 or a high speed camera7, main imaging methods for elastography remain ultrasound8 and magnetic resonance imaging (MRI)9. In MRI, computation of viscoelastic parameters has been performed with different algorithms, such as local frequency estimation (LFE)10, algebraic inversion of the equation of wave propagation11,12,13, identification methods using finite elements14, and temporal cross-correlation15,16.
In terms of shear wave generation, external actuators placed on the surface of the imaged tissue are mostly used in magnetic resonance elastography (MRE). In such cases, elastic waves propagate, and attenuate, from the surface to the region of interest while being tracked by the scanner. Pneumatic actuators have been widely used in MRE and have shown versatility through applications to liver17, brain18, heart19, and lung20. In this system, air pulses are generated by an active driver outside the MR room and routed through a plastic tube to a passive transducer attached to the patient. The setup is simple and handles well mono-frequency waveforms in the range of ~ 40 Hz to ~ 110 Hz. However, the uncontrolled acoustic response of pulsed air along the connection tube may render multi-frequency MRE difficult. Alternatively, a rigid rod has been attached directly to the pulsing active driver on one end and to a pad tied to the patient on the other end, which allowed for multi-frequency MRE21,22,23. Other successful surface actuation apparatus have been reported, such as piezoelectric24 and Lorentz coil drivers25 where the magnetic field of the MR scanner and the current injected in the coil combine to generate vibrations. So far, the presented solutions to produce displacement fields have in common the decrease of motion deflection as the driving frequency increases. This feature along with the inherent damping behavior of viscoelastic tissues limit in-vivo applications to low frequencies. To circumvent this limitation, mechanical devices relying on the centrifugal force have been presented, such as the “gravitational transducer”26. This system is made of a mass rotating about an axis of which the spinning frequency is driven by a motor outside the MR room. The centrifugal force exhibited in this type of actuator allows to increase the vibration amplitude by increasing the rotation frequency. This is an elegant solution to elicit displacement fields of sufficient amplitude at high frequencies within a reasonable distance from the surface actuator. However, challenges may remain at deep locations due to attenuation.
Instead of attempting to produce motion fields likely to reach the region of interest from the surface, an alternative may be to vibrate tissues closer to the targeted area thus shortening the propagation pathway and reducing the related wave damping. Ultrasound radiation pressure sources have been presented in that regard27 but carry practical implementation difficulties in the context of MR imaging. Intrinsic MRE, consisting in capitalizing on the natural pulsation of arteries as a motion source, requires no external actuator and has shown great potential at heartbeat frequencies16,28,29,30. Alternatively, the Lorentz force has been proposed for direct application of a motion (as opposed to Lorentz coils) to tissue mimicking phantoms by Grasland-Mongrain et al.31,32. They showed the possibility of producing shear waves within a conductive medium using the Lorentz force produced by a permanent portable magnet and external electrodes or a magnetic stimulation device, and an ultrasonic scanner for detection31,32,33.
Here, we hypothesized that a similar configuration could be used as a remote mechanism for shear wave induction in MRE by taking advantage of the strong magnetic field of MR scanners. Mechanical displacements of electromagnetic origin have been observed in soft solids using MRI34 but to our knowledge, no study has reported the use of the Lorentz force to produce in situ displacements in the context of MR elastography. In the present work, we propose to evaluate the feasibility of generating and detecting shear waves from a Lorentz force source within tissue mimicking samples. A specific MR compatible experimental setup was designed to isolate the Lorentz force contribution of the measured displacement field. Finally, preliminary conclusions on stiffness reconstructions from Lorentz force induced motion using the well-established LFE method are reported.
Method
The proposed method consists in producing a Lorentz force density in tissue mimicking gels (phantoms) which in turn induces shear waves that are recorded using an MR clinical system and processed to retrieve elasticity information. The expression of the Lorentz force density [N.m-3] is given by
where \({\varvec{j}}\) is the current density [A m−2] and \({\varvec{B}}\) the magnetic field [T]35. Here, the strong homogeneous magnetic field of the MR scanner and externally applied currents were used to produce the Lorentz force.
Experimental setup
In all the experiments reported in the present manuscript, agar phantoms were poured into a 3D printed mould and crossed by an electrical current \({\varvec{j}}\) and a magnetic field \({\varvec{B}}\) (Fig. 1a). Details about the different phantoms used are given in the paragraphs below and in Table 1. Saline baths (20%w) were designed on the two sides of the phantom holder to make the electrical contact between the phantom and external electrodes. This permitted to avoid mechanical contacts between electrodes and the phantom, thus allowing to isolate the Lorentz force density as the source of motion (i.e., avoid artefact movements). The electrical current was brought by means of electrodes (copper wires) plunged into the saline. Wires were placed parallel to the static magnetic field direction from the end of the bore to the phantom holder, and maintained fixed from above the phantom holder to the saline baths. Special care was taken to avoid mechanical contacts between wires and the rest of the setup to eliminate other movement artefacts. The phantom was placed in a 32-channel receive-only head coil in the bore of a 3 T MRI scanner (Achieva TX, Philips Healthcare, Best, Netherlands). The coil was positioned at the bore isocentre to minimise eddy currents and to facilitate the reproducibility of experiments. The phantom was oriented along the magnetic field (z axis, red arrow in Fig. 1a) and the current was mainly flowing along the y axis (green arrow in Fig. 1a) between the two electrodes so that the magnetic field and the current were perpendicular. The induced Lorentz force was oscillating along the x axis (black arrow in Fig. 1a) at the frequency of the current. A motion field was then produced with in-plane components parallel and perpendicular to the Lorentz force, i.e. along x and z axes. The electrical current was provided by a waveform generator (Agilent, 33250A 80 MHz), amplified (Brüel & Kjær, power amplifier type 2706) and routed through a 2-Ω shunt resistance in series with the phantom and the generator. Two sinusoidal waveforms of 60 Hz and 90 Hz were successively used in the experiments.
MRI acquisitions
For actuation and motion detection, current delivery to the phantom was synchronised with the motion encoding gradients (MEGs) of a 2D gradient echo sequence tailored to elastography experiments36 (Fig. 1b). The current generator was triggered by the MR scanner at the beginning of the MRE pulse sequence, in the same way as with other standard mechanical transducers. The phase accumulation experienced by spins undergoing harmonic actuation is given by:
where γ [rad s−1 T−1] is the sample’s gyromagnetic ratio, τ [s] the period of actuation, and G [T m−1] the motion encoding gradient amplitude. Generated motion was encoded along three orthogonal directions (anterior–posterior, right-left, foot-head). Parameters of the elastography sequence were: repetition time TR = 50 and 56 ms (adjusted to match a multiple of the shear wave vibration period), echo time TE = 15 and 20 ms, spatial resolution = 1.25 mm × 1.25 mm, slice thickness = 5 mm, elastography phases = 4, amplitude of the motion encoding gradients = 31 mT/m (highest value available to ensure maximum detection in the furthest regions from the motion source), and motion encoding over one cycle.
For displacement field measurements, two phase images were successively acquired with MEGs of opposite polarities. The second phase image was then subtracted from the first one using the exponential form of the complex MR signal to eliminate potential static phase offsets, thus producing a single phase difference image free of unwanted constant background. During the acquisition (whole TR) of both phase images, the generator was continuously delivering the same oscillating electrical current. The induced phase offset due to current injection in the phantom was then the same in these images and was consequently cancelled out in the phase difference operation. Additionally, receive coils were syntonised to the Larmor frequency of hydrogen’s proton and consequently show a low sensitivity to the low-frequency currents used in our study. To compute displacement maps from phase difference images, the encoding coefficient |ξ| in [rad m−1] was computed using \(\left|\xi \right|=\left|\frac{\gamma {G}_{0}\left(\mathit{sin}\left(\pi f\tau \right)-2sin\left(\frac{1}{2}\pi f\tau \right)\right)}{\pi f}\right|\), which approximates first moment nulling gradients and where γ [rad s−1 T−1] is the sample’s gyromagnetic ratio, τ [s] and G0 [T m−1] are the period and the amplitude of the MEG respectively, and f [Hz] is the excitation frequency37.
Inverse problem and data processing
Once shear waves were produced and detected, the next step consisted in generating maps of viscoelastic parameters. The forward problem relating harmonic displacements to tissue mechanical properties is described by the Navier’s equation:
where ρ is the material’s density [kg m−3], ω the motion angular velocity [rad s−1], u the motion field [m], λ [N m−2] the first Lamé coefficient38 and G [N m−2] the complex shear modulus. In agar and agar-doped phantoms, the real part of the complex shear modulus (storage modulus) was shown to considerably dominate the imaginary part (loss modulus)39,40. The LFE, which ignores the loss modulus, was then chosen to retrieve first elasticity information from Lorentz force induced displacements in agar phantoms, given its availability and wide use in the MRE community. The LFE reconstructions could allow validating the working hypothesis to show the possibility of implementing the Lorentz force MRE method on a clinical scanner. Additionally, the LFE is known to be more robust against noise than other direct methods involving the estimation of second or higher order derivatives11,13.
The LFE consists in applying pairs of filters to the selected harmonic of the displacement data in the k-space. These filters are called log-normal quadrature filters and are generally centered on frequencies separated by one octave41. The ratio of displacements processed by each filter of one pair is equal to the local wave number. As the actuation frequency ω is known (i.e., electrical current frequency), the wave number k can then be related to the local real value of the shear modulus in [N.m-2] using \(\mu =\rho {\left(\frac{\omega }{k}\right)}^{2}\). It is equivalent to solving the Helmholtz equation \(-\rho {\omega }^{2}{\varvec{u}}=\boldsymbol{ }\mu\Delta {\varvec{u}}\), which is a simplified version of the Navier’s equation under the assumption of tissue incompressibility, local homogeneity, and no shear wave attenuation. Shear modulus maps computed from displacements along the actuation direction (x axis) are reported in this study.
Prior to inversion, phase images were automatically unwrapped using a software embedded in the scanner (Resoundant Inc., Rochester, MN). Shear modulus images were then filtered to remove outliers using the smoothn Matlab function42,43. For displacement display (Fig. 3), a separate phase unwrapping procedure was used44,45,46.
Experiment 1
To demonstrate the proposed method, two experimental conditions were designed, both consisting in generating shear waves in tissue mimicking gels. The first one intended to show the feasibility of generating shear waves with the Lorentz force under favourable conditions (i.e., an electrically conductive, soft and homogeneous phantom, termed phantom A in Table 1), and to observe them with the MRI scanner. Two sinusoidal electrical currents of 137.5 mA peak-to-peak with frequencies of 60 Hz and 90 Hz were respectively used. Such intensity is 69 times as large as values reported in electrical stimulation (ES) protocols performed on humans (iES = 2 mA peak-to-peak)47. Results are shown Fig. 3.
Experiment 2
The purpose of the second experiment was to show that Lorentz force induced shear waves could be used to distinguish regions with different mechanical properties within the same heterogeneous structure using phantoms with different mechanical properties (termed B, C, D, and E in Table 1), and a less conductive medium corresponding to physiological conditions. To create the heterogeneous structure, phantom B was poured on top of phantom C in one phantom holder (Figs. 1a and 5A), and phantom D was poured on top of phantom E in another phantom holder (Figs. 1a and 5A). Sinusoidal currents of 62.22 mA peak-to-peak at frequencies of 60 Hz and 90 Hz were used, which corresponds to \(33\times {i}_{ES}\).
Simulations
A finite element physical model corresponding to the first experiment using phantom A was designed with Comsol 5.5 (COMSOL LiveLink with Matlab, Inc., Palo Alto, CA) to assess the displacement patterns obtained from a convolution of point-like sources, and to compare with experimental data. Simulations consisted in two steps, first computing the Lorentz force density distribution within the phantom, and second computing the wave field resulting from this Lorentz force density source term. To compute the Lorentz force density distribution, the following boundary value problem was solved using the electric current module:
where u is the electric potential [V], σ the electrical conductivity [S m−1], I the electrical current [A], S the surface areas [m2] of the electrical contact, and n is a unit vector perpendicular to the outer boundary. The externally applied electrical current I was modeled perpendicular to the phantom at the electrical contact location (Fig. 1a). The intensity and frequency of the current were the same as in the first experiment (60 Hz and 90 Hz, peak-to-peak amplitude of 137.5 mA). The phantom conductivity was homogeneous and set to 1 S/m as it impacts neither the current density distribution within the phantom nor its intensity, which are governed by the boundary conditions in this model. From the electric potential distribution u, the current density was computed using Ohm’s law: \({\varvec{J}}=-\sigma \nabla u\).. The Lorentz force density distribution in the phantom (Fig. 2) was then computed using Eq. (1):
where \(\left|{\varvec{B}}\right|=\left|{{\varvec{B}}}_{{\varvec{z}}}\right|=3 T\) is the magnetic field produced by the MR scanner.
To compute the displacement field resulting from the Lorentz force density distribution in the phantom, the following boundary value problem was solved using the solid mechanics module:
where u is the displacement field [m], ρ is the material’s density [kg m-3], ω is the pulsation [rad s−1], G is the complex shear modulus [N m−2], K is the bulk modulus [N m−2], and f is the Lorentz force density [N m−3]. The bulk modulus was set to 2.2 GPa (water). Although the approximation of no shear wave attenuation remains reasonable for agar phantoms in the calculation of the solution to the inverse problem, such approximation has large consequences on the calculated solution to the forward problem. The definition of a loss modulus was thus found necessary to account for the slight damping behavior of the phantom.
To fix the loss modulus G", the reported trend of the relationship between storage and loss moduli in agar phantoms was used. As aforementioned, the storage modulus was shown to dominate the loss modulus in agar and agar-doped phantoms. More specifically, around an order of magnitude difference can be observed between these two parameters39,40. The storage moduli at both frequencies were then fixed from the shear modulus distributions in Fig. 4, and the loss moduli were fixed from the above-mentioned G’–G” relationship. However, given the uncertainty on the phantom’s exact mechanical properties (variability in maps of Fig. 4 for G’ and approximated G”), the forward problem was solved multiple times using different sets of complex shear moduli. In the first simulation, G’ values at both frequencies were set to the mean values of the shear modulus distributions in Fig. 4: G’60 Hz = 1546 Pa and G’90 Hz = 1768 Pa. Loss moduli G" were set to 1/10th of the storage moduli: G”60 Hz = 155 Pa and G”90 Hz = 177 Pa. Results are shown in Fig. 3. In the second and third simulations, the same G’ values were used and G" values were varied and successively set to 1/5th and 1/20th of G’ values. In the fourth and fifth simulations, G’ values were changed towards the lower values in the shear modulus distributions in Fig. 4, and G" were set to 1/10th of G’ in each case. Results are shown in Fig. 3. All combinations of G’ and G" parameters are gathered in Table 2. This iterative framework intended to show that, despite uncertainties on the phantom’s exact mechanical characteristics, different possible storage and loss moduli do impact the shear wave propagation patterns and their wavelengths but do not question the Lorentz force density origin of the produced motion fields, which is what the simulation was meant to highlight.
Regarding the geometry, side boundaries (phantom holder’s walls) were fixed while the upper boundary was left free to mimic the configuration of Fig. 1. Both calculations (electric potential and displacement field) were performed in the frequency domain. Complex displacements in the imaged plane were then extracted and Fourier transformed into the time domain, aligned with the experimental displacement data, and Fourier transformed back into the frequency domain for display of the real and imaginary parts in Fig. 3. Mesh adaptations were performed to allow refinements where the gradients of solutions were high. No mesh dependence was observed in final results.
Results and discussion
Experiment 1 and simulations
Figure 2 shows the distribution of the Lorentz force density along the x, y, and z axes in the imaging plane (Fig. 2). A significant contribution along the x axis was observed while contributions along the y and z axes were found negligible. This behavior is consistent given the expression of the Lorentz force density (vector cross-product) and main directions of current density (along y axis) and magnetic field (along z axis).
Figure 3 displays the experimentally measured displacement fields along the three directions of space and the simulated ones computed from the Lorentz force density distribution. The storage moduli used to characterize the phantom in these simulations were obtained from the mean values of the shear modulus distributions displayed in Fig. 4. The loss moduli were approximated to 1/10th of the storage moduli. Figure 3 shows the line profiles of the polarized displacement fields along the propagation directions for each tested combination of G’ and G" (Table 2).
In Fig. 3, the experimental displacement patterns along x and z allow for clear identification of the shear wave source, i.e. where the Lorentz force density is the highest (Fig. 2), namely between electrical contacts within the saline bath-phantom structure (Fig. 1a). No other shear wave origin could be observed, indicating the absence of movement artefacts. Experimental and simulated displacement fields show similar patterns. They both present a monopole-like arrangement along the x axis and a dipole mode along the z axis, at both frequencies. These patterns correspond to the shear mode described by the Green’s function solution to the inhomogeneous Navier’s equation48. However, no clear pattern could be identified along the y axis in experimental data where the motion amplitude was very low resulting in measurement errors (broad areas with negative-only or positive-only displacement values).
Simulated displacements along the y axis show maximum amplitudes on the order of 10 nm, which cannot be measured with the MR scanner. Although presenting the same modes and propagation directions as experimental data, simulated displacement fields feature wavelengths larger than measured ones, which suggests a mismatch in the definition of the material properties in the simulations rather than an underlying modeling issue. The storage moduli were assigned based on experimental estimations, which implies that the storage shear modulus distribution from the LFE inversion was overestimated. The line profiles along propagation directions for possible complex shear modulus values present the same behavior around the center of the line, which matches the distribution of the Lorentz force density, regardless of the mechanical parameters used. The side lobes towards opposite edges have, however, their own individual behavior. All simulated configurations provided maximum displacement amplitudes larger than the measured ones. This suggests that the experimental displacement amplitudes may have been underestimated, which would be consistent with their low values close to the reported detection limit of a few hundreds of nanometers in MRE9.
Potential global under evaluation of displacement amplitudes have theoretically no impact on the LFE reconstruction, which does not depend on the displacement amplitude but on the wavelength estimation so long as enough signal-to-noise ratio is available. Qualitatively, the noticeable change in wavelength between measurements at 60 Hz and 90 Hz further confirms the wave field dependence on the frequency of the Lorentz force density. The impact of current intensity on the displacement field was also observed in other experiments where the current intensity was varied. With a current at 60 Hz and 113 mA peak-to-peak, the mean norm of the displacement field calculated over the whole field of view was 0.14 µm. This value accordingly dropped to 0.07 µm when the current was lowered by a factor of 2. The same conclusion was drawn at 90 Hz with the displacement mean norm decreasing from 0.08 µm to 0.06 µm (ratio of 1.2) when the current was lowered from 149 mA peak-to-peak to 113 mA peak-to-peak (ratio of 1.3).
Figure 4 shows the shear storage modulus distributions obtained with the LFE inversion performed on x displacements. The LFE relies on the evaluation of the wave number in motion field images making it particularly sensitive to reflections off boundaries and wave interferences, which result in apparent larger or smaller wavelengths and consequently higher or lower shear moduli even in homogeneous tissues. This is assumed to be the case at 60 Hz and 90 Hz where regions with higher and lower values respectively formed semi-circle branches on the left- and right-hand sides. These arcs of circles are consistent with the wave front geometries of the experimental x displacements shown in Fig. 3. At the center of both frequency maps, higher shear moduli correspond to regions where the assumption of a body force free solid underlying the derivation of the elastic Helmholtz equation is no longer valid due to the presence of the Lorentz force density (Fig. 2). The presence of the Lorentz force density translates into overestimated shear moduli to compensate for the inertia term. The mean values over the entire field of views are μ60 Hz = 1546 ± 383 Pa and μ90 Hz \(=1768\pm 440\) Pa.
Experiment 2
Figure 5 displays results of the second experimental condition consisting in generating shear waves using the Lorentz force in phantoms with different mechanical properties. The shear modulus maps were generated from x displacements, as in Fig. 4, for the evaluation of the impact of the Lorentz force density on reconstruction (the Lorentz force density pulses along the x axis). Phantom names corresponding to upper and lower parts of the phantom compartments are indicated in red in upper and lower right corners of shear modulus maps. Despite the higher electrical resistance due to lower salt content in those phantoms, currents with sufficient intensities 62 mA peak-to-peak could be applied to conduct these measurements with sufficient signal-to-noise ratios.
The separation between the two phantom layers is clearly visible on all shear modulus maps. However, the distribution within each layer shows a significant variance, especially in the phantom E. No major difference was measured between phantoms C and D at both frequencies despite the difference in agar concentration. The dotted line shows the separation between the two compartments where the shear modulus mean values and standard deviations were calculated. The dashed circle in the middle represents the extent to which the Lorentz force varied from its maximum magnitude (centre of the circle) to 1/10th of this maximum (edge of the circle). No impact of the Lorentz force could be observed in this circle as opposed to the centre of shear modulus maps in Fig. 4. This may be explained by the spatial coincidence of the Lorentz force distribution and the stiffness discontinuity. The LFE was reported to lack sensitivity at the boundary between media of different stiffness49 because of the spatial extent of filters and the assumption of local homogeneity. This leads to estimated stiffness values to be inaccurate within a certain distance to media boundary.
Another impact of the Lorentz force on the resolution of the inverse problem may appear in the coupling of the shear (divergence free) and pressure (curl free) fields near the source, as opposed to the far field where shear and pressure waves travel at different speeds50. The pressure field was also shown in Ref.13 to introduce a DC component into the displacement field produced by external surface actuators (i.e., no body force and no coupling of shear and pressure fields). Applying the curl operator to the Navier’s equation allows to isolate the contribution of the shear field at the cost of noise enhancement due to third-order derivatives13. As aforementioned, the presence of a body force in the field of view of the imaged tissue theoretically leads to overestimated mechanical parameters where the source is. Additional mapping of the motion source distribution may then be necessary to correct for the overestimation. In other reported cases where body forces (other than Lorentz forces) are present, such as supersonic shear imaging51, conventional methods to discard the body force term in the Navier’s equation consist in selecting displacement data after the shear wave source was switched off or selecting ROIs far from the source. Chatelin et al.52 have developed a method to solve the inverse problem while taking body forces into account in the context of high intensity focused ultrasound. It allows to estimate optimal stiffness values through a minimisation process between experimental displacements and simulated displacements using Green’s formalism. Such technique may provide the potential of including body forces in the calculation of the solution to the inverse problem in Lorentz force MR elastography.
Direct inversion techniques based on the Helmholtz equation conceptually allow for the addition of a source term in the formulation of the forward problem, but still rely on a local homogeneity assumption and high order derivatives. Identification techniques14,53 have shown good performance on brain MRE and involve the complete formulation of the viscoelastic forward problem, thus allowing for the addition of a source term. The gradient of mechanical parameters across the imaged domain is also taken into account. Additionally, such methods have been successful in configurations where the motion source, as in our case, is present in the field of view; that is in intrinsic MRE where the displacement field is produced by arteries pulsing within the imaged domain28,30.
In Lorentz force MRE, mapping the Lorentz force density distribution can be achieved using magnetic resonance electrical impedance tomography (MR-EIT), which allows to measure electrical current densities and conductivities of electrically inhomogeneous media54. Current in vivo techniques of MR-EIT resort to sets of electrodes attached to the tissue of interest55. Then, electric pulses are delivered to the tissue while being imaged, which presents many similarities with Lorentz force MRE. Additionally, MR-EIT is, like MRE, a phase contrast MR technique, which in turn may be combined using the Lorentz force as a tool for dual electrical–mechanical property mapping.
Limitations
The present study investigated the feasibility of generating shear waves using a Lorentz force produced in the imaged phantom from the combination of the MR magnetic field and electrical stimulation. Despite successful generation and detection of motion fields, limitations are severalfold. In terms of MR sequence, phase difference images were output by the scanner but no access to raw data was possible, which limited detailed investigation of phase image pre-processing steps. Comparison of displacement fields measured with different fully accessible MR sequences using the same experimental setup would be necessary for quantitative assessment of Lorentz force MRE efficiency and to identify any MR sequence related variability. On the reconstruction end, the LFE performed well at producing images with clear delineations of global heterogeneous structures, which is key in medical imaging but showed significant variance within homogeneous structures in Fig. 5, and artefacts due to wave reflection and interference in Fig. 4. This makes further mechanical testing and cross-comparisons of shear modulus estimates using standard actuation systems meaningless at this stage. Particularly, accurate evaluation of the impact of the source term on the retrieved properties should be addressed in full development by comparing Lorentz force MRE to other MRE procedures, as those listed in the previous paragraphs, which is beyond the scope of the presented study. Finally, the main limitation is the amplitude of the electrical current that is for now around 30 times above the human security threshold (2 mA peak-to-peak in humans). Nevertheless, future studies should focus on implementing our method on pre-clinical MR scanners with higher magnetic fields and stronger encoding gradients, which would allow reducing significantly the intensity of the current necessary to have robust displacement maps. For specific human applications, implementation on 7 T or higher gradient scanners might be envisaged.
Conclusion
In summary, this study demonstrated the feasibility of generating shear waves using the Lorentz force, producing stiffness maps from the displacement field, and reporting on potential impacts of the Lorentz force density on the reconstruction process. Overall, acquisitions were performed at two frequencies over three different phantom types implying removing and inserting phantom holders back into the experimental setup for each experimental condition. The consistent results obtained through multiple acquisitions indicate the robustness and reproducibility of the method. Despite significant variance in the shear modulus maps in Fig. 5, clear delineation of the two layers were observed, which is essential in medical imaging.
The LFE reconstructions could allow validating the working hypothesis to show the possibility of implementing the Lorentz force MRE method on a clinical scanner. This experimental condition using phantoms doped with physiological salt concentrations suggest that this method could be used for in vivo applications in MR elastography provided current intensities lower than 2 mA, as in transcranial alternative current stimulation (tACS) experiments56,57. This currently seems far from clinical applications but shows an interesting potential for preclinical studies where scanners with higher magnetic fields are available, thus producing a higher Lorentz force density for a given current value. Specific human applications on 7 T or higher magnetic field scanners might also be possible in the future.
Lowering the current intensity may also be reached by increasing other imaging parameters such as the motion encoding gradient amplitude (for instance to 285 mT m−1)58, the number of averaging over identical acquisitions repeated successively (to increase the signal-to-noise ratio), the number of MEG cycles (provided a suitable trade-off with the echo-time. Latitude thus remains to improve motion detection and reduce the current intensity. In terms of spatial selectivity, recent tACS results have shown that optimal electrode positioning on the scalp along with multi-frequency current application allow for delivering current to targeted deep regions into the brain59. Such current sequence design may be helpful in conceptualising Lorentz force MR brain elastography thus circumventing the low penetration of shear waves at high frequencies in conventional MR elastography due to significant attenuation. It may also provide the advantage of in situ localized vibrations to reduce diffraction artefacts by the skull when using an external vibration source.
As introduced above, a main advantage compared with conventional MR elastography is the fact that no external actuator is required thus excluding complex wave coupling between the external body and the imaged organ. The possibility of varying the shear wave frequency by a simple change in current frequency is another advantage. Eventually, Lorentz force MR elastography may be further developed as a remote mean to generate shear waves, notably with electric stimulation devices instead of electrodes, as in Ref. 18. At present, a framework involving tests of other MR sequences along with other reconstruction methods is currently being designed and may help sharpen the proposed technique. This will allow for quantitative characterization of the measured displacement fields using various encoding schemes and quantitative assessment of the impact of the Lorentz force density on the reconstructed mechanical parameter maps. The inclusion of MR-EIT protocols to Lorentz force MRE is also under development in an attempt to provide an original human or preclinical tool for assessment of both mechanical and electrical properties.
References
Friedrich-Rust, M., Poynard, T. & Castera, L. Critical comparison of elastography methods to assess chronic liver disease. Nat. Rev. Gastroenterol. Hepatol. 13, 402–411 (2016).
Hiscox, L. V. et al. Magnetic resonance elastography (MRE) of the human brain: technique, findings and clinical applications. Phys. Med. Biol. 61, R401–R437 (2016).
Mousavi, S. R., Rivaz, H., Sadeghi-Naini, A., Czarnota, G. J. & Samani, A. Breast ultrasound elastography using full inversion-based elastic modulus reconstruction. IEEE Trans. Comput. Imaging 3, 774–782 (2017).
Petitclerc, L., Sebastiani, G., Gilbert, G., Cloutier, G. & Tang, A. Liver fibrosis: Review of current imaging and MRI quantification techniques. J. Magn. Reson. Imaging 45, 1276–1295 (2017).
Lifshitz, E. M., Kosevich, A. M. & Pitaevskii, L. P. Theory of Elasticity 87–107 (Butterworth-Heinemann, 1986).
Kennedy, B. F., Kennedy, K. M. & Sampson, D. D. A review of optical coherence elastography: Fundamentals, techniques and prospects. IEEE J. Sel. Top. Quantum Electron. 20, 272–288 (2014).
Grasland-Mongrain, P. et al. Ultrafast imaging of cell elasticity with optical microelastography. Proc. Natl. Acad. Sci. 115, 861–866 (2018).
Gennisson, J. L., Deffieux, T., Fink, M. & Tanter, M. Ultrasound elastography: Principles and techniques. Diagn. Interv. Imaging 94, 487–495 (2013).
Muthupillai, R. et al. Magnetic resonance elastography by direct visualization of propagating acoustic strain waves. Science 269, 1854–1857 (1995).
Manduca, A., Muthupillai, R., Rossman, P.J., Greenleaf, J.F. & Ehman, R.L. in Proceedings of 3rd IEEE International Conference on Image Processing, Vol. 3 527–530 (IEEE, 1996).
Oliphant, T. E., Manduca, A., Ehman, R. L. & Greenleaf, J. F. Complex-valued stiffness reconstruction for magnetic resonance elastography by algebraic inversion of the differential equation. Magn. Reson. Med. 45, 299–310 (2001).
Manduca, A. et al. Magnetic resonance elastography: Non-invasive mapping of tissue elasticity. Med. Image Anal. 5, 237–254 (2001).
Sinkus, R. et al. Viscoelastic shear properties of in vivo breast lesions measured by MR elastography. Magn. Reson. Imaging 23, 159–165 (2005).
Van Houten, E. E. W., Paulsen, K. D., Miga, M. I., Kennedy, F. E. & Weaver, J. B. An overlapping subzone technique for MR-based elastic property reconstruction. Magn. Reson. Med. 42, 779–786 (1999).
Gallot, T. et al. Passive elastography: Shear-wave tomography from physiological-noise correlation in soft tissues. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 58, 1122–1126 (2011).
Zorgani, A. et al. Brain palpation from physiological vibrations using MRI. Proc. Natl. Acad. Sci. 112, 12917–12921 (2015).
Venkatesh, S.K. in Magnetic Resonance Elastography. (eds. S.K. Venkatesh & R.L. Ehman) 39–60 (Springer, 2014).
Perriñez, P. R., Kennedy, F. E., Houten, E. E. W. V., Weaver, J. B. & Paulsen, K. D. Magnetic resonance poroelastography: An algorithm for estimating the mechanical properties of fluid-saturated soft tissues. IEEE Trans. Med. Imaging 29, 746–755 (2010).
Kolipaka, A. et al. MR elastography as a method for the assessment of myocardial stiffness: Comparison with an established pressure–volume model in a left ventricular model of the heart. Magn. Reson. Med. 62, 135–140 (2009).
Mariappan, Y. K. et al. MR elastography of human lung parenchyma: Technical development, theoretical modeling and in vivo validation. J. Magn. Reson. Imaging 33, 1351–1361 (2011).
Asbach, P. et al. Assessment of liver viscoelasticity using multifrequency MR elastography. Magn. Reson. Med. 60, 373–379 (2008).
Klatt, D., Hamhaber, U., Asbach, P., Braun, J. & Sack, I. Noninvasive assessment of the rheological behavior of human organs using multifrequency MR elastography: a study of brain and liver viscoelasticity. Phys. Med. Biol. 52, 7281–7294 (2007).
Johnson, C. L. et al. 3D multislab, multishot acquisition for fast, whole-brain MR elastography with high signal-to-noise efficiency. Magn. Reson. Med. 71, 477–485 (2014).
Guo, J. et al. Towards an Elastographic Atlas of Brain Anatomy. PLoS ONE 8, e71807 (2013).
Garteiser, P. et al. Rapid acquisition of multifrequency, multislice and multidirectional MR elastography data with a fractionally encoded gradient echo sequence. NMR Biomed. 26, 1326–1335 (2013).
Runge, J. H. et al. A novel magnetic resonance elastography transducer concept based on a rotational eccentric mass: preliminary experiences with the gravitational transducer. Phys. Med. Biol. 64, 045007 (2019).
Wu, T., Felmlee, J. P., Greenleaf, J. F., Riederer, S. J. & Ehman, R. L. MR imaging of shear waves generated by focused ultrasound. Magn. Reson. Med. 43, 111–115 (2000).
Weaver, J. B. et al. Brain mechanical property measurement using MRE with intrinsic activation. Phys. Med. Biol. 57, 7275–7287 (2012).
Hirsch, S. et al. In vivo measurement of volumetric strain in the human brain induced by arterial pulsation and harmonic waves. Magn. Reson. Med. 70, 671–683 (2013).
Solamen, L. M. et al. Poroelastic mechanical properties of the brain tissue of normal pressure hydrocephalus patients during lumbar drain treatment using intrinsic actuation MR elastography. Acad. Radiol. 28, 457–466 (2021).
Grasland-Mongrain, P. et al. Imaging of shear waves induced by Lorentz force in soft tissues. Phys. Rev. Lett. 113, 038101–038105 (2014).
Grasland-Mongrain, P., Miller-Jolicoeur, E., Tang, A., Catheline, S. & Cloutier, G. Contactless remote induction of shear waves in soft tissues using a transcranial magnetic stimulation device. Phys. Med. Biol. 61, 2582–2593 (2016).
Sun, Z., Giammarinaro, B., Birer, A., Liu, G. & Catheline, S. Shear wave generation by remotely stimulating aluminum patches with a transient magnetic field and its preliminary application in elastography. IEEE Trans. Biomed. Eng., 1 (2020).
Basford, A. T., Basford, J. R., Kugel, J. & Ehman, R. L. Lorentz-force-induced motion in conductive media. Magn. Reson. Imaging 23, 647–651 (2005).
Griffiths, D. J. Introduction to Electrodynamics 4th edn. (Cambridge University Press, 2017).
Venkatesh, S. K., Yin, M. & Ehman, R. L. Magnetic resonance elastography of liver: technique, analysis, and clinical applications. J. Magn. Reson. Imaging 37, 544–555 (2013).
Hirsch, S., Braun, J. & Sack, I. in Magnetic Resonance Elastography: Physical Background And Medical Applications 41–59 (2016).
Royer, D. & Dieulesaint, E. Elastic Waves in Solids I: Free and Guided Propagation (Springer, 1999).
Doyley, M. M., Perreard, I., Patterson, A. J., Weaver, J. B. & Paulsen, K. M. The performance of steady-state harmonic magnetic resonance elastography when applied to viscoelastic materials. Med. Phys. 37, 3970–3979 (2010).
McIlvain, G. et al. Reliable preparation of agarose phantoms for use in quantitative magnetic resonance elastography. J. Mech. Behav. Biomed. Mater. 97, 65–73 (2019).
Knutsson, H., Westin, C. & Granlund, G. in Proceedings of 1st International Conference on Image Processing, Vol. 1 36–40 (IEEE, 1994).
Garcia, D. Robust smoothing of gridded data in one and higher dimensions with missing values. Comput. Stat. Data Anal. 54, 1167–1178 (2010).
Garcia, D. A fast all-in-one method for automated post-processing of PIV data. Exp. Fluids 50, 1247–1259 (2011).
Herráez, M. A., Burton, D. R., Lalor, M. J. & Gdeisat, M. A. Fast two-dimensional phase-unwrapping algorithm based on sorting by reliability following a noncontinuous path. Appl. Opt. 41, 7437–7444 (2002).
Ghiglia, D. C. & Romero, L. A. Robust two-dimensional weighted and unweighted phase unwrapping that uses fast transforms and iterative methods. J. Opt. Soc. Am. A 11, 107–117 (1994).
Kasim, M.F. (File Exchange, MathWorks; 1996).
Fröhlich, F. in Network Neuroscience. (ed. F. Fröhlich) 197–210 (Academic Press, 2016).
Hirsch, S., Braun, J. & Sack, I. Magnetic Resonance Elastography 61–129 (Wiley-VCH, 2016).
Hirsch, S., Braun, J. & Sack, I. Magnetic Resonance Elastography 199–228 (Wiley-VCH, 2016).
Aki, K. & Richards, P. G. Quantitative Seismology 63–117 (University Science Books, 2002).
Bercoff, J., Tanter, M. & Fink, M. Supersonic shear imaging: a new technique for soft tissue elasticity mapping. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 51, 396–409 (2004).
Chatelin, S., Charpentier, I., Corbin, N., Meylheuc, L. & Vappou, J. An automatic differentiation-based gradient method for inversion of the shear wave equation in magnetic resonance elastography: Specific application in fibrous soft tissues. Phys. Med. Biol. 61, 5000–5019 (2016).
Testu, J. et al. Viscoelastic power law parameters of in vivo human brain estimated by MR elastography. J. Mech. Behav. Biomed. Mater. 74, 333–341 (2017).
Seo, J. K. & Woo, E. J. Magnetic resonance electrical impedance tomography (MREIT). SIAM Rev. 53, 40–68 (2011).
Sajib, S. Z. K., Lee, M. B., Kim, H. J., Woo, E. J. & Kwon, O. I. Extracellular total electrolyte concentration imaging for electrical brain stimulation (EBS). Sci. Rep. 8, 290 (2018).
Bland, N. S. & Sale, M. V. Current challenges: the ups and downs of tACS. Exp. Brain Res. 237, 3071–3088 (2019).
Bächinger, M. et al. Concurrent tACS-fMRI reveals causal influence of power synchronized neural activity on resting state fMRI connectivity. J. Neurosci. 37, 4766 (2017).
Millward, J. M. et al. Tissue structure and inflammatory processes shape viscoelastic properties of the mouse brain. NMR Biomed. 28, 831–839 (2015).
Lee, S., Lee, C., Park, J. & Im, C.-H. Individually customized transcranial temporal interference stimulation for focused modulation of deep brain structures: A simulation study with different head models. Sci. Rep. 10, 11730–11731 (2020).
Acknowledgements
The authors would like to thank Professor Elijah Van Houten from the University of Sherbrooke, Québec, Canada for valued discussions, Professor Stéphane Bedwani from the University of Montreal, Québec, Canada for 3D printing of our experimental designs, and the Department of Radiology of the University of Montreal Hospital Research Center for providing free access to the MR scanner. This work was supported by postgraduate scholarships given to Dr Grasland-Mongrain by the Natural Sciences and Engineering Research Council of Canada (NSERC) and the Fondation pour la Recherche Médicale (FRM ARF20170938675). Partial funding for this work was also provided by NSERC (Grant #503381-16).
Author information
Authors and Affiliations
Contributions
G.F. and G.G. conducted the experiments and processed the data; G.F. designed the experiment, performed the simulations, and wrote the paper; G.F. and P.G.-M. proposed the method; and G.C. headed the project, provided general supervision and funding. All authors participated in reviewing the paper.
Corresponding author
Ethics declarations
Competing interests
Dr Guillaume Gilbert is a full-time employee of Philips Healthcare Canada. Other authors declare no competing interests.
Additional information
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Flé, G., Gilbert, G., Grasland-Mongrain, P. et al. Lorentz force induced shear waves for magnetic resonance elastography applications. Sci Rep 11, 12785 (2021). https://doi.org/10.1038/s41598-021-91895-9
Received:
Accepted:
Published:
DOI: https://doi.org/10.1038/s41598-021-91895-9
- Springer Nature Limited