Skip to main content
Log in

Intraimage reproducibility of measurements in the macular area using a computerized system

  • Published:
International Ophthalmology Aims and scope Submit manuscript

Abstract

An accurate analysis of the morphological changes which take place during pathological processes of the posterior pole is important for a correct diagnosis and therapeutic approach. The purpose of the study was to determine the intraobserver and interobserver reproducibility of the Image-net system 100 (Topcon, Japan) to take measurements on the retina.

The program ’Linear/Areal Measurement functions‘ of Image-net system 100 which is an image digitalization technique, was tested. Twelve patients were consecutively selected from the patients of the Retina Center of the Department of Ophthalmology, University of Genoa. Three images of each eye were taken from each subject and only the best image was used in this study. The intraobserver and interobserver reproducibility of both the distance between two pre-set points (linear measurement), and the perimeter and area of preselected retinal zones were calculated.

The repeatability (or intraobserver reproducibility) of the linear sizes was measured by the coefficient of variation and ranged from 0.32% to 7.38%, while the interobserver reproducibility ranged from 0.46% to5.22%. The repeatability and reproducibility of the perimeters ranged from 0.72% to 9.63% and from0.6%to 5.7%, respectively, while the repeatability and reproducibility of the areas ranged from 0.72% to 9.63%and from 0.6% to 5.7%, respectively.

Although the results were quite good, the quality of the image of the fundus and the number of observers influenced the coefficient of variation; furthermore, the anatomy of the areas to be measured and the computer’mouse‘ could increase the value of the coefficient of variation.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Friberg TR, Reakopf PG, Warnicki SE, Eller AW. Use of directly acquired digital fundus and fluorescein angiographic images in the diagnosis of retinal disease. Retina 1987; 7: 246–51.

    Google Scholar 

  2. Lichter PR. Variability of expert observers in evaluating the optic disc. Trans Am Ophthalmol Soc 1976; 74: 532–72.

    Google Scholar 

  3. Iester M, Traverso CE, Rolando M, Calabria G, Zingirian M. Study of the correlation between average value of optic disc parameters and their measurement variability using stereovideographic digital analysis. Ophthalmologica 1995; 209: 177–81.

    Article  Google Scholar 

  4. Takamota T, Schwarts B. Reproducibility of photogrammetric optic disc cup measurements. Invest Ophthalmol Vis Sci 1985; 26: 814–7.

    Google Scholar 

  5. Caprioli J, Klingbeil U, Sears M, Pope B. Reproducibility of optic disc measurements with computerized analysis of stereoscopic video image. Arch Ophthalmol 1986; 104: 1035–9.

    Google Scholar 

  6. Mikelberg FS, Douglas GR, Schulzer M, Cornsweet TN, Wijsman K. Reliability of optic disk topographic measurements recorded with a video-ophthalmograph. Am J Ophthalmol 1984; 98: 89–102.

    Google Scholar 

  7. Friberg TR, Campagna J. Central serious chorioretinopathy: an analysis of the clinical morphology using image-processing techniques. Graefe’s Arch Clin Exp Ophthalmol 1989; 227: 201–5.

    Article  Google Scholar 

  8. Eleanor E Faye. Low vision. In: Tasman W (ed) ‘Duane’s clinical ophthalmology’. Lippincott JB Company, Philadelphia, pp 46:6, 1993.

    Google Scholar 

  9. Iester M, Rolando M, Campagna P, Borgia L, Traverso CE, Calabria G. Valutazione della riproducibilità di analisi morfometriche papillari ottenute mediante il sistema IMAGE-net. 1991 nov 20–23 Roma. Roma: 71 Cong. Soc. Oftalmol Italiana, 1992: 359–66.

  10. Rolando M, Iester M, Campagna P, Borgia L, Traverso CE, Calabria G. Measurement variability in digital analysis of optic disc. Documenta Ophthalmologica 1994; 85: 211–22.

    Article  Google Scholar 

  11. Varma R, Steimann WC, Spaeth GL, Wilson RP. Variability in digital analysis of optic disc topography. Graefe’s Arch Clin Exp Ophthalmol 1989; 226: 435–42.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Iester, M., Mochi, B., Lai, S. et al. Intraimage reproducibility of measurements in the macular area using a computerized system. Int Ophthalmol 21, 153–159 (1997). https://doi.org/10.1023/A:1026490416530

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1023/A:1026490416530

Navigation