Abstract
In order to determine multiple elements by k 0 standardization-based neutron induced prompt gamma-ray analysis (PGA), measurements and evaluation of k 0 factors for the elements have been performed. The k 0 factors using Cl as a comparator for 27 elements, which are effective for the analysis by PGA, were measured with precision almost less than 3% using the cold and thermal guided neutron beams at JRR-3M. The accuracy of the k 0 factors were better than 10% except for the non-1/v elements of Cd and Sm. Neutron spectrum correction is necessary for common use of k 0 factors of the non-1/v elements.
Similar content being viewed by others
References
C. Yonezawa, Non-destructive elemental analysis, Z. B. Alfassi (Ed.), Blackwell Science, 2001, p. 58.
G. L. MolnÁr, Zs. RÉvay, R. L. Paul, R. M. Lindstrom, J. Radioanal. Nucl. Chem., 234 (1998) 21.
Zs. RÉvay, G. L. MolnÁr, T. Belgya, Zs. Kasztovszky, R. B. Firestone, J. Radioanal. Nucl. Chem., 244 (2000) 383.
R. L. Paul, R. M. Lindstrom, in: Proc. 2nd Intern. k0-Users Workshop, Ljubljana, Slovenia, 1997, p. 54.
M. A. Lone, S. F. Mughabghab, R. Paviotti-Corcuera, Summary Report of the 2nd IAEA Research Co-ordination Meeting on Development of a Database for Prompt Gamma-Ray Neutron Activation Analysis, 14.17 May 2001, Vienna, Austria, INDC(NDS)-424, 2001.
S. Raman, C. Yonezawa, H. Matsue, H. Iimura, N. Shinohara, Nucl. Instr. Meth. Phys. Res., A454 (2000) 389.
H. Matsue, C. Yonezawa, J. Radioanal. Nucl. Chem., 245 (2000) 189.
H. Matsue, C. Yonezawa, J. Radioanal. Nucl. Chem., 249 (2001) 11.
C. Yonezawa, Anal. Sci., 9 (1993) 185.
C. Yonezawa, A. K. Haji Wood, M. Hoshi, Y. Ito, E. Tachikawa, Nucl. Instr. Meth. Phys. Res., A329 (1993) 207.
Zs. RÉvay, G. L. MolnÁr, in: Summary Report of the 2nd IAEA Research Co-ordination Meeting on Development of a Database for Prompt Gamma-Ray Neutron Activation Analysis, 14.17 May 2001, Vienna, Austria, INDC(NDS)-424, 2001, p. 57.
T. Nakagawa, K. Shibata, S. Chiba, T. Fukahori, Y. Nakajima, Y. Kikuchi, T. Kawano, Y. Kanda, T. Ohsawa, H. Matsunobu, M. Kawai, A. Zukeran, T. Watanabe, S. Igarasi, K. Kosako, T. Asami, J. Nucl. Sci. Technol., 32 (1995) 1259.
R. B. Firestone, in: Summary Report of the 1st IAEA Research Co-ordination Meeting on Development of a Database for Prompt Gamma-Ray Neutron Activation Analysis, 2-4 November 1999, Vienna, Austria, INDC(NDS)-411, 2000, p. 45.
S. F. Mughabghab, M. Divadeenam, N. E. Holden, Neutron Cross Sections, Vol. 1: Neutron Resonance Parameters and Thermal Cross Sections, Part A, Z = 1-60, Academic Press, New York, 1981.
P. De Bievre, P. D. P. Taylor, Intern. J. Mass Spectrom. Ion Phys., 123 (1993) 149.
M. A. Lone, R. A. Leavitt, D. A. Harrison, Atomic Data Tables, 26 (1981) 511.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Matsue, H., Yonezawa, C. Measurement and evaluation of k0 factors for PGA at JAERI. Journal of Radioanalytical and Nuclear Chemistry 257, 565–571 (2003). https://doi.org/10.1023/A:1025492615803
Issue Date:
DOI: https://doi.org/10.1023/A:1025492615803