Abstract
A method of replacing the triple point of mercury by the melting point of gallium when reproducing the ITS-90 International Temperature Scale is considered. A method of estimating the systematic error and of determining the conditions under which the method can be used is proposed.
Similar content being viewed by others
References
P. V. Novitskii and I. A. Zograf, Estimating the Errors of Means of Measurement [in Russian], Énergoatomizdat, Leningrad (1985).
Handbook of Algorithms and Programs in BASIC for Personal Computers [in Russian], Nauka, Moscow (1987).
F. P. Tarasenko, Nonparametric Statistics [in Russian], Tomsk (1976).
A. V. Yashin, Kontrol'no-Izmeritel'nye Pribory i Sistemy, No. 2, 28 (2000).
Rights and permissions
About this article
Cite this article
Marinko, S.V., Yashin, A.V. A Method of Reproducing the Triple Point of Mercury in the Working Standards of the Unit of Temperature. Measurement Techniques 44, 181–188 (2001). https://doi.org/10.1023/A:1010969526161
Issue Date:
DOI: https://doi.org/10.1023/A:1010969526161