Abstract
We have shown that it is sufficiently accurate to use the MCNP peak-to-total calibration results to correct for cascade summing effects in a gamma-spectrum. Also, it is sufficient to use only approximate detector characterization data with empirical peak-to-total to obtain good cascade summing correction results. The intrinsic P/T-curve for detectors with the same efficiency is very similar and it may be considered a common characteristic of the whole detector's family with given efficiency.
Similar content being viewed by others
References
D. S. Andreev, K. I. Erokhina, V. S. Zvonov, I. Kh. Lemberg, Instr. Expt.Techn., 15 (1972) 1358.
J. C. Verplancke, Nucl. Instr. Meth. Phys. Res., 96 (1971) 557.
R. J. Gehrke, R. C. Helmer, R. C. Greenwood, Nucl. Instr. Meth. Phys. Res., 147 (1977) 405.
K. Debertin, U. SchÖtzig, Nucl. Instr. Meth. Phys. Res., 158 (1979) 471.
V. P. Kolotov, V. V. Atrashkevich, S. J. Gelsema, J. Radioanal. Nucl. Chem., 210 (1996) 183.
V. P. Kolotov, M. J. Koskelo, J. Radioanal. Nucl. Chem., 233 (1998) 95.
F. L. Bronson, B. M. Young, Mathematical Calibrations of Ge Detectors and the Instruments that Use Them, Proc. 5th Annual NDA/NDE Waste Characterization Conference, Salt Lake City, UT, Jan 11, 1997.
J. F. Briesmeister (Ed.), MCNP. A General Monte Carlo N-Particle Transport Code, Version 4a, Los Alamos National Laboratory Report LA-12625-M, November, 1993.
R. Venkataraman, F. Bronson, V. Atrashkevich, B. M. Young, M. Field, Nucl. Instr. Meth. Phys. Res., A442 (1999) 450.