Reproducibility of Bipolar Endocavitary Electrogram Measurements at sites of Radiofrequency Energy Delivery in Patients with the Wolff-Parkinson-White Syndrome

  • Published:
Journal of Interventional Cardiac Electrophysiology Aims and scope Submit manuscript

Abstract

Background: Radiofrequency ablation of atrioventricular accessory pathway is widely used to cure patients with the Wolff-Parkinson-White syndrome. The site of successful ablation is determined using electrophysiological parameters, endocavitary bipolar electrogram measurements being the most commonly used. Interobserver reproducibility of these measurements may limit the reliability of ablation criteria based upon bipolar measurements only but, to our knowledge, this reproducibility has not been evaluated so far. Such was the aim of this study.

Methods: Three independent observers reviewed the bipolar electrograms recorded at sites were radiofrequency energy was delivered (successfully or not) in 28 consecutive patients with the Wolff-Parkinson-White syndrome. In each tracing, 4 intervals were measured: (1) A0V

0 (onset of the atrial electrogram to onset of the ventricular electrogram), (2) A

aV

a (activation time of the atrial electrogram to activation time of the ventricular electrogram), (3) V

0-QRS (onset of the ventricular electrogram to onset of delta wave on the surface ECG) and (4) V

a-QRS (activation time of the ventricular electrogram to onset of delta wave on the surface ECG).

Results: The interobserver reproducibility was low since only 50% of A

0V

0 intervals were measured with an interobserver difference lower than 10ms and up to 43% of V

a-QRS intervals were measured with an interobserver difference greater than 30ms. The reproducibility of interval measurement was graded from the highest to the lowest as follows: A

0V

0, A

aV

a, V

0-QRS and V

a-QRS (Chi-square statistic, χ

2

= 71.72, p < 0.0001). Kappa values were lower than 0.40, indicating a poor interobserver reproducibility.

Conclusions: Our study suggests that interobserver reproducibility of

only bipolar electrograms interval measurements at sites of radiofrequency ablation of atrioventricular accessory pathway is poor, which limits the reliability of bipolar criteria to predict a successful ablation site.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Calkins H, Kim YN, Schmaltz S, Sousa J, El-Atassi R, Leon A, Kadish A, Landberg JJ, Morady F. Electrogram criteria for identification of appropriate target sites for radiofrequency catheter ablation of accessory atrioventricular connections. Circulation 1992;85:565-573.

    Google Scholar 

  2. Silka MJ, Kron J, Halperin BD, Griffith K, Crandall B, Oliver RP, Walance CG, McAnulaty JH. Analysis of local electrogram characteristics correlated with successful radiofrequency catheter ablation of accessory atrioventricular pathways. PACE 1992;15:1000-1006.

    Google Scholar 

  3. HaÏssaguerre M, Dartigues JF, Warin JF, Le Metayer P, Monserrat P, Salamon R. Electrogrampatterns predictive of successful catheter ablation of accessory pathways. Value of unipolar recording mode. Circulation 1991;84:188-202.

    Google Scholar 

  4. HaÏssaguerre M, Fisher B, Warin JF, Dartigues JF, Le Metayer P, Egloff P. Electrogram patterns predictive of successful radiofrequency catheter ablation of accessory pathways. PACE 1992;15:2138-2145.

    Google Scholar 

  5. Simmers TA, Hauer RNW, Waever EFD, Wittkampf FHM, Robles de Medina EO. Unipolar electrogram models for prediction of outcome in radiofrequency ablation of accessory pathways. PACE 1994;17:186-198.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

de Chillou, C., Magnin-Poull, I., Sadoul, N. et al. Reproducibility of Bipolar Endocavitary Electrogram Measurements at sites of Radiofrequency Energy Delivery in Patients with the Wolff-Parkinson-White Syndrome. J Interv Card Electrophysiol 2, 377–381 (1998). https://doi.org/10.1023/A:1009772923689

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1023/A:1009772923689

Keywords

Navigation