Inappropriate Shocks from Implanted Cardioverter Defibrillators Caused by Sensing of Diaphragmatic Myopotentials

  • Published:
Journal of Interventional Cardiac Electrophysiology Aims and scope Submit manuscript

Abstract

Inappropriate shocks remain one of the major problems associated with the use of implantable cardioverter defibrillators (ICD). We describe 3 patients who had inappropriate shocks due to oversensing of diaphragmatic myopotentials. In all 3, oversensing could be demonstrated during deep inspiration or Valsalva maneuver. The problem was remedied in one by insertion of a separate rate sensing lead in the right ventricular outflow tract, in the second by decreasing the rate of antibradycardia pacing of the ICD, and in the third by decreasing the sensitivity of the ICD. We conclude that the possibility of sensed diaphragmatic myopotentials should be considered at the time of ICD implant so that appropriate preventive measures can be taken.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Hamer ME, Clair WK, Wilkinson WE, Greenfield RA, Pritchett LC, Page RL. Evaluation of outpatients experiencing implantable cardioverter defibrillator shocks associated with minimal symptoms. PACE 1994;17:938-943.

    Google Scholar 

  2. Reiter MJ, Mann DE. Sensing and tachyarrhythmia detection problems in implantable cardioverter defibrillators. J Cardiovasc Electrophysiol 1996;7:542-558.

    Google Scholar 

  3. Peters RW, Foster AH, Shorofsky SR, Nolan DA, Gold MR. Spurious discharges due to late insulation break in endocardial sensing leads for cardioverter defibrillators. PACE 1995;18:478-481.

    Google Scholar 

  4. Sticherling C, Klingenheben T, Hohnloser SH. An unusual cause of an ICD shock. PACE 1997;20:2265-2267.

    Google Scholar 

  5. Barold SS, Ong LS, Falkoff MD, Heinle RA. Inhibition of bipolar demand pacemaker by diaphragmatic myopotentials. Circulation 1977;56:679-683.

    Google Scholar 

  6. Barold SS, Falkoff MD, Ong LS, Heinle RA. Diaphragmatic myopotential inhibition in multiprogrammable unipolar and bipolar pacemakers. In: Steinbach K, ed. Cardiac Pacing Proceedings of the VIIIth World Symposium on Cardiac Pacing. Durmstadt, Germany: Steinkopff Verlag, 1983;537-540.

    Google Scholar 

  7. Damle RS, Mann DE, Kelly PA, Landers M J Reiter MJ. Noise during bradycardia pacing with Ventritex implantable cardioverter defibrillators: Role of diaphragmatic myopotentials and clinical implications. Circulation 1997;I-721. (Abstract).

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Peters, R.W., Cooklin, M., Brockman, R. et al. Inappropriate Shocks from Implanted Cardioverter Defibrillators Caused by Sensing of Diaphragmatic Myopotentials. J Interv Card Electrophysiol 2, 367–370 (1998). https://doi.org/10.1023/A:1009768822780

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1023/A:1009768822780

Navigation