Skip to main content

Wind tunnel comparison of the collection efficiency ofthree Hirst-type volumetric sampler drum coatings

Abstract

A large number of sites worldwide measure airbornepollen and fungal spore concentrations using aHirst-type volumetric impaction sampler. In Europe anumber of adhesive drum coatings are routinely used.The drum coating routinely used in the UK, V/P,consisting of 90% petroleum Jelly (Vaseline)/10%paraffin wax, has not previously been compared withother coatings, for fungal spores or pollens. Thecollection efficiency of different coatings wascompared in a wind tunnel, at IACR-Rothamsted. Sporesof different species were released and trapped inidentical miniature suction traps which reproduce theaerodynamic features of a full size Hirst-type sporetrap. The trapping surfaces were coated in one ofthree commonly used drum coatings: (1) V/P; (2) SiliconeFluid; (3) Glycerol Jelly. These are all used in Europeand both V/P and Glycerol Jelly are used the UK.

Four spore types, Lycopodium clavatum (moss);Alternaria alternata; Botrytis cinerea;and Calvatia excipuliformis, which represent thesize range of spores most frequently found in the air,were used. With the exception of B cinerea, thenumbers of spores collected on Glycerol Jelly weresignificantly smaller than on the other two coatings.For all spore types, the numbers caught on the V/P andSilicone Fluid were not significantly different.

This is a preview of subscription content, access via your institution.

References

  1. British Aerobiology Federation: 1995, Airborne Pollens and Fungal Spores: A Guide to Trapping and Counting. Kimberly Clark Ltd., Aylesford, ISBN: 0-9525617-0-0.

    Google Scholar 

  2. Comtois P. and Mandrioli P.: 1997, Pollen capture media: a comparative study. Aerobiologia 13, 149-154.

    Google Scholar 

  3. Fitt D.L., Walklate J.J., McCartney H.A., Bainbridge A., Creighton N.F., Hirst J.M., Lacey M.E. and Legg B.J.: 1986, A raintower and wind tunnels for studying the dispersal of plant pathogens by rain and wind. Ann. Appl. Biol. 109, 661-671.

    Google Scholar 

  4. Galán C. and Dominguez-Vilches E.: 1997, The capture media in aerobiological sampling. Aerobiologia 13, 155-160

    Google Scholar 

  5. Käpylä M.: 1989, Adhesive and mounting media in aerobiological sampling. Grana 28, 215-218.

    Google Scholar 

  6. McCartney H.A. and Aylor D.E.: 1987, Relative contribution of sedimentation and impaction to the deposition of particles in a crop canopy. Agric. Forest. Meterol. 40, 342-358.

    Google Scholar 

Download references

Author information

Affiliations

Authors

Corresponding author

Correspondence to F.E. Warner.

Rights and permissions

Reprints and Permissions

About this article

Cite this article

Warner, F., McCartney, H. & Emberlin, J. Wind tunnel comparison of the collection efficiency ofthree Hirst-type volumetric sampler drum coatings. Aerobiologia 16, 25–28 (2000). https://doi.org/10.1023/A:1007637105259

Download citation

  • adhesive
  • capture efficiency
  • fungal spores
  • Glycerol Jelly
  • Vaseline/praffin wax