Inter- and Intraobserver Reliability of Pelvic Obliquity Measurement Methods in Patients With Cerebral Palsy

  • Published:
Spine Deformity Aims and scope Submit manuscript

Abstract

Study Design

Retrospective, chart review.

Objectives

The identification and management of pelvic obliquity in neuromuscular scoliosis can be difficult; therefore, appropriate evaluation of this pathology is important. Variations in presentation have resulted in various methods of measurement, without a consensus or gold standard measurement. This study is the evaluation of reliability of five methods commonly used to determine pelvic obliquity in the frontal plane.

Summary of Background Data

Previous studies have used five different methods (ie, the Maloney method, the O’Brien method, the Osebold method, the Allen and Ferguson method, and the Lindseth method) to determine pelvic obliquity in the frontal plane.

Methods

Radiographic images of 50 patients with neuromuscular scoliosis involving the pelvic girdle were identified and evaluated by a team of 5 raters. Each rater was instructed to apply five commonly used methods of measuring pelvic obliquity. The same raters were asked to rate the same radiographs again after a minimum of one month. The interobserver reliability was assessed using the intraclass correlation coefficient (ICC) and the intraobserver analysis was assessed using the Pearson correlation. A rating greater than 0.8 was considered excellent reliability.

Results

The Maloney method showing the highest interobserver reliability of ICC of 0.965 and 0.964. The Lindseth method had the lowest ICC, but all the methods were considered highly reliable by the rating classification system. The Maloney method also showed the highest intraobserver reliability, ranging from 0.845 to 0.962. The Allen and Ferguson method had the lowest intraobserver reliability.

Conclusions

Many of the methods tested had excellent reliability in this study, but our data suggest the Maloney method is the most reliable method of measuring pelvic obliquity on a frontal view radiograph.

Level of Evidence

Level III, retrospective, comparative study.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Whitaker C, Burton DC, Asher M. Treatment of selected neuromuscular patients with posterior instrumentation and arthrodesis ending with lumbar pedicle screw anchorage. Spine 2000;25:2312–8.

    Article  CAS  Google Scholar 

  2. McCarthy J, D’Andrea LP, Betz RR, et al. Scoliosis in the child with cerebral palsy. J Am Acad Orthop Surg 2006;14:367–75.

    Article  Google Scholar 

  3. Mubarak SJ, Morin WD, Leach J. Spinal fusion in Duchenne muscular dystrophy—fixation and fusion to the sacropelvis? J Pediatr Orthop 1993;13:752–7.

    Article  CAS  Google Scholar 

  4. Yazici M, Asher MA, Hardacker JW. The safety and efficacy of Isola-Galveston instrumentation and arthrodesis in the treatment of neuromuscular spinal deformities. J Bone Joint Surg Am 2000;82:524–43.

    Article  CAS  Google Scholar 

  5. Maloney WJ, Rinsky LA, Gamble JG. Simultaneous correction of pelvic obliquity, frontal plane, and sagittal plane deformities in neuromuscular scoliosis using a unit rod with segmental sublaminar wires: a preliminary report. J Pediatr Orthop 1990;10:742–9.

    Article  CAS  Google Scholar 

  6. Osebold WR, Mayfield JK, Winter RB, et al. Surgical treatment of paralytic scoliosis associated with myelomeningocele. J Bone Joint Surg Am 1982;64:841–56.

    Article  CAS  Google Scholar 

  7. Allen BL, Ferguson RL. The Galveston technique of pelvic fixation with L-rod instrumentation of the spine. Spine 1984;9:388–94.

    Article  Google Scholar 

  8. Lindseth RE. Posterior iliac osteotomy for fixed pelvic obliquity J Bone Joint Surg Am 1978;60:17–22.

    Article  CAS  Google Scholar 

  9. O’Brien M, Kuklo T, Blanke K, et al; Spinal Deformity Study Group. Radiographic measurement manual. Memphis, TN: Medtronic Sofamor Danek USA; 2008.

    Google Scholar 

  10. Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics 1977;33:159–74.

    Article  CAS  Google Scholar 

  11. Gupta MC, Wijesekera S, Sossan A, et al. Reliability of radiographic parameters in neuromuscular scoliosis. Spine 2007;32:691–5.

    Article  Google Scholar 

  12. Heidt C, Hollander K, Wawrzuta J, et al. The radiological assessment of pelvic obliquity in cerebral palsy and the impact on hip development. Bone Joint J 2015;97-B: 1435–40.

    Article  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to M. Wade Shrader MD.

Additional information

Author disclosures: none.

Financial Disclosure: No financial support was provided for this study.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Shrader, M.W., Andrisevic, E.M., Belthur, M.V. et al. Inter- and Intraobserver Reliability of Pelvic Obliquity Measurement Methods in Patients With Cerebral Palsy. Spine Deform 6, 257–262 (2018). https://doi.org/10.1016/j.jspd.2017.10.001

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1016/j.jspd.2017.10.001

Keywords

Navigation