Abstract
We present a newly developed instrument that uses proton-transfer ion trap-mass spectrometry (PIT-MS) for on-line trace gas analysis of volatile organic compounds (VOCs). The instrument is based on the principle of proton-transfer reaction-mass spectrometry (PTR-MS): VOCs are ionized using PTRs and detected with a mass spectrometer. As opposed to a quadrupole mass filter in a PTR-MS, the PIT-MS instrument uses an IT-MS, which has the following advantages: (1) the ability to acquire a full mass spectrum in the same time as one mass with a quadrupole and (2) extended analytical capabilities of identifying VOCs by performing collision-induced dissociation (CID) and ion molecule reactions in the IT. The instrument described has, at its current status, limits of detection between 0.05 and 0.5 pbbv for 1-min measurements for all tested VOCs. The PIT-MS was tested in an ambient air measurement in the urban area of Boulder, Colorado, and intercompared with PTR-MS. For all measured compounds the degree of correlation between the two measurements was high (r 2 > 0.85), except for acetonitrile (CH3CN), which was close to the limit of detection of the PIT-MS instrument. The two measurements agreed within less than 25%, which was within the combined measurement uncertainties. Automated CID measurements on m/z 59 during the intercomparison were used to determine the contributions of acetone and propanal to the measured signal; both are detected at m/z 59 and thus are indistinguishable in PTR-MS. It was determined that m/z 59 was mainly composed of acetone. An influence of propanal was detected only during a high pollution event. The advantages and future developments of PIT-MS are discussed.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Hewitt, C. N. Reactive Hydrocarbons in the Atmosphere; Academic Press: San Diego, 1999.
Crutzen, P. J. Role of NO and NO2 in the Chemistry of the Troposphere and Stratosphere. Annu. Rev. Earth Planet. Sci 1979, 7, 443–472.
Lindinger, W.; Fall, R.; Karl, T. Environmental, Food and Medical Applications of Proton-Transfer-Reaction Mass Spectrometry (PTR-MS). Adv. Gas Phase Ion Chem. 2001, 4, 1–48.
Lindinger, W.; Hansel, A.; Jordan, A. Online Monitoring of Volatile Organic Compounds at pptv Levels by Means of Proton-Transfer-Reaction Mass Spectrometry (PTR-MS): Medical Applications, Food Control and Environmental Research. Int. J. Mass. Spectrom. Ion Process. 1998, 173, 191–241.
Williams, J.; Pöschl, U.; Hansel, A.; Holzinger, R.; Warneke, C.; Lindinger, W. An Atmospheric Chemistry Interpretation of Mass Scans Obtained from a Proton Transfer Mass Spectrometer Flown Over the Tropical Rainforest of Surinam. J. Atmos. Chem. 2001, 38, 133–166.
Karl, T.; Fall, R.; P. J., C.; Jordan, A.; Lindinger, W. High Concentrations of Reactive Biogenic VOCs at a High Altitude Site in Late Autumn. Geophys. Res. Lett. 2001, 28, 507–510.
Warneke, C.; de Gouw, J. A.; Goldan, P. D.; Kuster, W. C.; Fall, R. Validation of Atmospheric Measurements by Proton-Transfer-Reaction Mass Spectrometry Using a Gas-chromatographic Pre-separation Method. Environ. Sci. Tech. 2003, 37, 2494–2501.
de Gouw, J. A.; Warneke, C.; Karl, T.; Eerdekens, G. ; van der Veen, C.; Fall, R. Sensitivity and Specificity of Atmospheric Trace Gas Detection by Proton-Transfer-Reaction Mass Spectrometry. Int. J. Mass Spectrom. 2003, 365–382.
Prazeller, P.; Palmer, P. T.; Boscaini, E.; Jobson, T.; Alexander, M. Proton Transfer Reaction Ion Trap Mass Spectrometry. Rapid Commun. Mass Spectrom. 2003, 17, 1593–1599.
Warneke, C.; Rosen, S.; Lovejoy, E. R.; de Gouw, J. A.; Fall, R. Two Additional Advantages of Proton Transfer Ion Trap Mass Spectrometer (PT-ITMS). Rapid Commun. Mass Spectrom. 2004, 18, 133–134.
Warneke, C.; van der Veen, C.; Luxembourg, S. L.; de Gouw, J. A.; Kok, A. Measurements of Benzene and Toluene in Ambient Air Using Proton-transfer-reaction Mass Spectrometry: Calibration, Humidity Dependence and Field Intercomparison. Int. J. Mass Spectrom. 2001, 207, 167–182.
Lovejoy, E. R.; Wilson, R. R. Kinetic Studies of Negative Ion Reactions in a Quadrupole Ion Trap: Absolute Rate Coefficients and Ion Energies. J. Phys. Chem. A 1998, 102, 2309–2315.
Curtius, J.; Froyd, K. D.; Lovejoy, E. R. Cluster Ion Thermal Decomposition (I): Experimental Kinetics Study and Ab Initio Calculations for HSO −4 (H2SO4)x(HNO3)y. J. Phys. Chem. A 2001, 105, 10867–10873.
March, R. E.; Hughes, P. J. Practical Aspects of Ion Trap Mass Spectrometry, Vol. I–III; Wiley: New York; CRC: Boca Raton, FL, 1995.
Cheng, L.; Wang, T. C. L.; Ricca, T. C.; Marshall, A. G. Phase-modulated Stored Waveform Inverse Fourier Transform Excitation for Trapped Ion Mass Spectrometry. Anal. Chem. 1987, 449–454.
Stafford, G.; Kelley, P.; Syka, J.; Reynolds, W.; Todd, J. Recent Improvements In and Analytical Applications of Advanced Ion Trap Technology. Int. J. Mass Spectrom. Ion Process. 1984, 60.
de Gouw, J. A.; Goldan, P. D.; Warneke, C.; Kuster, W. C.; Roberts, J. M.; Marchewka, M.; Bertman, S. B.; Pszenny, A. A. P.; Keene, W. C. Validation of Proton-Transfer-Reaction Mass Spectrometry (PTR-MS) Measurements of Gas-Phase Organic Compounds in the Atmosphere During the New England Air Quality Study (NEAQS) in 2002. J. Geophys. Res. 2003, 108, 10.1029/2003JD003863.
Author information
Authors and Affiliations
Corresponding author
Additional information
Published online June 24, 2005
Rights and permissions
About this article
Cite this article
Warneke, C., de Gouw, J.A., Lovejoy, E.R. et al. Development of proton-transfer ion trap-mass spectrometry: on-line detection and identification of volatile organic compounds in air. J Am Soc Mass Spectrom 16, 1316–1324 (2005). https://doi.org/10.1016/j.jasms.2005.03.025
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1016/j.jasms.2005.03.025