Skip to main content
Log in

Development of respiratory gated myocardial SPECT system

  • Original Articles
  • Published:
Journal of Nuclear Cardiology Aims and scope

Abstract

Background

The superposition of the diaphragm and abdominal structures on the inferior wall of the left ventricle has often distorted single photon emission computed tomography (SPECT). We developed a respiratory gated SPECT (RGS) system to diminish artifacts caused by overlap between the inferior wall and upper abdomen and have validated its feasibility for clinical use.

Methods and Results

A 2-detector SPECT system equipped with a respiratory monitor based on impedance plethysmography and an original triggering apparatus was used for RGS in 7 healthy male volunteers. A pulse triggered 100 ms after every expiratory peak was processed in a SPECT system as well as an electrocardiogram (ECG) gating pulse. Inspiratory and expiratory frames were determined using the respiratory curve derived from fluctuation of the gall bladder uptake. Both sets of images were reoriented into short-axis and vertical long-axis slices. For quantification, data were reconstructed into polar plots and count density estimated in 9 myocardial segments. The mean percentage uptake of inferior segments at inspiration was significantly greater than that at expiration (81±8.3 versus 76±7.1; P<.0001). The inferior-lateral activity ratio improved from 0.78 at expiration to 0.81 at inspiration (P<.01). The coefficient of variance for each segment of inspiratory data was significantly smaller than that at expiration, indicating improved homogeneity of tracer distribution. The lowest cutoff threshold of the tomograms to separate the inferior uptake from that of the upper abdomen was significantly lower at inspiration than at expiration, suggesting smaller scatter from abdominal structures on inspiratory images.

Conclusions

RGS yielded improved tracer uptake of the inferior wall in healthy male subjects and may be suitable as an alternative method for attenuation and scatter correction. However, further clinical validation is needed.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Goldensohn ES, Zablow L. An electrical impedance spirometer. J Appl Physiol 1959;14:463–4.

    PubMed  CAS  Google Scholar 

  2. Baker LE, Geddes LA. The measurement of respiratory volumes in animals and man with use of electrical impedance. Ann N Y Acad Sci 1970;170:667–88.

    Article  Google Scholar 

  3. Zubal IG, Bizais Y, Bennett GW, Brill AB. Dual gated cardiac images. IEEE Trans Nucl Sci 1984;31:566–9.

    Article  Google Scholar 

  4. Ehman RL, McNamara MT, Pallack M, Hricak H, Higgins CB. Magnetic resonance imaging with respiratory gating: techniques and advantages. AJR Am J Roentgenol 1984;143:1175–82.

    PubMed  CAS  Google Scholar 

  5. Natarajan TK, Wise RA, Karam M, Permutt S, Wagner HN Jr. Immediate effect of expiratory loading on left ventricular stroke volume. Circulation 1987;75:139–45.

    PubMed  CAS  Google Scholar 

  6. Kim BH, Ishida Y, Tsuneoka Y, et al. Effects of spontaneous respiration on right and left ventricular function: evaluation by respiratory and ECG gated radionuclide ventriculography. J Nucl Med 1987;28:173–7.

    PubMed  CAS  Google Scholar 

  7. Mori M, Murata K, Takahashi M, et al. Accurate contiguous sections without breath-holding on chest CT: value of respiratory gating and ultrafast CT. AJR Am J Roentgenol 1994;162:1057–62.

    PubMed  CAS  Google Scholar 

  8. Tung C-H, Gullberg GT, Zeng GL, Christian PE, Datz FL, Morgan HT. Non-uniform attenuation correction using simultaneous transmission and emission converging tomography. IEEE Trans Nucl Sci 1992;39:1134–43.

    Article  CAS  Google Scholar 

  9. Frey EC, Tsui BM, Perry JR. Simultaneous acquisition of emission and transmission data for improved thallium-201 cardiac SPECT imaging using a technetium-99m transmission source. J Nucl Med 1992;33:2238–45.

    PubMed  CAS  Google Scholar 

  10. Liew SC, Hasegawa BH, Brown JK, Lang TF. Noise propagation in SPECT images reconstructed using an iterative maximum-likelihood algorithm. Phys Med Biol 1993;38:1713–26.

    Article  PubMed  CAS  Google Scholar 

  11. Ficaro EP, Fessler JA, Ackermann RJ, Rogers WL, Corbett JR, Schwaiger M. Simultaneous transmission-emission thallium-201 cardiac SPECT: effect of attenuation correction on myocardial tracer distribution. J Nucl Med 1995;36:921–31.

    PubMed  CAS  Google Scholar 

  12. Ficaro EP, Fessler JA, Shreve PD, Kritzman JN, Rose PA, Corbett JR. Simultaneous transmission/emission myocardial perfusion tomography. Diagnostic accuracy of attenuation-corrected 99mTc-sestamibi single-photon emission computed tomography. Circulation 1996;93:463–73.

    PubMed  CAS  Google Scholar 

  13. Prvulovich EM, Lonn AH, Bomanji JB, Jarritt PH, Ell PJ. Effect of attenuation correction on myocardial thallium-201 distribution in patients with a low likelihood of coronary artery disease. Eur J Nucl Med 1997;24:266–75.

    PubMed  CAS  Google Scholar 

  14. He ZX, Scarlett MD, Mahmarian JJ, Verani MS. Enhanced accuracy of defect detection by myocardial single-photon emission computed tomography with attenuation correction with gadolinium 153 line sources: evaluation with a cardiac phantom. J Nucl Cardiol 1997;4:202–10.

    Article  PubMed  CAS  Google Scholar 

  15. Kluge R, Sattler B, Seese A, Knapp WH. Attenuation correction by simultaneous emission-transmission myocardial single-photon emission tomography using a technetium-99m-labelled radiotracer: impact on diagnostic accuracy. Eur J Nucl Med 1997;24:1107–14.

    PubMed  CAS  Google Scholar 

  16. Kijima T, Kumita S, Mizumura S, et al. The usefulness and the problems of attenuation correction using simultaneous transmission and emission data acquisition method: studies on normal volunteers and phantom. Kaku Igaku 1997;34:167–76.

    PubMed  CAS  Google Scholar 

  17. Heller EN, DeMan P, Liu YH, et al. Extracardiac activity complicates quantitative cardiac SPECT imaging using a simultaneous transmission-emission approach. J Nucl Med 1997;38:1882–90.

    PubMed  CAS  Google Scholar 

  18. Prigent FM, Hyun M, Berman DS, Rozanski A. Effect of motion on thallium-201 SPECT studies: A simulation and clinical study. J Nucl Med 1993;34:1845–50.

    PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Cho, K., Kumiata, Si., Okada, S. et al. Development of respiratory gated myocardial SPECT system. J Nucl Cardiol 6, 20–28 (1999). https://doi.org/10.1016/S1071-3581(99)90061-2

Download citation

  • Received:

  • Accepted:

  • Issue Date:

  • DOI: https://doi.org/10.1016/S1071-3581(99)90061-2

Key Words

Navigation