Abstract
An electrospray interface is described that provides high sensitivity and signal stability for mass spectrometric detection of analytes in solvents with high water content including 100% water. The electrospray capillary tip section is heated close to the boiling point of the solvent. An approximately 20°C hotter airstream, with flow coaxial to the electrospray tip and codirectional to the electrospray.. is also used. With this arrangement, the analyte signal sensitivity and stability obtained with neat water is equal to that obtained with neat methanol. The heated electrospray also affords the use of a wide range of flow rates: 1–100 μL/min.
Article PDF
Similar content being viewed by others
References
Smith, R. D.; Loo, J. A.; Edmonds, C. G.; Barniaga, C. J.; Udseth, H. R. Anal. Chern. 1990, 62, 882.
Smith, D. P. H. IEEE Trans. Ind. Appl. 1986, IA-22, 527.
Ikonomou, M. G.; Blades, A. T.; Kebarle, P. J. Am. Soc. Mass Spectrom. 1991, 2, 497.
Gomez, A.; Tang, K. Physics of Fluids (in print).
Tang, L.; Kebarle, P. Anal. Chem. 1993, 65, 3654. Kebarle, P.; Tang, L. Anal. Chem., 1993, 65, 972A.
Fernandez de la Mora, J., Rossel-Liompart, J. Proceedings of the 39th ASMS Conference on Mass Spectrometry and Allied Topics; Nashville, 1991; p 441.
Davies, C. N. In Fundamentals of Aerosol Science; Shaw, D. T., Ed.; Wiley-Intersdence: Wiley, New York, 1978; p 154.
Ikonomou, M. G.; Blades, A. T.; Kebarle, P. Anal. Chern. 1991, 63, 1989–1998.
Iribarne, J. V.; Thomson, B. A. J. Phys. Chern. 1976, 64, 2287; Thomson, B. A.; Iribarne, J. V. J. Phys. Chem. 1979, 71, 4451.
Schmelzeisen-Redeker, G.; Buttering, L.; Rölgen, F. W. Int. J. Mass Spectrom. Ion Processes 1989, 90, 139.
Wong, S. F.; Meng, C. K.; Fenn, J. B. J. Phys, Chern. 1988, 92, 546.
Smith, R. D.; Barniaga, C. J.; Udseth, H. R. Anal. Chem. 1988, 60, 1948 -52.
Chowdhury, S. K.; Katta, V.; Chait, B. T. Rapid Commun. Mass Spectrom. 1990, 4, 81.
LeBlanc, J.C. Y.; Beauchemin, D.; Siu, K. W. M.; Guerzernont, R.; Berman, S. S. Org. Mass Spectrom. 1991, 26, 831.
Mirza, U. A.; Cohen, S. L.; Chait, B. T. Anal. Chem. 1993, 65, 1.
Lee, E. D.; Henion, J. D. Rapid Commun. Mass Spectrom. 1992, 6, 727.
Reid, N.; Buckley, J. A.; French, J. B.; Poon, C. C. Adv. Mass Spectrom. 1979, 8B, 1843.
Ikonomou, M. G.; Blades, A. T.; Kebarle, P. Anal Chem. 1990, 62, 957.
Ikonornou, M. G.; Blades, A. T.; Kebarle, P. Anal Chem. 1991, 63, 1989.