Abstract
A “micro-electrospray” ionization source has been developed that markedly increases the sensitivity of the conventional electrospray source. This was achieved by optimization of the source to accommodate nanoliter flow rates from 300 to 800-nL/min spraying directly from a capillary needle that, for the analysis of peptides, contained C18 liquid chromatography packing as an integrated concentration-desalting device. Thus, a total of 1 fmol of methionine enkephalin was desorbed from the capillary column spray needle, loaded as a 10-μL injection of 100-amol/μL solution. The mass spectrum showed the [M + H]+ ion at m/z 574.2 with a signal-to-noise ratio of better than 5:1 from a chromatographic peak with a width of about 12 s. A narrow range (15-u) tandem mass spectrum was obtained for methionine enkephalin from the injection of 500 amol, and a full-scan tandem-mass spectrum was obtained from 50 fmol. For proteins, the average mass measurement accuracy was approximately 100–200 ppm for the injection of 2.5 fmol of apomyoglobin and 20–40 ppm for 200 fmol. Carbonic anhydrase B and bovine serum albumin showed similar mass measurement accuracies.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Fenn, J. B.; Mann, M.; Meng, C. K.; Wong, S. F.; Whitehouse, C. M. Mass Spectrom. Rev. 1990, 9, 37–70.
Mann, M. Org. Mass Spectrom. 1990, 25, 575–587.
Moon, D.-C.; Kelly, J. A. Biomed. Mass Spectrom. 1988, 17, 229–237.
Desiderio, D. M. Life Sci. 1992, 51, 169–176.
Dass, C.; Kusmierz, J. J.; Desiderio, D. M.; Jarvis, S. A.; Green, B. N. J. Am. Soc. Mass Spectrom. 1991, 2, 149–156.
Kusmierz, J. J.; Sumrada, R.; Desiderio, D. M. Anal. Chem. 1990, 62, 2395–2400.
Tomer, K. B.; Perkins, J. R.; Parker, C. E.; Deterding, L. J. Biomed. Mass Spectrom. 1991, 20, 783–788.
Wahl, J. H.; Goodlett, D. R.; Udseth, H. R.; Smith, R. D. Anal. Chem. 1992, 64, 3194–3196.
Hunt, D. F.; Henderson, R. A.; Shabanowitz, J.; Sakaguchi, K.; Michel, H.; Sevilir, N.; Cox, A. L.; Appella, E.; Engelhard, V. H. Science 1992, 255, 1261–1263.
Wahl, J. H.; Goodlett, D. R.; Udseth, H. R.; Smith, R. D. Electrophoresis 1993, 14, 448–457.
Gale, D. C.; Smith, R. D. Rapid Commun. Mass Spectrom. 1993, 7, 1017–1021.
12. Van Berkel, G. J.; McLuckey, S. A.; Glish, G. L. Anal. Chem. 1992, 64, 1586–1593.
Emmett, M. R.; Caprioli, R. M. Proceedings of the 41st ASMS Conference on Mass Spectrometry and Allied Topics; San Francisco, CA, May 30–June 4, 1993; Abstract 47.
Austell, T. L.; Jorgenson, J. W. Abstracts of the 5th Annual HPCE Meeting; Orlando, FL, 1993; Abstract M106.
Austell, T. L.; Jorgenson, J. W. Abstracts of the 6th Annual HPCE Meeting; San Diego, CA, 1994; Abstract P308.
Allen, M. H.; Vestal, M. L. J. Am. Soc. Mass Spectrom. 1992, 3, 18–26.
Allen, M. H.; Hutchens, T. W. Rapid Commun. Mass Spectrom. 1992, 6, 308–312.
Kennedy, R. T.; Jorgenson, J. W. Anal. Chem. 1989, 61, 1128–1135.
Mosely, M. A.; Deterding, L. J.; Tomer, K. B.; Jorgenson, J. W. Anal. Chem. 1991, 63, 1467–1473.
Roepstorff, P.; Pohlmann, J. Biomed. Mass Spectrom. 1984, 1, 601.
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Emmett, M.R., Caprioli, R.M. Micro-electrospray mass spectrometry: Ultra-high-sensitivity analysis of peptides and proteins. J Am Soc Mass Spectrom 5, 605–613 (1994). https://doi.org/10.1016/1044-0305(94)85001-1
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1016/1044-0305(94)85001-1