Abstract
By using a modified ion trap mass spectrometer, resolution in excess of 30,000 (FWHM) at m I z 502 is demonstrated. The method of increasing resolution in the ion trap mass spectrometer operated in the mass-selective instability mode depends on decreasing the rate of scanning the primary radio frequency amplitude as well as using resonance ejection at the appropriate frequency and amplitude. A theoretical basis for the method is introduced.
Article PDF
Similar content being viewed by others
Avoid common mistakes on your manuscript.
References
Stafford, C. C., Jr.; Kelley, P. E.; Syka, J. E. P.; Reynolds, W. E.; Todd, J. F. J. Int. J. Mass Spectrom. Ion Proc. 1984, 60, 85.
March, R. E.; Hughes, R. J. Quadrupole Storage Mass Spectrometry; Wiley: New York, 1989.
Brodbelt, J.; Louris, J. N.; Cooks, R. G. Anal. Chem. 1981, 186, 219.
Brodbelt, J. S.; Cooks, R. G. Anal. Chem. Acta 1988, 206, 239.
Louris, J. N.; Amy, J. W.; Ridley, T. Y.; Cooks, R. G., Int. J. Mass Spec. Ion Proc. 1989, 88, 97.
Kaiser, R. E., Jr.; Louris, J. N.; Amy, J. W.; Cooks, R. G. Rapid Commun. Mass Spectrom. 1989, 3, 225.
Kaiser, R. E., Jr.: Cooks, R. G.; Syka, J. E. P.; Stafford, G. C., Jr. Rapid Commun. Mass Spectrom. 1990, 4, 30.
Van Berkel, C. J.; Clish, G. L.; McLuckey, S. A. Anal. Chem. 1990, 62, 1284.
Jardine, I.; Hail, M. E.; Lewis, S.; Zhou, J.; Schwartz, J. C; Mylchreest, I. C. Presented at the 38th ASMS Conference on Mass Spectrometry and Allied Topics, Tucson, AZ, June 3–8, 1990.
Louris, J. N.; Brodbelt-Lustig, J. S.; Cooks, R. G.; Glish, G. L.; Van Berkel, G. J.; McLuckey, S. A. Int. J. Mass Spec. Ion. Proc. 1990, 96, 117.
Johnson, J. V.; Pedder, R. E; Yost, R. A. Int. J. Mass Spectrom, Ion Proc., in press.
Kaiser, R. E., Jr.; Cooks, R. C.; Stafford, C. C., Syka, J. E. P.; Hemberger, P. H. Int. J. Mass Spectrom. Ion Proc., in press.
Schoen, A. E; Syka, J. E. P. Presented at the 36th ASMS Conference on Mass Spectrometry and Allied Topics, San Francisco, CA, June 5–10, 1988; p. 843.
Whitehouse, C. M.; Dreyer, R. N.; Yamashita, M.; Fenn, J. B. Anal. Chem. 1985, 57, 675.
Fischer, E. Z. Phys. 1959, 156, 1.
Kaiser, R. E., Jr.; Cooks, R. G.; Moss, J.; Hemberger, P. H. Rapid Commun. Mass Spec. 1989, 3, 50.