Skip to main content
Log in

Preclinical Bench Testing on a Novel Posterior Dynamic Deformity Correction Device for Scoliosis

  • Published:
Spine Deformity Aims and scope Submit manuscript

Abstract

Study Design

Biomechanical test.

Objective

To summarize the preclinical tests performed to assess the durability of a novel fusionless dynamic device for the treatment of adolescent idiopathic scoliosis (AIS).

Summary of Background Data

The minimal invasive deformity correction (MID-C) system is a distractible posterior dynamic deformity correction device designed to reduce scoliosis for AIS patients, to maintain curve correction, and to preserve spinal motion. To overcome the challenges of wear and fatigue of this procedure, the system has two unique features: polyaxial joints at the rod-screw interface and a ceramic coating of the moving parts.

Methods

Five biomechanical tests were performed: Static compression to failure, fatigue loading per ASTM F 1717 with 5.5-mm screws for 10 million cycles (MC) at 5 Hz, wear assessment, wear test of the polyaxial joint under 100 N load for 10 MC, and wear particle implantation in rabbits.

Results

The system failed through buckling of the rod with loads over 3000 N (400% of human body weight). Dynamically, the system maintained 700 N for 10 MC with 5.5 mm screws. The maximum total steady-state wear rate was 0.074 mg/MC (0.03 per polyaxial joint and 0.014 mg/MC for the ratchet mechanism). Histologic evaluation of the particle injection sites indicated no difference in the local tissue response between the control and test articles. At 3 and 6 months postinjection, there were neither adverse local effects nor systemic effects observed.

Conclusions

The unique design features of the MID-C system, based on polyaxial joints and ceramic coating, resulted in favorable static, fatigue, and wear resistance properties. Wear properties were superior to those published for artificial spinal discs. Long-term outcomes from clinical use will be required to correlate these bench tests to the in vivo reality of clinical use.

Level of Evidence

Level V.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Connolly PJ, Von Schroeder HP, Johnson GE, Kostuik JP. Adolescent idiopathic scoliosis. Long-term effect of instrumentation extending to the lumbar spine. J Bone Joint Surg Am. 1995;77:1210–6.

    Article  CAS  Google Scholar 

  2. Wilk B, Karol LA, Johnston CE, et al. The effect of scoliosis fusion on spinal motion: a comparison of fused and nonfused patients with idiopathic scoliosis. Spine (Phila Pa 1976). 2006;31:309–14.

    Article  Google Scholar 

  3. Cook S, Asher M, Lai SM, Shobe J. Reoperation after primary posterior instrumentation and fusion for idiopathic scoliosis. Toward defining late operative site pain of unknown cause. Spine (Phila Pa 1976). 2000;25:463–8.

    Article  CAS  Google Scholar 

  4. Luhmann SJ, Lenke LG, Bridwell KH, Schootman M. Revision surgery after primary spine fusion for idiopathic scoliosis. Spine (Phila Pa 1976). 2009;34:2191–7.

    Article  Google Scholar 

  5. Erwin WD, Dickson JH, Harrington PR. Clinical review of patients with broken Harrington rods. J Bone Joint Surg Am. 1980;62:1302–7.

    Article  CAS  Google Scholar 

  6. Hallab NJ. A review of the biologic effects of spine implant debris: fact from fiction. SAS J. 2009;3:143–60.

    Article  Google Scholar 

  7. Nachemson A, Elfström G. Intravital wireless telemetry of axial forces in Harrington distraction rods in patients with idiopathic scoliosis. J Bone Joint Surg Am. 1971;53:445–65.

    Article  CAS  Google Scholar 

  8. Duke H, Hill DL, M M, et al. Research Deformities 2. In: Stokes IAF, editor. Series studies in Health Technology and Informatics, 109–12. Amsterdam: IOS Press; 1999.

    Google Scholar 

  9. Shah S, Garrido E, Farooq N, Stewart Tucker HN. In Vivo Distruction Force and Length Measurements of Growing Rods: Which Factors Influence On The Ability To Lengthen? In International Congress on Early Onset Scoliosis and Growing Spine 20–21 (Journal of Child Orthopedics), 2009.

  10. Rohlmann A, Arntz U, Graichen F, Bergmann G. Loads on an internal spinal fixation device during sitting. J Biomech. 2001;34:989–93.

    Article  CAS  Google Scholar 

  11. Rohlmann A, Graichen F, Weber U, Bergmann G. Volvo Award winner in biomechanical studies: monitoring in vivo implant loads with a telemeterized internal spinal fixation device. Spine (Phila Pa 1976). 2000;25:2981–6.

    Article  CAS  Google Scholar 

  12. Lorenz M, Patwardhan A, Vanderby R. Load-bearing characteristics of lumbar facets in normal and surgically altered spinal segments. Spine. 1983;8:122–30.

    Article  CAS  Google Scholar 

  13. Appelt K. Biomechanics of the thoracic spine, development of a method to measure the influence of the rib cage in thoracic spine movement. Doctoral dissertation. Ulm, Germany: University of Ulm; 2012.

    Google Scholar 

  14. Yao X, Blount TJ, Suzuki N, et al. A biomechanical study on the effects of rib head release on thoracic spinal motion. Eur Spine J. 2012;21:606–12.

    Article  Google Scholar 

  15. Edward Stanford R, Herman Loefler A, Mark Stanford P, Walsh WR. Multiaxial pedicle screw designs: static and dynamic mechanical testing. Spine (Phila Pa 1976). 2004;29:367–75.

    Article  Google Scholar 

  16. Ellipse Technologies Inc. 510(k) Summary MAGEC: Spinal Bracing and Distraction System. Premarket Notification Number: K(140178). 2014.

  17. Lachiewicz PF, Heckman DS, Bsn ESS, et al. Femoral head size and wear of highly cross-linked polyethylene at 5 to 8 years. Clin Orthop Relat Res. 2009;467:3290–6.

    Article  Google Scholar 

  18. Joyce TJ, Smith SL, Rushton PRP, et al. Analysis of explanted magnetically controlled growing rods from seven UK spinal centers. Spine (Phila Pa 1976). 2018;43:E16–22.

    Article  Google Scholar 

  19. Holewijn RM, de Kleuver M, van der Veen AJ, et al. A novel spinal implant for fusionless scoliosis correction: a biomechanical analysis of the motion preserving properties of a posterior periapical concave distraction device. Global Spine J. 2017;7:400–9.

    Article  Google Scholar 

  20. Appelt K. Biomechanics of the thoracic spine development of a method to measure the influence of the rib cage on thoracic spine movement. Medizinischen Fakultät. Ulm, Germany: Universitát Ulm; 2012.

    Google Scholar 

  21. Floman Y, Gavriliu S, Potaczek T, et al. A new posterior dynamic device for correction of moderate adolescent idiopathic scoliosis: 27 cases with over two years of follow-up. In: Paper presented at: International Meeting of Advanced Spine Techniques, Los Angeles, California, July 11-15 2018.

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to R. El-Hawary MD.

Additional information

Author disclosures: UA (personal fees from ApiFix Ltd, during the conduct of the study; in addition, UA has a patent PCT/US11/035278 licensed, and a patent PCT/US13/20453 licensed), RE-H (personal fees from Apifix Ltd., during the conduct of the study; grants from DePuy Synthes Spine and Medtronic Spine; personal fees from DePuy Synthes Spine, Medtronic Spine, and Apifix Ltd.; nonfinancial support from Children’s Spine Foundation and Pediatric Orthopaedic Society of North Americal; grants from Canadian Institute of Health Research, Atlantic Canada Opportunities Agency, Scoliosis Research Society, Pediatric Orthopaedic Society of North America, and EOS Imaging; nonfinancial support from Scoliosis Research Society, outside the submitted work), RRB (personal fees and other from ApiFix, during the conduct of the study; personal fees and other from Abyrx, DePuy Synthes Spine, and SpineGuard; other from Advanced Vertebral Solutions, Electrocore, Medovex, MiMedx, Orthobond, and Lifeunit; personal fees from Globus Medical, Medtronic, MacKeith Publishers, and Wishbone, outside the submitted work; and Son Randal Jr. is an employee of DePuy Synthes Spine), BSL (grants from Setting Scoliosis Straight Foundation; personal fees from DePuy Synthes Spine, K2M, Paradigm Spine, Spine Search, and Ethicon; nonfinancial support from Spine Deformity Journal; grants from John and Marcella Fox Fund Grant, OREF; personal fees from Zimmer Biomet, Apifix, and Unyq Align, outside the submitted work), YF (reports personal fees from Apifix, during the conduct of the study).

IRB Approval: Not required as this was a biomechanical testing study.

Funding Sources: This study was funded by Apifix Ltd., Israel.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Arnin, U., El-Hawary, R., Betz, R. et al. Preclinical Bench Testing on a Novel Posterior Dynamic Deformity Correction Device for Scoliosis. Spine Deform 7, 203–212 (2019). https://doi.org/10.1016/j.jspd.2018.08.010

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1016/j.jspd.2018.08.010

Keywords

Navigation