High mass selectivity for top-down proteomics by application of SWIFT technology

Article

DOI: 10.1016/j.jasms.2009.11.011

Cite this article as:
Guan, S. & Burlingame, A.L. J Am Soc Mass Spectrom (2010) 21: 455. doi:10.1016/j.jasms.2009.11.011

Abstract

Stored waveform inverse Fourier transform (SWIFT) technology has been implemented on a commercial Fourier transform ICR mass spectrometer. Complex ejection/isolation waveforms can be generated on an arbitrary waveform generator (AWG) and applied to the instrument by use of a high speed analog switch. High mass selectivity and subsequent electron capture dissociation (ECD) of the SWIFT isolated ions has been demonstrated with analysis of intact Bovine histone H4. A mass selectivity about 0.1 m/z unit for isolation of the 18+ charge state ions was achieved. Adaptation of SWIFT on the commercial instrument provides an enhanced capability for characterizing intact proteins by top-down analysis.

Copyright information

© American Society for Mass Spectrometry 2010