Sub part-per-million mass accuracy by using stepwise-external calibration in fourier transform ion cyclotron resonance mass spectrometry

Articles

Abstract

A new external calibration procedure for FT-ICR mass spectrometry is presented, stepwise-external calibration. This method is demonstrated for MALDI analysis of peptide mixtures, but is applicable to any ionization method. For this procedure, the masses of analyte peaks are first accurately measured at a low trapping potential (0.63 V) using external calibration. These accurately determined (<1 ppm accuracy) analyte peaks are used as internal calibrant points for a second mass spectrum that is acquired for the same sample at a higher trapping potential (1.0 V). The second mass spectrum has a ∼10-fold improvement in detection dynamic range compared with the first spectrum acquired at a low trapping potential. A calibration equation that accounts for local and global space charge is shown to provide mass accuracy with external calibration that is nearly identical to that of internal calibration, without the drawbacks of experimental complexity or reduction of abundance dynamic range. For the 609 mass peaks measured using stepwise-external calibration method, the root-mean-square error is 0.9 ppm. The errors appear to have a Gaussian distribution; 99.3% of the mass errors are shown to lie within three times the sample standard deviation (2.6 ppm) of their true value.

References

  1. 1.
    Yamashita, M.; Fenn, J. B. Electrospray Ion Source—Another Variation on the Free-Jet Theme. J. Phys. Chem. 1984, 88, 4451–4459.CrossRefGoogle Scholar
  2. 2.
    Yamashita, M.; Fenn, J. B. Negative-Ion Production with the Electrospray Ion Source. J. Phys. Chem. 1984, 88, 4671–4679.CrossRefGoogle Scholar
  3. 3.
    Fenn, J. B.; Mann, M.; Meng, C. K.; Wong, S. F.; Whitehouse, C. M. Electrospray Ionization for Mass Spectrometry of Large Biomolecules. Science 1989, 246, 64–71.CrossRefGoogle Scholar
  4. 4.
    Karas, M.; Hillenkamp, F. Laser Desorption Ionization of Proteins with Molecular Masses Exceeding 10,000 Daltons. Anal. Chem. 1988, 60, 2299–2301.CrossRefGoogle Scholar
  5. 5.
    Tanaka, K.; Waki, H.; Ido, Y.; Akita, S.; Yoshida, Y.; T, Y. Protein and Polymer Analyses up to m/z 100,000 by Laser Ionization Time-of-Flight Mass Spectrometry. Rapid Commun. Mass Spectrom. 1988, 2, 151–153.CrossRefGoogle Scholar
  6. 6.
    Hillenkamp, F.; Karas, M.; Beavis, R. C.; Chait, B. T. Matrix-Assisted Laser Desorption Ionization Mass-Spectrometry of Biopolymers. Anal. Chem. 1991, 63, A1193-A1202.CrossRefGoogle Scholar
  7. 7.
    Ge, Y.; Lawhorn, B. G.; El-Naggar, M.; Strauss, E.; Park, J. H.; Begley, T. P.; McLafferty, F. W. Top Down Characterization of Larger Proteins (45 kDa) by Electron Capture Dissociation Mass Spectrometry. J. Am. Chem. Soc. 2002, 124, 672–678.CrossRefGoogle Scholar
  8. 8.
    Meng, F. Y.; Du, Y.; Miller, L. M.; Patrie, S. M.; Robinson, D. E.; Kelleher, N. L. Molecular-Level Description of Proteins from Saccharomyces cerevisiae Using Quadrupole FT Hybrid Mass Spectrometry for Top- Down Proteomics. Anal. Chem. 2004, 76, 282–2858.CrossRefGoogle Scholar
  9. 9.
    Kelleher, N. L. Top-Down Proteomics. Anal. Chem. 2004, 76, 196A-203A.CrossRefGoogle Scholar
  10. 10.
    Yates, J. R.; McCormack, A. L.; Eng, J. Mining Genomes with MS. Anal. Chem. 1996, 68, A534-A540.CrossRefGoogle Scholar
  11. 11.
    Gygi, S. P.; Rist, B.; Gerber, S. A.; Turecek, F.; Gelb, M. H.; Aebersold, R. Quantitative Analysis of Complex Protein Mixtures Using Isotope-Coded Affinity Tags. Nat. Biotechnol. 1999, 17, 994–999.CrossRefGoogle Scholar
  12. 12.
    Gao, H. Y.; Shen, Y. F.; Veenstra, T. D.; Harkewicz, R.; Anderson, G. A.; Bruce, J. E.; Pasa-Tolic, L.; Smith, R. D. Two-Dimensional Electrophoretic/Chromatographic Separations Combined with Electrospray Ionization FTICR Mass Spectrometry for High Throughput Proteome Analysis. J. Microcolumn Sep. 2000, 12, 383–390.CrossRefGoogle Scholar
  13. 13.
    Brock, A.; Horn, D. M.; Peters, E. C.; Shaw, C. M.; Ericson, C.; Phung, Q. T.; Salomon, A. R. An Automated Matrix-Assisted Laser Desorption/Ionization Quadrupole Fourier Transform Ion Cyclotron Resonance Mass Spectrometer for “Bottom-Up” Proteomics. Anal. Chem. 2003, 75, 3419–3428.CrossRefGoogle Scholar
  14. 14.
    Wolters, D. A.; Washburn, M. P.; Yates, J. R. An Automated Multidimensional Protein Identification Technology for Shotgun Proteomics. Anal. Chem. 2001, 73, 5683–5690.CrossRefGoogle Scholar
  15. 15.
    Smith, R. D.; Anderson, G. A.; Lipton, M. S.; Pasa-Tolic, L.; Shen, Y. F.; Conrads, T. P.; Veenstra, T. D.; Udseth, H. R. An Accurate Mass Tag Strategy for Quantitative and High-Throughput Proteome Measurements. Proteomics 2002, 2, 513–523.CrossRefGoogle Scholar
  16. 16.
    McDonald, W. H.; Yates, J. R. Shotgun Proteomics and Biomarker Discovery. Disease Markers 2002, 18, 99–105.CrossRefGoogle Scholar
  17. 17.
    Washburn, M. P.; Ulaszek, R. R.; Yates, J. R. Reproducibility of Quantitative Proteomic Analyses of Complex Biological Mixtures by Multidimensional Protein Identification Technology. Anal. Chem. 2003, 75, 5054–5061.CrossRefGoogle Scholar
  18. 18.
    Dieguez-Acuna, F. J.; Gerber, S. A.; Kodama, S.; Elias, J. E.; Beausoleil, S. A.; Faustman, D.; Gygi, S. P. Characterization of Mouse Spleen Cells by Subtractive Proteomics. Mol. Cell. Proteom. 2005, 4, 1459–1470.CrossRefGoogle Scholar
  19. 19.
    Wong, R. L.; Amster, I. J. Combining Low and High Mass Ion Accumulation for Enhancing Shotgun Proteome Analysis by Accurate Mass Measurement. J. Am. Soc. Mass Spectrom. 2006, 17, 205–212.CrossRefGoogle Scholar
  20. 20.
    Henzel, W. J.; Billeci, T. M.; Stults, J. T.; Wong, S. C.; Grimley, C.; Watanabe, C. Identifying Proteins from 2-Dimensional Gels by Molecular Mass Searching of Peptide-Fragments in Protein-Sequence Databases. Proc. Natl. Acad. Sci. U.S.A. 1993, 90, 5011–5015.CrossRefGoogle Scholar
  21. 21.
    Henzel, W. J.; Watanabe, C.; Stults, J. T. Protein Identification: The Origins of Peptide Mass Fingerprinting. J. Am. Soc. Mass Spectrom. 2003, 14, 931–942.CrossRefGoogle Scholar
  22. 22.
    Futcher, B.; Latter, G. I.; Monardo, P.; McLaughlin, C. S.; Garrels, J. I. A Sampling of the Yeast Proteome. Mol. Cell. Biol. 1999, 19, 7357–7368.CrossRefGoogle Scholar
  23. 23.
    Krause, E.; Wenschuh, H.; Jungblut, P. R. The Dominance of Arginine-Containing Peptides in MALDI-Derived Tryptic Mass Fingerprints of Proteins. Anal. Chem. 1999, 71, 4160–4165.CrossRefGoogle Scholar
  24. 24.
    Comisarow, M. B.; Marshall, A. G. Fourier-Transform Ion-Cyclotron Resonance Spectroscopy. Chem. Phys. Lett. 1974, 25, 282–283.CrossRefGoogle Scholar
  25. 25.
    Comisarow, M. B.; Marshall, A. G. Frequency-Sweep Fourier-Transform Ion-Cyclotron Resonance Spectroscopy. Chem. Phys. Lett. 1974, 26, 489–490.CrossRefGoogle Scholar
  26. 26.
    Gorshkov, M. V.; Guan, S. H.; Marshall, A. G. Masses of Stable Neon Isotopes Determined at Parts-Per-Billion Precision by Fourier-Transform Ion-Cyclotron Resonance Mass-Spectrometry. Int. J. Mass Spectrom. Ion Processes 1993, 128, 47–60.CrossRefGoogle Scholar
  27. 27.
    Rodgers, R. P.; White, F. M.; Hendrickson, C. L.; Marshall, A. G.; Andersen, K. V. Resolution, Elemental Composition, and Simultaneous Monitoring by Fourier Transform Ion Cyclotron Resonance Mass Spectrometry of Organosulfur Species Before and After Diesel Fuel Processing. Anal. Chem. 1998, 70, 4743–4750.CrossRefGoogle Scholar
  28. 28.
    He, F.; Hendrickson, C. L.; Marshall, A. G. Baseline Mass Resolution of Peptide Isobars: A Record for Molecular Mass Resolution. Anal. Chem. 2001, 73, 647–650.CrossRefGoogle Scholar
  29. 29.
    Clauser, K. R.; Baker, P.; Burlingame, A. L. Role of Accurate Mass Measurement (±10 ppm) in Protein Identification Strategies Employing MS or MS/MS and Database Searching. Anal. Chem. 1999, 71, 2871–2882.CrossRefGoogle Scholar
  30. 30.
    Conrads, T. P.; Anderson, G. A.; Veenstra, T. D.; Pasa-Tolic, L.; Smith, R. D. Utility of Accurate Mass Tags for Proteome-Wide Protein Identification. Anal. Chem. 2000, 72, 3349–3354.CrossRefGoogle Scholar
  31. 31.
    Burton, R. D.; Matuszak, K. P.; Watson, C. H.; Eyler, J. R. Exact Mass Measurements Using a 7 Tesla Fourier Transform Ion Cyclotron Resonance Mass Spectrometer in a Good Laboratory Practices-Regulated Environment. J. Am. Soc. Mass Spectrom. 1999, 10, 1291–1297.CrossRefGoogle Scholar
  32. 32.
    Jeffries, J. B.; Barlow, S. E.; Dunn, G. H. Theory of Space-Charge Shift of Ion-Cyclotron Resonance Frequencies. Int. J. Mass Spectrom. Ion Processes 1983, 54, 169–187.CrossRefGoogle Scholar
  33. 33.
    Francl, T. J.; Sherman, M. G.; Hunter, R. L.; Locke, M. J.; Bowers, W. D.; McIver, R. T. Experimental Determination of the Effects of Space Charge on Ion Cyclotron Resonance Frequencies. Int. J. Mass Spectrom. Ion Processes 1983, 54, 189–199.CrossRefGoogle Scholar
  34. 34.
    Ledford, E. B.; Rempel, D. L.; Gross, M. L. Space-Charge Effects in Fourier-Transform Mass-Spectrometry—Mass Calibration. Anal. Chem. 1984, 56, 2744–2748.CrossRefGoogle Scholar
  35. 35.
    Chen, S. P.; Comisarow, M. B. Simple Physical Models for Coulomb-Induced Frequency-Shifts and Coulomb-Induced Inhomogeneous Broadening for Like and Unlike Ions in Fourier-Transform Ion-Cyclotron Resonance Mass Spectrometry. Rapid Commun. Mass Spectrom. 1991, 5, 450–455.CrossRefGoogle Scholar
  36. 36.
    Chen, S. P.; Comisarow, M. B. Modeling Coulomb Effects in Fourier-Transform Ion-Cyclotron Resonance Mass-Spectrometry by Charged Disks and Charged Cylinders. Rapid Commun. Mass Spectrom. 1992, 6, 1–3.CrossRefGoogle Scholar
  37. 37.
    Easterling, M. L.; Mize, T. H.; Amster, I. J. Routine Part-Per-Million Mass Accuracy for High-Mass Ions: Space-Charge Effects in MALDI FT-ICR. Anal. Chem. 1999, 71, 624–632.CrossRefGoogle Scholar
  38. 38.
    Mize, T. H.; Amster, I. J. Broad-Band Ion Accumulation with an Internal Source MALDI-FTICR-MS. Anal. Chem. 2000, 72, 5886–5891.CrossRefGoogle Scholar
  39. 39.
    Taylor, P. K.; Amster, I. J. Space Charge Effects on Mass Accuracy for Multiply Charged Ions in ESI-FTICR. Int. J. Mass Spectrom. 2003, 222, 351–361.CrossRefGoogle Scholar
  40. 40.
    Shen, Y. F.; Zhao, R.; Belov, M. E.; Conrads, T. P.; Anderson, G. A.; Tang, K. Q.; Pasa-Tolic, L.; Veenstra, T. D.; Lipton, M. S.; Udseth, H. R.; Smith, R. D. Packed Capillary Reversed-Phase Liquid Chromatography with High-Performance Electrospray Ionization Fourier Transform Ion Cyclotron Resonance Mass Spectrometry for Proteomics. Anal. Chem. 2001, 73, 1766–1775.CrossRefGoogle Scholar
  41. 41.
    Conrads, T. P.; Alving, K.; Veenstra, T. D.; Belov, M. E.; Anderson, G. A.; Anderson, D. J.; Lipton, M. S.; Pasa-Tolic, L.; Udseth, H. R.; Chrisler, W. B.; Thrall, B. D.; Smith, R. D. Quantitative Analysis of Bacterial and Mammalian Proteomes Using a Combination of Cysteine Affinity Tags and N-15-Metabolic Labeling. Anal. Chem. 2001, 73, 2132–2139.CrossRefGoogle Scholar
  42. 42.
    Shen, Y. F.; Tolic, N.; Zhao, R.; Pasa-Tolic, L.; Li, L. J.; Berger, S. J.; Harkewicz, R.; Anderson, G. A.; Belov, M. E.; Smith, R. D. High-Throughput Proteomics Using High Efficiency Multiple-Capillary Liquid Chromatography with On-Line High-Performance ESI FTICR Mass Spectrometry. Anal. Chem. 2001, 73, 3011–3021.CrossRefGoogle Scholar
  43. 43.
    Shen, Y. F.; Jacobs, J. M.; Camp, D. G.; Fang, R. H.; Moore, R. J.; Smith, R. D.; Xiao, W. Z.; Davis, R. W.; Tompkins, R. G. Ultra-High-Efficiency Strong Cation Exchange LC/RPLC/MS/MS for High Dynamic Range Characterization of the Human Plasma Proteome. Anal. Chem. 2004, 76, 1134–1144.CrossRefGoogle Scholar
  44. 44.
    Masselon, C.; Tolmachev, A. V.; Anderson, G. A.; Harkewicz, R.; Smith, R. D. Mass Measurement Errors Caused by “Local” Frequency Perturbations in FTICR Mass Spectrometry. J. Am. Soc. Mass Spectrom. 2002, 13, 99–106.CrossRefGoogle Scholar
  45. 45.
    Hannis, J. C.; Muddiman, D. C. A Dual Electrospray Ionization Source Combined with Hexapole Accumulation to Achieve High Mass Accuracy of Biopolymers in Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. J. Am. Soc. Mass Spectrom. 2000, 11, 876–883.CrossRefGoogle Scholar
  46. 46.
    Belov, M. E.; Zhang, R.; Strittmatter, E. F.; Prior, D. C.; Tang, K.; Smith, R. D. Automated Gain Control and Internal Calibration with External Ion Accumulation Capillary Liquid Chromatography-Electrospray Ionization-Fourier Transform Ion Cyclotron Resonance. Anal. Chem. 2003, 75, 4195–4205.CrossRefGoogle Scholar
  47. 47.
    O’Connor, P. B.; Costello, C. E. Internal Calibration on Adjacent Samples (InCAS) with Fourier Transform Mass Spectrometry. Anal. Chem. 2000, 72, 5881–5885.CrossRefGoogle Scholar
  48. 48.
    Muddiman, D. C.; Oberg, A. L. Statistical Evaluation of Internal and External Mass Calibration Laws Utilized in Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. Anal. Chem. 2005, 77, 2406–2414.CrossRefGoogle Scholar
  49. 49.
    Syka, J. E. P.; Marto, J. A.; Bai, D. L.; Horning, S.; Senko, M. W.; Schwartz, J. C.; Ueberheide, B.; Garcia, B.; Busby, S.; Muratore, T.; Shabanowitz, J.; Hunt, D. F. Novel Linear Quadrupole Ion Trap/FT Mass Spectrometer: Performance Characterization and Use in the Comparative Analysis of Histone H3 Post-Translational Modifications. J. Proteome Res. 2004, 3, 621–626.CrossRefGoogle Scholar
  50. 50.
    Peterman, S. M.; Mulholland, J. J. A Novel Approach for Identification and Characterization of Glycoproteins Using a Hybrid Linear Ion Trap/FT-ICR Mass Spectrometer. J. Am. Soc. Mass Spectrom. 2006, 17, 168–179.CrossRefGoogle Scholar
  51. 51.
    Page, J. S.; Bogdanov, B.; Vilkov, A. N.; Prior, D. C.; Buschbach, M. A.; Tang, K.; Smith, R. D. Automatic Gain Control in Mass Spectrometry Using a Jet Disrupter Electrode in an Electrodynamic Ion Funnel. J. Am. Soc. Mass Spectrom. 2005, 16, 244–253.CrossRefGoogle Scholar
  52. 52.
    Ramanathan, R. Introductory Econometrics with Applications; 3rd ed.; Harcourt Brace and Co.: Fort Worth, TX, 1995, p 23.Google Scholar
  53. 53.
    Shi, S. D. H.; Drader, J. J.; Freitas, M. A.; Hendrickson, C. L.; Marshall, A. G. Comparison and Interconversion of the Two Most Common Frequency-to-Mass Calibration Functions for Fourier Transform Ion Cyclotron Resonance Mass Spectrometry. Int. J. Mass Spectrom. 2000, 196, 591–598.CrossRefGoogle Scholar
  54. 54.
    Gorshkov, M. V.; Marshall, A. G.; Nikolaev, E. N. Analysis and Elimination of Systematic-Errors Originating from Coulomb Mutual Interaction and Image Charge in Fourier-Transform Ion-Cyclotron Resonance Precise Mass Difference Measurements. J. Am. Soc. Mass Spectrom. 1993, 4, 855–868.CrossRefGoogle Scholar

Copyright information

© American Society for Mass Spectrometry 2006

Authors and Affiliations

  1. 1.Department of ChemistryUniversity of GeorgiaAthensUSA

Personalised recommendations