Development and optimization of a receive-only surface array with purely geometrical decoupling for rat brain MRI at 2 T

  • Original Article
  • Published:
Research on Biomedical Engineering Aims and scope Submit manuscript

Abstract

Introduction

Magnetic resonance imaging (MRI) experiments on small animals, as well as in humans, require specific radiofrequency (RF) coil designs in order to maximize field homogeneity during excitation and signal-to-noise ratio (SNR) during reception. Therefore, the use of dedicated transmit-only and receive-only coil geometries has been preferred as standard technology configuration. Among the receive coil geometries, the surface coil has been widely used to increase the SNR in localized regions of interest. With the advent of MRI-phased arrays, the high sensitivity of surface coils can be achieved in extended regions, further allowing parallel imaging acquisition methodologies to accelerate acquisition time.

Objectives

This work describes the development and characterization of a two-channel receive-only RF coil set specifically designed for MRI acquisition of rat brain in a 2 T system for neuroscience studies.

Methods

We have developed and compared the performance of one common surface coil with a two-channel phased array with purely geometric decoupling between coil elements (i.e., without using low input impedance preamplifiers), both actively detuned during reception and having exactly the same dimensions.

Results

The results have shown that the single-channel surface coil achieved up to 5 times higher SNR at 5 mm depth compared to a volume coil commonly used for rat brain MRI in our laboratory, while the two-channel array achieved 4 times higher SNR at the same depth, with the additional advantage of producing an extended area with higher sensitivity.

Conclusion

In vivo brain images of anesthetized rats confirmed the SNR gain for both receive-only single-channel surface coil and two-channel array in comparison with a volume coil. We have verified that only using the partial overlapping technique does not provide sufficient isolation to eliminate mutual coupling interaction between elements, since residual coupling can still affect the coil performance.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Subscribe and save

Springer+ Basic
$34.99 /Month
  • Get 10 units per month
  • Download Article/Chapter or eBook
  • 1 Unit = 1 Article or 1 Chapter
  • Cancel anytime
Subscribe now

Buy Now

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

References

Download references

Acknowledgments

The authors are grateful to Jaqueline Moraes Malheiros for her excellent skills and help with animal handling. The authors thank the Coordination for the Improvement of Higher Education Personnel (CAPES) for support on the scholarship for D. Papoti.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Daniel Papoti.

Ethics declarations

This research was supported by Sao Paulo Research Foundation (FAPESP) grant # 2005/56663-1. All applicable national and institutional guidelines set by the Animal Care and Use of the University of Sao Paulo were followed.

Conflict of interest

The authors declare that they have no conflict of interest.

Additional information

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Papoti, D., Vidoto, E.L.G., Matins, M.J. et al. Development and optimization of a receive-only surface array with purely geometrical decoupling for rat brain MRI at 2 T. Res. Biomed. Eng. 36, 341–348 (2020). https://doi.org/10.1007/s42600-020-00074-8

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s42600-020-00074-8

Keywords

Navigation