Abstract
Introduction
Power output can be used to control training load. However, it is important to know if different power meters are reliable, because variations in their measures have direct repercussions on the training program. The aim of this study was to verify the reliability of the power output during two different pedaling protocols (incremental and variable tests) obtained with GARMIN Vector pedals and to compare it with the gold standard LODE Excalibur cycle ergometer.
Methods
Fourteen male cyclists (30 (6) years of age, maximal power 423 (23) W) performed a maximal test (initial load 100 W with 25 W/min increments) and a variable load test (100–400 W) with the LODE Excalibur cycle ergometer instrumented with GARMIN Vector pedals (coupled on the LODE Excalibur cranks). Student’s t test, intraclass correlation coefficient (ICC), and Bland-Altman analysis were used to evaluate meters’ differences and reliability, respectively.
Results
All measures performed in the incremental test (100–400 W) and two measures (100 W, 150 W) of the variable load test showed no power output measurements’ reliability between the GARMIN Vector pedals and the LODE Excalibur cycle ergometer (p < 0.05).
Conclusion
The high non-systematic errors in power output measured by the GARMIN Vector pedals (low repeatability) can generate problems when using this equipment for the training periodization in cycling.
Similar content being viewed by others
References
Abbiss CR, Quod MJ, Levin G, Martin DT, Laursen PB. Accuracy of the Velotron ergometer and SRM power meter. Int J Sports Med. 2009;30(2):107–12. https://doi.org/10.1055/s-0028-1103285.
Batterham AM, Hopkins WG. Making meaningful inferences about magnitudes. Int J Sports Physiol Perform. 2006;1(1):50–7. https://doi.org/10.1123/ijspp.1.1.50.
Bertucci W, Duc S, Villerius V, Grappe F. Validity and reliability of the Axiom PowerTrain cycle ergometer when compared with an SRM powermeter. Int J Sports Med. 2005;26(1):59–65. https://doi.org/10.1055/s-2004-817855.
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–10. https://doi.org/10.1016/S0140-6736(86)90837-8.
Bouillod A, Pinot J, Soto-Romero G, Bertucci W, Grappe F. Validity, sensitivity, reproducibility, and robustness of the PowerTap, Stages, and Garmin Vector power meters in comparison with the SRM device. Int J Sports Physiol Perform. 2017;12(8):1023–30. https://doi.org/10.1123/ijspp.2016-0436.
Coyle EF, Feltner ME, Kautz SA, Hamilton MT, Montain SJ, Baylor AM, et al. Physiological and biomechanical factors associated with elite endurance cycling performance. Med Sci Sports Exerc. 1991;23(1):93–107. https://doi.org/10.1249/00005768-199101000-00015.
Earnest CP, Wharton RP, Church TS, Lucia A. Reliability of the Lode Excalibur Sport ergometer and applicability to Computrainer electromagnetically braked cycling training device. J Strength Cond Res. 2005;19(2):344–8. https://doi.org/10.1519/R-15714.1.
Gonzalez-Alonso J, Crandall CG, Johnson JM. The cardiovascular challenge of exercising in the heat. J Physiol. 2008;586(1):45–53. https://doi.org/10.1113/jphysiol.2007.142158.
Hopker J, Myers S, Jobson SA, Bruce W, Passfield L. Validity and reliability of the Wattbike cycle ergometer. Int J Sports Med. 2000;31(10):731–6. https://doi.org/10.1055/s-0030-1261968.
Hopkins WG. Measures of reliability in sports medicine and science. Sports Med. 2000;30(1):1–15. https://doi.org/10.2165/00007256-200030010-00001.
Hopkins WG, Schabort EJ, Hawley JA. Reliability of power in physical performance tests. Sports Med. 2001;31(3):211–34. https://doi.org/10.2165/00007256-200131030-00005.
Inoue A, Sa Filho AS, Mello FC, Santos TM. Relationship between anaerobic cycling tests and mountain bike cross-country performance. J Strength Cond Res. 2012;26(6):1589–93. https://doi.org/10.1519/JSC.0b013e318234eb89.
Koo TK, Li MY. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chir Med. 2016;15(2):155–63. https://doi.org/10.1016/j.jcm.2016.02.012.
Lucas RD, Machado CEP, Souza KM, Oliveira MFM, Guglielmo LGA, Vleck V, et al. Aspectos fisiológicos do mountain biking competitivo. Braz J Sports Med. 2010;16(6):459–64. https://doi.org/10.1590/S1517-86922010000600013.
Lucia A, Pardo J, Durantez A, Hoyos J, Chicharro JL. Physiological differences between professional and elite road cyclists. Int J Sports Med. 1998;19(5):342–8. https://doi.org/10.1055/s-2007-971928.
Lucia A, Hoyos J, Santalla A, Perez M, Chicharro JL. Curvilinear VO(2):power output relationship in a ramp test in professional cyclists: possible association with blood hemoglobin concentration. Jap J Physiol. 2002;52(1):95–103. https://doi.org/10.2170/jjphysiol.52.95.
Marfell-Jones M, Olds T, Stewart A, Carter L. International standards for anthropometric assessment. South Africa: Potchefstroom; 2006.
Nimmerichter A, Schnitzer L, Prinz B, Simon D, Wirth K. Validity and reliability of the Garmin Vector power meter in laboratory and field cycling. Int J Sports Med. 2017;38(6):439–46. https://doi.org/10.1055/s-0043-101909.
Novak AR, Dascombe BJ. Agreement of power measures between Garmin Vector and SRM cycle power meters. Meas Phys Educ Exerc Sci. 2016;20(3):1–6. https://doi.org/10.1080/1091367X.2016.1191496.
Rizzo L, Thompson MW. Cardiovascular adjustments to heat stress during prolonged exercise. J Sports Med Phys Fitness. 2018;58(5):727–43. https://doi.org/10.23736/S0022-4707.17.06831-1.
Rossato M, Bini RR, Carpes FP, Diefenthaeler F, Moro AR. Cadence and workload effects on pedaling technique of well-trained cyclists. Int J Sports Med. 2008;29(9):746–52. https://doi.org/10.1055/s-2008-1038375.
Sanders D, Myers T, Akubat I. Training-intensity distribution in road cyclists: objective versus subjective measures. Int J Sports Physiol Perform. 2017;12(9):1232–7. https://doi.org/10.1123/ijspp.2016-0523.
Schwenk M, Gogulla S, Englert S, Czempik A, Hauer K. Test-retest reliability and minimal detectable change of repeated sit-to-stand analysis using one body fixed sensor in geriatric patients. Physiol Meas. 2012;33(11):1931–46. https://doi.org/10.1088/0967-3334/33/11/1931.
Sparks SA, Dove B, Bridge CA, Midgely AW, McNaughton LR. Validity and reliability of the Look Keo power pedal system for measuring power output during incremental and repeated sprint cycling. Int J Sports Physiol Perform. 2015;10(1):39–45. https://doi.org/10.1123/ijspp.2013-0317.
Stapelfeldt B, Schwirtz A, Schumacher YO, Hillebrecht M. Workload demands in mountain bike racing. Int J Sports Med. 2004;25(4):294–300. https://doi.org/10.1055/s-2004-819937.
Vogt S, Heinrich L, Schumacher YO, Blum A, Roecker K, Dickhuth HH, et al. Power output during stage racing in professional road cycling. Med Sci Sports Exerc. 2006;38(1):147–51. https://doi.org/10.1249/01.mss.0000183196.63081.6a.
Weir JP. Quantifying test-retest reliability using the intraclass correlation coefficient and the SEM. J Strength Cond Res. 2005;19(1):231–40. https://doi.org/10.1519/15184.1.
Acknowledgments
The authors wish to specially thank the athletes for their participation and cooperation in this study.
Funding
MAV is a recipient of a research grant from the Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq-Brasil).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Lanferdini, F.J., Sperb, C.H., da Rocha, E.S. et al. Power output reliability between Garmin® Vector meter and Lode Excalibur Sport® cycle ergometer. Res. Biomed. Eng. 36, 333–340 (2020). https://doi.org/10.1007/s42600-020-00073-9
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s42600-020-00073-9