Introduction

Many electrochemical scientists studied the effect of the redox current peaks of mercury ions in different medium and different solid electrodes with nanomaterials by cyclic voltammetric technique [14].

Electrochemical methods provide high potential for the investigation of antioxidant compounds, assessment of antioxidant capacity, and measurement of electrochemical index. The devices can be stationary or flow through, and based on cyclic or differential pulse voltammetry as well as potentiostatic analysis. The methods are known for their suitability for food control and monitoring the levels of antioxidant capacity in other biological samples and matrices. The application of electrochemical methods for the analysis of plant and clinical samples with respect to study of their antioxidant properties was studied by different researchers [5].

The electrochemical behavior of ascorbic acid (AA) was studied by a description of the potentiometric, voltammetric and amperometric methods for vitamin C content estimation in various media. Different methods for the development of electrochemical sensors are used, from unmodified electrodes to different composites incorporating carbon nanotubes (CNTs), ionic liquids or various mediators. The interaction between the functional groups of the sensor’s material and the analyte molecule is essential for the analytical characteristics obtained. The numerous applications of AA electrochemical sensors in fields like food, pharmaceutical or clinical analysis are a key of analyte [6].

Mercury is reported as a heavy metal that induces blood disorders and immunological effects. This study was performed to determine the haematological toxicity of mercury evaluated the protective antioxidant role of vitamin C. Studies have shown that vitamin C supplementation has varied effects on induced toxicity. AA has been found to interact with several elements in such a way as to render them less available for animals. The final results of this study of mercury and vitamin C had a protective effect on the potential harmful metals [7].

The studies are made to find therapeutic agents capable of minimizing genotoxicity of various natural and man-made compounds. The genotoxicity induced by mercury compounds remains controversial. It was found that the genotoxic effect of mercuric chloride at three concentrations (1.052, 5.262, and 10.524 μM) and AA (vitamin C) at a concentration of 9.734 μM on mercuric chloride-treated short-term human leucocyte cultures. The results showed that mercury ions have no effect on cell-cycle kinetics. Hg(II) also significantly induced C-anaphases (abnormal mitosis) in blood cultures. These effects were prevented by the addition of vitamin C to mercuric chloride-treated cultures. The data indicate that the mutagenic activity of mercuric chloride and the protective role of vitamin C on mercury-induced genotoxicity in human blood cultures is probably due to its strong antioxidant and nucleophilic nature [8].

Other study that the protective effects of vitamin C and vitamin E against mercuric chloride-induced reproductive toxicity and the mechanisms underlying these effects. The results provide further insight into the mechanisms of mercuric chloride-induced reproductive toxicity and confirm the antioxidant potential of both vitamin C and vitamin E [9].

MWCNT/TiO2 composite was modified onto glassy carbon electrode (GCE) and fabricated by mechanical attachment, then applied to detection of AA using cyclic voltammetry (CV). The electrocatalytic study of AA using single-walled CNT/tungsten oxide (SWCNT/WO3)-modified GCE prepared by mechanical attachment was conducted. The oxidation peaks obtained indicate that the SWCNT/WO3-modified GCE has good electrochemical behavior in terms of sensitivity [10, 11].

The present work involves studying of the biosensors as a good catalyst for the electro-analysis of the interference of mercuric ion with ascorbic ion and effected on the blood medium by cyclic voltammetric technique.

Materials and methods

CNT (Fluka, 98 %), blood samples were used from healthy human. Other chemicals and solvents were of annular grade and used as received from the manufacturer. Deionized water was used for the preparation of aqueous solutions. All solutions were deaerated with oxygen free nitrogen gas for 10–15 min prior to making the measurement.

Instruments

Two types of instruments were used in this work; the first one is electrochemical workstations of Bioanalytical System, Inc., USA: models BAS CV 50W with potentiostat driven by electroanalytical measuring software were connected to personal computer to perform CV. The other one of instrument is EZstat series (potentiostat/galvanostat) NuVant Systems, Inc. pioneering electrochemical technologies USA. Electrochemical workstations of Bioanalytical System, Inc. with potentiostat driven by electroanalytical measuring softwares were connected to personal computer to perform CV; the Ag/AgCl (3 M NaCl) and platinum wire (1 mm diameter) were used as a reference and counter electrode, respectively. The working electrodes used in this study were GCE and modified with CNT to present CNT/GCE by mechanical attachment method [12].

Preparing for CNT modified GCE

A mechanical attachment technique method [13, 14] used direct attachment (about thirty times) of CNT in powder form and modified GCE by abrasively adhere CNT nanoparticles onto the GCE surface. The modification of GCE was done every run of CV to avoid dropping of CNT from the surface of GCE into the solution of cell.

Results and discussion

Effect of ascorbic acid

Figure 1 shows the effect for the oxidation current peaks of Hg2+ with and without AA on modified GCE/CNT working electrode using CV. It was observed that the oxidation peak of Hg2+ appeared at +260 mV. It was considerably enhanced by three–four times in the presence of 1 mM AA, five–seven times in 2 mM AA, and six–eight times in 3 mM AA.

Fig. 1
figure 1

Voltammograms of oxidation current peak of 10 mM Hg(II) in 0.1 M KCl effected by the different concentrations of ascorbic acid (a) without AA, (b) 1 mM AA, (c) 2 mM AA, and (d) 3 mM AA (using GCE/CNT, 100 mV s−1 vs. Ag/AgCl)

It was found that mercuric ions exhibit oxidizing agent behavior when used in acidic media such as AA solution which enhances the oxidation current peak at CV analytic study as shown in Figs. 2 and 3.

Fig. 2
figure 2

Cyclic voltammogram of the oxidation current peak of 10 mM Hg2+ in 0.1 M KCl (using GCE, 100 mV s−1 vs. Ag/AgCl)

Fig. 3
figure 3

Voltammograms of the oxidation current peak of 10 mM Hg2+ 0.1 M KCl and 10 mM AA (using GCE, 100 mV s−1 vs. Ag/AgCl)

Effect of lower pH (acidity AA)

The effect of acidic solution at lower pH using AA solution on Hg(II) with KCl as supporting electrolyte. It was observed that the oxidation current of Hg2+ gradually increases linearly with lowering of pH from 6 to 2 accompanied by linear shift in oxidation potential, E(mV) = −5.546 pH + 224.09 with correlation coefficient of R 2 = 09267 and I(μA) = 1.646 pH − 0.7637 with correlation coefficient of R 2 = 0.9844; the oxidation peak increased in about twofolds and shifting toward origin potential [1] as shown in Fig. 4.

Fig. 4
figure 4

Voltammograms for effect of AA in the oxidation current peaks of 1 mM Hg2+ in 0.1 M KCl (using GCE/CNT, 100 mV s−1 vs. Ag/AgCl)

Effect of AA on Hg(II) in blood medium

In this study, the effect of AA on the high oxidative compounds such as Hg(II) in blood medium was studied using cyclic voltammetric technique. The danger lies in the case of a perimeter of an acid such as AA which works as electrocatalysis, so for its acidity media to enhance the oxidative current peak which causes blood to disintegrate and the formation of a complex leads to death.

Figure 5 illustrates high current peak of mercury ions in blood medium as an oxidative agent in human blood. The more danger is due to the presence of acidic pH like AA solution which enhances the oxidation current peak of Hg(II) as shown in Figs. 6 and 7.

Fig. 5
figure 5

Cyclic voltammograms of the oxidation current peak of 0.1 M Hg2+ in mix of blood medium and 0.1 M KCl (using GCE, 100 mV s−1 vs. Ag/AgCl)

Fig. 6
figure 6

Cyclic voltammogram of the oxidation current peak of 10 mM Hg2+ in mix of blood medium with 0.1 M KCl and 1 mM AA (using GCE, 100 mV s−1 vs. Ag/AgCl)

Fig. 7
figure 7

Cyclic voltammogram of the oxidation current peak of 10 mM Hg2+ in mix of blood medium with 0.1 M KCl and 10 mM AA (using GCE, 100 mV s−1 vs. Ag/AgCl)

Effect of different concentrations of AA

The calibration plot (Fig. 8) was performed at the GCE in 10 mM Hg2+ solutions with a good linearity of anodic current versus Hg2+ described by Y = 1.4X + 19.5, R 2 = 0.98.

Fig. 8
figure 8

Plot of oxidation current versus different concentrations of AA (1–4 mM) in blood medium with 10 mM HgCl2 and 0.1 M KCl using GCE versus Ag/AgCl

Conclusion

Voltammetric determination of Hg(II) ions in blood media was studied using modified GCE with and without modified CNT. The determination was based on the finding of the results of the oxidation current peak by cyclic voltammetric technique. The enhancement of oxidation current peak of Hg(II) in the presence of AA causes the decomposition and damage in blood composition. This contribution accentuates the role of electrochemical techniques in the determination of oxidant activity in the biological blood sample. Electrochemical techniques represent due to selectivity and sensitivity suitable tool for the determination of oxidant capacity in biological blood sample, where these methods are sensitive under the low concentrations of AA affected on the oxidation current peak of Hg(II). In addition, obtaining oxidation current peaks and their position enable us to estimate the possible oxidant agent present in a blood sample. These findings suggest that mercury present in the environment and in particular in foodstuffs especially in seafood of first necessity causes haematological disturbances in the blood composition.