Abstract
3D printing has provided a new prospective in the manufacturing of personalized medical implants, including fistulas for haemodialysis (HD). In the current study, an optimized fused modelling deposition (FDM) 3D printing method has been validated, for the first time, to obtain cylindrical shaped fistulas. Printing parameters were evaluated for the manufacturing of fistulas using blank and 0.25% curcumin-loaded filaments that were produced by hot melt extrusion (HME). Four different fistula types have been designed and characterized using a variety of physicochemical characterization methods. Each design was printed three times to demonstrate printing process accuracy considering outer and inner diameter, wall thickness, width, and length. A thermoplastic polyurethane (TPU) biocompatible elastomer was chosen, showing good mechanical properties with a high elastic modulus and maximum elongation, as well as stability at high temperatures with less than 0.8% of degradation at the range between 25 and 250 °C. Curcumin release profile has been evaluated in a saline buffer, obtaining a low release (12%) and demonstrating drug could continue release for a longer period, and for as long as grafts should remain in patient body. Possibility to produce drug-loaded grafts using one-step method as well as 3D printing process and TPU filaments containing curcumin printability has been demonstrated.
Graphical abstract
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.Avoid common mistakes on your manuscript.
Introduction
Chronic kidney disease (CKD) is a pathologic condition characterized by the decreased function and altered structure of the kidneys. The kidneys have a variety of functions, including filtering the blood to eliminate waste and excess fluid, to support red blood cells and to help controlling the blood pressure [1]. In 2017, there were 697.5 million cases of CKD in the world, with 9.1% of prevalence and 1.2 million deaths [2]. According to the World Health Organization (WHO), from 2000 to 2019, kidney diseases have risen from the world’s 13th leading cause of death to the 10th. Mortality has increased from 813,000 in 2000 to 1.3 million in 2019 [3].
When impaired kidneys are functional only at 10–15%, patients undergo a dialysis treatment for replacing healthy kidneys [1]. Haemodialysis (HD) is a technique where blood circulates through dialyser membrane channels, while the dialysis fluid with an electrolyte composition like extracellular fluid is passing outside these channels in the opposite direction, in order to eliminate wastes from the patient blood [4].
Entrance points obtained with a non-invasive surgical procedure are required into patient blood vessels for patients under HD to connect blood to the dialyser [1]. A well-functioning vascular access (VA) will lead to an efficient HD procedure. Native arteriovenous fistula (AVF) is the first preferred approach, arteriovenous graft (AVG) and central venous catheter being the main types of dialysis access used [5]. AVG is an AVF that has been made with prosthetic interposition between an artery and a vein to link two vessels, which could not be connected by AVF due to the distance between them [5]. AVF remains the first choice for haemodialysis access, compared to AVG and catheter; however, when AVF become unappropriated due to unsuitable veins or maturation issues, AVG is the second choice to mitigate catheter use, mostly in elderly population [6].
A variety of synthetic vascular grafts were made in the past decade that composed of polytetrafluoroethylene (ePTFE) or polyethylene terephthalate (Dacron©), which can successfully apply to replace large diameter vessels (> 6.0 mm). These products show some issues that are related to the replacement of small diameter blood vessels and to their susceptibility to infection [7]. Furthermore, grafts could need to be placed in a looped configuration; if the loop is too tight, it could lead to kinking at the apex and reducing diameter dimension [8]. Therefore, the aim is to fabricate grafts according to the vasculature of each patient. Tissue-engineering technologies could apply for overcoming commercial-related graft problems considering possibility nowadays to obtain bio-mimicking vascular tubes [9].
In 1960s, xenogeneic, and later allogeneic, grafts are used which did not lead to immunogenic reaction after implantation; however, they needed biological treatment to remove collagen, connective tissue proteins and cellular remnants, with the aim of decrease in immunogenicity [10]. Furthermore, cryopreserved grafts or femoral vein homograft results failed in clinical environments and are rarely used, even if they are available in commerce [11].
Electrospinning is a well-established technique for the fabrication of small diameter cell-free scaffold tissue-engineered vascular grafts (TEVGs), with adjustable nano- and micro-fibres, porous structures favourable for cell infiltration and regeneration [12]. Electrospinning parameters, such as electric field, distance needle collector, flow rate and needle diameter, and solution parameters including solvent, polymer concentration, viscosity and solution conductivity and environmental parameters (humidity, temperature) could affect electrospun fibre characteristics (e.g. smooth, bead-free) [13].
3D printing (3DP) is a new prospective in the medical technologies’ era, to produce TEVGs, due to the possibility to print structures composed of cells, growth factors, drugs and scaffolds assembled in an organized way [14]. 3DP systems have shown a promising solution compared to the electrospinning inability to control a three-dimensional structure [15]. In addition, 3DP could reduce fabrication time and costs, as both cells and scaffolds can be processed at the same time [14].
One of the recurring issues using 3DP technologies is that often a scaffold is used to maintain layers position for achieving the requested structure, such as cylindrical shape [16]. However, 3DP method could be a potential application for creating patient characteristics that matched vascular grafts, and the versatility of using different types of materials during printing process, providing the possibility to add drugs to the final device [17]. In vascular graft manufacturing, anti-coagulant drugs could be added to reduce thrombosis events, which is one of the main vascular graft related consequences. In the last 4 years, the use of 3DP has been one of the latest prospects in this field; however, only a few studies are present in the literature regarding vessel constructs or vascular grafts [8].
For the successful manufacturing of grafts, printing adjustments are required due to the specific vessel shape, such as the rotary system to obtain cylindrical geometries and the avoidance on using support materials for printing a cell-laden bioink [18]. Different materials could be used in order to obtain biocompatible vascular grafts [19], such as sodium alginate, 4-arm-poly(-ethylene glycol)-tetra-acrylate (PEGTA) and gelatin methacryloyl (GelMA), using an inkject bioprinter with coaxial nozzles [20]. Bioinks based on hydrophilic polymer, such as 4-acryloylmorpholine (ACMO), have been employed for 3D printed grafts using a digital light processing (DLP) technology [8].
In this work, vascular grafts were produced using 3DP by adjusting different parameters to find a suitable method to print required shapes. Required cylindrical shapes are hard to obtain using a 3DP method, mainly due to material characteristics, 3D printer working methods and shape geometry. However, the fused deposition modelling (FDM) 3DP has been demonstrated, in our studies, to be suitable to produce vascular grafts with small size diameter (< 6.0 mm). FDM technology provides an accurate pore dimension, morphology and interconnectivity of grafts architectures control, while giving important benefits such as price, high speed and simplicity [21]. After testing different TPU grades (TPU70, TPU80 and TPU95), TPU 80 was the selected polymer, which has been generally used for biomedical applications [22], such as catheter production [17]. Curcumin has been chosen in this study to be added consequently due to its anticoagulant [23, 24] and anti-bacterial properties [25], in order to overcome graft related problems mentioned above. Hot melt extrusion (HME) was used to combine TPU 80 with curcumin to create 0.25% drug-loaded filaments, which could then be used to print drug-loaded vascular grafts. The grafts were then characterized using a variety of physicochemical characterization techniques.
Materials and methods
Materials
Elastollan thermoplastic polyurethane (TPU) pellets were kindly provided by DistruPol Ltd. (Dublin, Ireland). Curcumin was purchased from Tokyo Chemical Industry Ltd (UK), the Castor oil was obtained from Ransom (Hitchin, UK) and the phosphate buffer solution (PBS) at pH 7.4 was prepared using PBS tablets, Merck (Darmstadt, Germany). Methanol, ethanol and Tween 80 were purchased from Merck (Darmstadt, Germany), all in analytical grade.
CAD
Using a computer design system (Thinkercad©), four different vascular grafts were designed (Fig. 1; Table 1), to avoid kinking problems related to straight shape.
Preparation of vascular grafts with TPUs using FDM 3D printer
TPUs vascular grafts were prepared exploiting the FDM 3DP system. Three different TPU elasticity grades (95A, 80A, 70A) were investigated to create vascular grafts (Table 1) using an Ultimaker 2 (Ultimaker B.V., Geldermalsen, the Netherlands) fused filament fabrication (FFF) system and Cura software. The Ultimaker 2 FFF system was equipped with a 0.4-mm nozzle, which is a RepRap Open Source FFF equipment. Different optimized printing parameters are used for the three TPUs (e.g. 95, 80, 70) due to their different characteristics (Table 2).
Preparation of TPU filaments containing curcumin and vascular grafts
TPU filaments containing curcumin were prepared using a single-screw HME. Oil method using castor oil was used to ensure curcumin coating on TPU pellet surface. After filling a 50-mL tube with TPU 80 pellets (30 g), castor oil (30 μL) was added into the tube, and the solution was vortexed to achieve a homogeneous coating of pellets. Oil-coated pellets were then transferred to a new 50-mL tube to avoid loss of drug, by sticking to leftover oil on the tube walls. After that, 0.25% of curcumin was added and vortexed to coat all pellets. Coated pellets were added to filament extruder (3Devo, Utretch, the Netherlands) at an extrusion speed of 3 rpm and a filament fan speed of 80%. The temperature was adjusted according to the material between 170 and 200° C and filament with 2.85 mm diameter extruded.
3D printed vascular graft characterization
Microscopy
Filaments and 3D printed grafts were assessed using a Leica EZ4 W digital microscope (Leica, Wetzlar, Germany). The surface morphology of the 3D printed grafts with and without curcumin was also evaluated using scanning electron microscope (SEM) (Hitachi TM3030, Tokyo, Japan).
FT-IR spectroscopy
Pellet TPU80, curcumin and TPU filaments containing curcumin were analysed by attenuated total reflection-Fourier transform infrared spectroscopy (ATR-FTIR) to establish if any interactions were present between drug and TPU. Spectra were recorded using a Nicolet™ iS50 FTIR Spectrometer (Thermo Fisher Scientific, UK).
Thermal analysis
Thermal degradation behaviour was investigated due to high temperatures required during HME and 3DP. A Q500 thermogravimetric analyser (TGA, TA Instruments, New Castle, DE, USA) was used to perform TGA on 5–10 mg TPU80, curcumin and TPU filaments containing curcumin, to measure their weight loss. The analyses performed from 25 to 250 °C with a 10 °C increase per minute and a nitrogen flow rate of 40 mL min−1.
Moreover, differential scanning calorimetry (DSC) was also performed to determinate material transitions subjected to high temperatures. DSC 214 Polyma for Polymer Characterization from Netzsch was used, with a range temperature from 25 to 250 °C with a 10 °C increase per minute and a nitrogen flow rate of 40 to 60 mL min−1.
Mechanical properties
TA.XT plus texture analyser (Stable Micro Systems, Surrey, UK) was used to evaluate the elastic modulus and the fracture force of the TPU 80 and filaments containing curcumin. Pieces of filament (6.0 cm) were cut and clamped vertically into the texture analyser, where the distance between both clamps was 2.0 cm and a test speed of 3.0 mm s−1. The elastic modulus was calculated from the slope from obtained stress/strain curves.
In vitro drug release studies
Drug release study was performed to calculate the amount of curcumin eluting from vascular grafts. Due to curcumin low solubility, vascular grafts were placed in 3 mL of PBS solution containing 0.5% v/v Tween 80 [26,27,28]. PBS with 0.5% Tween 80 and 3% methanol (PTM) was placed in vial placed in a shaking incubator at 37 °C and 40 rpm, according to literature [29]. Different intervals were chosen to analyse the samples (e.g. 2, 4, 6, 24, 48, and 72 h, 1 week and 2 weeks). At each time point, grafts were removed from previous PTM, dried and placed in a fresh 3 mL PTM.
Degradation study
Degradation study took place using gravimetric analysis by placing 1 cm vascular graft in PBS (7.2 pH) and in PTM at 37 °C and were performed at 24, 48, 72 h, 1, 2 weeks and 3 weeks, to establish TPU filaments containing curcumin and TPU 80 mass loss into the two buffers.
Statistical analysis
Quantitative data was expressed as a mean ± standard deviation, n = 3. The statistical analysis was performed using a one-way analysis of variance; p < 0.05 was considered statistically significant.
Results and discussions
Fabrication of vascular grafts
The FDM 3DP system has been used to print a commercial TPU 95A (Ninjaflex) filament, to optimize the process printing parameters using three samples for each design and to demonstrate the possibility printing the required shapes with high grades of accuracy. Several parameters needed to be adjusted to print TPU 95; printing process has been performed at different speeds decreasing to 5.0 mm s−1 as optimized speed and 0.2 mm layer height, with 100% infill. Three samples for each design were printed with an outer diameter (OD) of 6.0 mm, an inner diameter (ID) of 4.0 mm and a wall thickness of 2.0 mm (Fig. 2). Obtained samples have been measured using light microscopy or a calliper, as shown in Table 3. The most significance differences are related to wall thickness, mostly in V-shape, U-shape and C-shape, due to difficulties of printing these specific shapes. Straight sizes are more closed than other designed shapes/sizes. However, size differences between designed and obtained grafts were included within the statistically accepted values; consequently, printing process accuracy has been demonstrated.
Three different TPUs (95A, 80, 70) with differing degrees of elasticity were tested at first considering their elasticity, flexibility and printability. Printing process speed has been tested starting from 5.0 mm s−1 for TPU 95 to 1.0 or 2.0 mm s−1 for TPU 80 and 70, respectively. Low speed is required due to sticky material characteristics and shape instability. Nozzle moving with high speed generally deposited excess molten TPU causing shape instability and losing some material, which could result in the presence of holes in the final graft. Nozzle travel speed from side to side was considered too; high speed during process could move the deposited graft on the stage causing possible holes to be present at the end of printing. Surface morphology of different printed vascular grafts was then studied with SEM (Fig. 3) to evaluate which filament created smooth and linear surfaces. Light microscopy was also used to evaluate graft size and wall thickness. From SEM analysis, 3D printed vascular grafts showed a linear surface. Light microscopy showed good outline scaffold and accurate wall thickness. However, TPU 80 showed smoother surface. TPU 70 was almost immediately discarded from the study due to difficulty in printing of this material. Consequently, due to TPU 70 strong sticky characteristics, Bowden tube needed to be lubricated using glycerine to avoid it getting stuck during printing process.
TPU 80 was chosen for this work consequently to its favoured physical characteristics of elasticity and flexibility, compared to TPU 95 (Fig. 4). TPU 80 has a better printability compared to TPU 70; printing process parameters have been adjusted as listed in Table 2 for each material. As listed in Table 2, 1 mm s−1 speed was chosen to print TPU 80, as well as layer height which was decreased to the lowest value of 0.1 mm; consequently, lower values will produce slower prints in high resolution. Initial layer height has also been set to 0.1 mm and line width to 0.3 mm. Infill density has been settled at 100% to obtain regular surface, avoiding the presence of holes. Bowden lubricated tube had to be used during printing process due to sticky characteristics.
Smooth filaments of 2.85 mm in diameter were produced through HME using TPU 80 pellets and 0.25% curcumin powder. Curcumin was mixed successfully with TPU matrix, as SEM analysis (Fig. 5) confirmed the homogenous distribution of curcumin along the filament with no aggregates within the extruded filament. This can be also inferred by filament colour (Fig. 6) where a homogenous colour is present throughout the tube.
Vascular grafts were printed and characterized using the curcumin loaded extruded filaments with straight shape of 65 mm length, OD of 6.0 mm and ID of 4.0 mm (Figs. 6 and 7), using the same adjusted printing process parameters used for TPU 80, demonstrating no changes due to the presence of curcumin on graft printability.
FT-IR characterization
FT-IR spectra (Fig. 8) were obtained from curcumin powder, TPU filaments containing curcumin and TPU 80 pellet, to establish possible interaction between different materials. No interaction between curcumin and TPU were shown, probably due to the lower concentration of curcumin used for this study.
Thermal analysis
TGA (Fig. 9) and DSC (Fig. 10) analyses were conducted on three samples of curcumin powder, TPU 80 pellet and filament not printed (NP), TPU filaments containing curcumin pellet NP, in a range temperature between 25 and 250 °C, to investigate also the HME (200 °C at maximum) and printer (210 °C nozzle temperature) temperatures.
Thermal analysis also did not show any curcumin influence in filament stability, considering TPU has maintained a good stability at high temperature, despite curcumin instability. Curcumin started degrading around 180 °C, according to its melting point [30]; otherwise, no peaks associated with curcumin degradation could be noticed in filament containing curcumin and pellet. However, neat TPU naturally include hard segment and soft segment with two different melting temperatures [31]. Soft segment peak usually around 20 °C could not be noticed in the analysis because 25 °C was used as initial temperature; hard segments melting temperature could be noticed in Fig. 9, around 110 °C in TPU 80 pellet and TPU pellet containing curcumin [32]. However, thermal analysis demonstrated TPU stability at high temperature used for extrusion and printing.
Degradation studies
Degradation studies were performed by placing grafts containing curcumin into PBS, to analyse surface modifications after 3 weeks (Fig. 11). The graft weight was also considered, with differences remaining within statistical accepted values starting from 156.3 mg at time 0 to 160.9 mg on 3rd week.
Mechanical properties of vascular grafts
TPU 80 and TPU containing curcumin were evaluated to demonstrate no changes in mechanical properties after curcumin loading. The filament containing curcumin samples did not break even after 20 cm distance, which was the maximum distance input into the texture analyser. Same results were also obtained with TPU 80 samples. Ultimate tensile strength (UTS), maximum elongation, and elastic modulus obtained from stress and strain curve (Fig. 12) and stiffness (Table 4) were calculated for each sample (n = 3).
As shown in Table 4, clear differences could be noticed between curcumin-loaded TPU and TPU 80 UTS, while there are no changes with max elongation. However, the extruded filament reserves good elasticity considering no breaks were noticed during analysis of all three samples.
In vitro release of curcumin from vascular grafts
Drug release studies were analysed using UV analysis at 425 nm length. Due to curcumin’s low solubility, vascular grafts were placed in 3 mL of PBS solution containing 0.5% v/v Tween 80. It is well known that curcumin exhibits a pH-dependent stability in acidic conditions; in vitro release study was carried out at acidic pH (pH 6.0) [33]. After several literature studies, Tween 80 turned out to be a good additive into buffer, which has been used in different studies in order to sustain curcumin release [26,27,28]. Several manuscripts have also added methanol into buffer, in order to have a certain curcumin release, even if methanol is not assessed in physiological conditions [34, 35]. An initial burst release was detected at day 1, and the release gradually continued until the end of the 2-week study (Fig. 13). Curcumin released was around 61 μg compared to 75 mg used for filament extrusion. Considered the small loss of curcumin during HME, around 80% was incorporated into grafts; as previously reported [17, 36], curcumin might be interacting with TPU due to TPU hydrophobic nature, limiting drug release; therefore, curcumin will continue to release also after the 2 weeks. Using material such as TPU, drug could remain in the graft for a longer period if grafts remain permanently in the patient body. However, considering curcumin activity, the amount of curcumin required to exhibit anticoagulant effects is around 5.5 mg to match heparin activity. Seventy-five milligrams of curcumin was used to prepare filament; in this way, around 10.83 mg of curcumin could be enough to be used in a 20-cm vascular graft. Low curcumin concentration (0.25%) was chosen at first to demonstrate printability of drug within main material. In order to have a higher release concentration, curcumin amount in filament should be increased. The curcumin amount which could be released from 20 cm graft might be capable in exhibiting an anti-bacterial effect; curcumin MIC is between 129 and 293 μg mL−1 [25], 10.83 mg is present in 20 cm vascular graft as cited before and 411.35 μg mL−1 (12%) will be released in 2 weeks.
Conclusions
This study demonstrates the potential use of 3DP FDM method in the manufacture of vascular grafts, in order to overcome inter-personal characteristics and obtain personalized grafts suitable for different vein and arterial sizes. User-friendly CAD software and speedy printing process allow personalized vascular graft production. Different shaped designs could overcome kinking problems caused by straight shape in commerce. Thanks to 3DP technique, in several years, it could be possible to manufacture directly patient veins and arteries to create a personalized graft in a short time, avoiding failing maturation problem and infection. 3DP method was validated using a FDM printing technique despite difficult cylindrical configuration to obtain. Moreover, incorporation of drugs in TPU using HME has been demonstrated with SEM and TGA. It has also been shown that the presence of drug does not influence TPU mechanical properties. Filaments obtained were successfully used to prepare vascular grafts using 3DP. Anti-coagulant drug incorporation could avoid thrombosis problem generally caused using grafts.
The methods, described in this manuscript, could be used for different medical applications, as well as ease incorporation of drugs. One-step filament production could lead to short time production process and low costs; drug melting point and temperature degradation have to be investigated in order to demonstrate the possibility of incorporation.
Data availability
The datasets generated during and/or analysed during the current study are available from the corresponding author on reasonable request.
References
National Kidney Foundation. https://www.kidney.org/. Accessed 15 Jan 2021.
Bikbov B, Purcell C. Global, regional, and national burden of chronic kidney disease, 1990–2017: a systematic analysis for the Global Burden of Disease Study 2017. The Lancet. 2020. https://doi.org/10.1016/S0140-6736(20)30045-3.
World Health Organisation. The top 10 causes of deaht. 2020. https://www.who.int/news-room/fact-sheets/detail/the-top-10-causes-of-death. Accessed Jul 2021
Mallick N, Gokal R. Haemodialysis. The Lancet. 1999. https://doi.org/10.1016/s0140-6736(97)09411-7.
Santoro D, Benedetto F. Vascular access for hemodialysis: current perspectives. Int J Nephrol Renovasc Dis. 2014. https://doi.org/10.2147/IJNRD.S46643.
Agarwal A, Haddad N. Innovations in vascular access for hemodialysis. Kidney Int. 2019. https://doi.org/10.1016/j.kint.2018.11.046.
Radakoovic D, Reboredo J. A multilayered electrospun graft as vascular access for hemodialysis. PLoS ONE. 2017. https://doi.org/10.1371/journal.pone.0185916.
Cheng B, Xing Y. The formulation and characterization of 3D printed grafts as vascular access for potential use in hemodialysis. RSC Adv. 2018;8:15471–9.
Awad N, Niu H. Electrospun fibrous scaffolds for small-diameter blood vessels: a review. Memb. 2018. https://doi.org/10.3390/membranes8010015.
Gage SM, Lawson JH. Bioengineered hemodialysis access grafts. J of Vasc Access. 2017;18:S56–63.
Peck M, Dusserre N. New biological solutions for hemodialysis access. J of Vasc Access. 2011. https://doi.org/10.5301/JVA.2011.6451.
Jin X, Geng X. Preparation of small-diameter tissue-engineered vascular grafts electrospun from heparin end-capped PCL and evaluation in a rabbit carotid artery replacement model. Macromol Biosci. 2019. https://doi.org/10.1002/mabi.201900114.
Haider A, Haider S. Comprehensive review summarizing the effect of electrospinning parameters and potentialapplications of nanofibers in biomedical and biotechnology. Arabian J of Chem. 2015. https://doi.org/10.1016/j.arabjc.2015.11.015.
Wenger R, Giraud M. 3D Printing applied to tissue engineered vascular grafts. Appl Sci. 2018. https://doi.org/10.3390/app8122631.
Saghati S, Akbarzedeh A. Electrospinning and 3D printing: prospects for market opportunity in Kny E, Ghosal K, Thomas S, editors. Electrospinning: from basic research to commercialization, 2018.
Patra S, Young V. A review of 3D printing techniques and the future in biofabrication of bioprinted tissue. Cell Biochem Biophys. 2016. https://doi.org/10.1007/s12013-016-0730-0.
Mathew E, Dominguez RJ. Fused deposition modelling as an effective tool for anti-infective dialysis catheter fabrication. ACS Biomat Sci and Eng. 2019;5:6300–10.
Freeman S, Ramos R. A bioink blend for rotary 3D bioprinting tissue engineered small- diameter vascular constructs. Acta Biomat. 2019. https://doi.org/10.1016/j.actbio.2019.06.052.
Wang X, Liu F. Synthetic polymers for organ 3D printing. Polymer. 2020. https://doi.org/10.3390/polym12081765.
Jia W, Gungor-Ozkerim P. Direct 3D bioprinting of perfusable vascular constructs using a blend bioink. Biomat. 2016. https://doi.org/10.1016/j.biomaterials.2016.07.038.
Hassan M, Dave K. 3D printing of biopolymer nanocomposites for tissue engineering: Nanomaterials, processing and structure-function relation. Eur Pol J. 2019. https://doi.org/10.1016/j.eurpolymj.2019.109340.
Bergmeister H, Seyidova N. Biodegradable, thermoplastic polyurethane grafts for small diameter vascular replacements. 2015. https://doi.org/10.1016/j.eurpolymj.2019.109340.
Lee MJ, Kim SE. Curcumin-loaded biodegradable polyurethane scaffolds modified with gelatin using 3D printing technology for cartilage tissue engineering. Adv Healthc Mater. 2019. https://doi.org/10.1002/adhm.201400018.
Kim DC, Ku S. Anticoagulant activities of curcumin and its derivative. BMP Rep. 2012;45:221–6.
Gunes H, Gulen D. Antibacterial effects of curcumin: an in vitro minimum inhibitory concentration study. Toxicol Ind Health. 2016. https://doi.org/10.1177/0748233713498458.
Sun XZ, Williams GR. Electrospun curcumin-loaded fibers with potential biomedical applications. Carbohydr Polym. 2013. https://doi.org/10.1016/j.carbpol.2012.12.064.
Moradkhannejhad L, Adbouss M. The effect of molecular weight and content of PEG on in vitro drug release of electrospun curcumin loaded PLA/PEG nanofibers. J of Drug Delivery Sci and Technol. 2020. https://doi.org/10.1016/j.jddst.2020.101554.
Ayubi M, Karimi M. Magnetic nanoparticles decorated with PEGylated curcumin as dual targeted drug delivery: synthesis, toxicity and biocompatibility study. Mater Sci Eng C Mater Biol Appl. 2019. https://doi.org/10.1016/j.msec.2019.109810.
Correia GS, Bruglia ML. Microfluidic manufacturing of phospholipid nanoparticles: stability, encapsulation efficacy, and drug release. Int J Pharm. 2017. https://doi.org/10.1016/j.ijpharm.2016.11.025.
PubChem. Curcumin. https://pubchem.ncbi.nlm.nih.gov/. Accessed 20 Feb 2021.
Xu J, Cheng L. Highly exfoliated montmorillonite clay reinforcedthermoplastic polyurethane elastomer:in situpreparation and efficient strengthening. RSC Adv. 2019. https://doi.org/10.1039/C8RA10121C.
Silva G, Calado H. Polymer nanocomposites based on P3OT, TPU and SWNT: preparation and characterization. 2006. https://doi.org/10.1109/ICONN.20066.340581.
Kharat M, Du Z. Physical and chemical stability of curcumin in aqueous solutions and emulsions; impact of pH, temperature, and molecular environment. J of Agricoultur and Food Chem. 2017. https://doi.org/10.1021/acs.jafc.6b04815.
Lee MJ, Koim SE. Curcumin‐loaded biodegradable polyurethane scaffolds modified with gelatin using 3D printing technology for cartilage tissue engineering. 2019. https://doi.org/10.1002/adhm.201400018.
Kasoju N, Bora U. Fabrication and characterization of curcumin-releasing silk fibroin scaffold. 2012. https://doi.org/10.1002/jbm.b.32753.
Farmer ZL, Utomo E. 3D printed estradiol-eluting urogynecological mesh implants: Influence of material and mesh geometry on their mechanical properties. 2021. https://doi.org/10.1016/j.ijpharm.2020.120145
Acknowledgements
The authors would like to thank the ERASMUS+ programme for the mobile scholarship to S.B.
Author information
Authors and Affiliations
Contributions
All authors contributed to the study conception and design. Material preparation, data collection and analysis were performed by Sara Basile, Essyrose Mathew and Dimitrios A. Lamprou. The first draft of the manuscript was written by Sara Basile, and all authors commented on previous versions of the manuscript. All authors read and approved the final manuscript.
Corresponding authors
Ethics declarations
Ethics approval and consent to participate
Not applicable.
Consent for publication
Not applicable.
Conflict of interests
The authors declare no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Basile, S., Mathew, E., Genta, I. et al. Optimization of FDM 3D printing process parameters to produce haemodialysis curcumin-loaded vascular grafts. Drug Deliv. and Transl. Res. 13, 2058–2071 (2023). https://doi.org/10.1007/s13346-021-01078-2
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s13346-021-01078-2