A case of false hypoglycemia by SMBG due to improper storage of glucometer test strips

  • Case Report
  • Published:
Diabetology International Aims and scope Submit manuscript

Abstract

Many types of glucometer test strips deteriorate in the presence of humidity. Storing test strips outside their container or removing them from their individual wrappers could worsen their measurement accuracy. While the measurement accuracy can gradually deteriorate under such conditions, some patients are unaware of this aspect of test strip care and handling. Although such a deterioration in glucometer accuracy is often seen in clinical practice, there are few reports that warn about this issue. This is the first case report of pseudohypoglycemia caused by the inappropriate storage of test strips, which resulted in inadequate use of medications and deterioration of glycemic control. In clinical practice, patients sometimes handle glucometers in unexpected ways. This can arise due to inadequate instruction. Patients should be advised to read the manual associated with the device carefully and to store the test strips in their original container, closing the cap tightly, and to take each test strip out of its individual wrapper just before it is used. Careful patient instruction is needed to ensure safe and stable diabetic control.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

References

  1. Shue HW. Addressing self-monitoring of blood glucose: advocating paired glycemic testing for people with type 2 diabetes. J Diabetes Investig. 2012;3:337–8.

    Google Scholar 

  2. Tamaki M, Shimizu T, Kanazawa A, et al. Effects of changes in basal/total daily insulin ratio in type 2 diabetes patients on intensive insulin therapy including insulin glargine (JUN-LAN Study 6). Diabetes Res Clin Pract. 2008;81:e1–3.

    Article  CAS  PubMed  Google Scholar 

  3. Varanauskiene E. Can blood glucose self-monitoring improve treatment outcomes in type 2 diabetes? Diabetes Res Clin Pract. 2008;82(Suppl 2):S112–7.

    Article  PubMed  Google Scholar 

  4. Bamberg R, Schulman K, MacKenzie M, et al. Effect of adverse storage conditions on performance of glucometer test strips. Clin Lab Sci. 2005;18:203–9.

    PubMed  Google Scholar 

  5. King JM, Eigenmann CA, Colagiuri S. Effect of ambient temperature and humidity on performance of blood glucose meters. Diabet Med. 1995;12:337–40.

    Article  CAS  PubMed  Google Scholar 

  6. Fink KS, Christensen DB, Ellsworth A. Effect of high altitude on blood glucose meter performance. Diabetes Technol Ther. 2002;4:627–35.

    Article  PubMed  Google Scholar 

  7. Haller MJ, Shuster JJ, Schatz D, et al. Adverse impact of temperature and humidity on blood glucose monitoring reliability: a pilot study. Diabetes Technol Ther. 2007;9:1–9.

    Article  CAS  PubMed  Google Scholar 

  8. Yamaoka H, Sode K. SPCE based glucose sensor employing novel thermostable glucose dehydrogenase, FADGDH: blood glucose measurement with 150 nL sample in one second. J Diabetes Sci Technol. 2007;1:28–35.

    Google Scholar 

  9. Park S, Boo H, Chung TD. Electrochemical non-enzymatic glucose sensors. Anal Chim Acta. 2006;556:46–57.

    Article  CAS  PubMed  Google Scholar 

  10. Lu J, Bu RF, Sun ZL, et al. Comparable efficacy of self-monitoring of quantitative urine glucose with self-monitoring of blood glucose on glycaemic control in non-insulin-treated type 2 diabetes. Diabetes Res Clin Pract. 2011;93:179–86.

    Article  CAS  PubMed  Google Scholar 

  11. Kilo C Sr, Dickey WT Jr, Joynes JO, et al. Evaluation of a new blood glucose monitoring system with auto-calibration for home and hospital bedside use. Diabetes Res Clin Pract. 2006;74:66–74.

    Article  CAS  PubMed  Google Scholar 

  12. Price DA. Case study: recurrent diabetic ketoacidosis resulting from spurious hypoglycemia: a deleterious consequence of inadequate detection of partial strip filling by a glucose monitoring system. Clin Diabetes. 2009;27:164–6.

    Article  Google Scholar 

  13. Silverstein JH, House DV, Schatz DA, et al. Accuracy of self-monitoring blood glucose meters in a high humidity, high temperature summer camp setting. Diabetes Care. 1995;18:1200–1.

    CAS  PubMed  Google Scholar 

  14. Ishima H, Noriko N, Ogata S, et al. The effects of preserving SMBG test strips under unwrapped condition on glucose readings. J Japan Diab Soc. 2010;53:42–7 (in Japanese).

    Google Scholar 

Download references

Acknowledgments

We thank the staff at Tokushima Prefectural Central Hospital for their excellent patient care.

Conflict of interest

The authors declare no conflict of interest.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Akio Kanazawa.

About this article

Cite this article

Tamaki, M., Kanazawa, A., Shirakami, A. et al. A case of false hypoglycemia by SMBG due to improper storage of glucometer test strips. Diabetol Int 5, 199–201 (2014). https://doi.org/10.1007/s13340-013-0147-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s13340-013-0147-6

Keywords

Navigation