Comparison of forehead infrared thermometry with axillary digital thermometry in neonates

  • Research Brief
  • Published:
Indian Pediatrics Aims and scope Submit manuscript

Abstract

Fluctuations in core body temperature beyond a narrow range are of concern as they indicate changed homeostasis. Thermoregulation remains a key aspect of neonatal care. New ‘non touch’ method of measuring temperature may reduce infection rate and discomfort of neonates. Body temperature of neonates admitted in Neonatal Intensive Care Unit was measured using axillary digital thermometer and a handheld infrared non touch thermometer. The two methods did not agree well (mean difference = −0.5, 95% limits of agreement: [−2.3, 1.2]). The agreement was similar with a negligible difference when patients in open care warmers were excluded (mean difference = −0.6, 95% limits of agreement: [−2.3, 1.1]). As the accuracy is unsatisfactory, the said technology needs further improvement before it can be adapted for routine patient care.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. Martin SA, Kline AM. Can there be a standard for temperature measurement in the pediatric intensive care unit? AACN Clin Issues. 2004;15:254–266.

    PubMed  Google Scholar 

  2. Chiappini E, Sollai S, Longhi R, Morandini L, Laghi A, Osio CE, et al. Performance of non-contact infrared thermometer for detecting febrile children in hospital and ambulatory settings. J Clin Nurs. 2011;20:1311–1318.

    Article  PubMed  Google Scholar 

  3. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–310.

    Article  CAS  PubMed  Google Scholar 

  4. Fortuna EL, Carney MM, Macy M, Stanley RM, Younger JG, Bradin SA. Accuracy of non-contact infrared thermometry versus rectal thermometry in young children evaluated in the emergency department for fever. J Emerg Nurs. 2010;36:101–104.

    Article  PubMed  Google Scholar 

  5. Sener S, Karcioglu O, Eken C, Yaylaci S, Ozsarac M. Agreement between axillary, tympanic, and mid-forehead body temperature measurements in adult emergency department patients. Eur J Emerg Med. 2012;19:252–256.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Somashekhar Nimbalkar.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Sethi, A., Patel, D., Nimbalkar, A. et al. Comparison of forehead infrared thermometry with axillary digital thermometry in neonates. Indian Pediatr 50, 1153–1154 (2013). https://doi.org/10.1007/s13312-013-0302-y

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s13312-013-0302-y

Keywords

Navigation