Accuracy and efficiency of published film dosimetry techniques using a flat-bed scanner and EBT3 film

  • Scientific Paper
  • Published:
Australasian Physical & Engineering Sciences in Medicine Aims and scope Submit manuscript

Abstract

Gafchromic EBT3 film is widely used for patient specific quality assurance of complex treatment plans. Film dosimetry techniques commonly involve the use of transmission scanning to produce TIFF files, which are analysed using a non-linear calibration relationship between the dose and red channel net optical density (netOD). Numerous film calibration techniques featured in the literature have not been independently verified or evaluated. A range of previously published film dosimetry techniques were re-evaluated, to identify whether these methods produce better results than the commonly-used non-linear, netOD method. EBT3 film was irradiated at calibration doses between 0 and 4000 cGy and 25 pieces of film were irradiated at 200 cGy to evaluate uniformity. The film was scanned using two different scanners: The Epson Perfection V800 and the Epson Expression 10000XL. Calibration curves, uncertainty in the fit of the curve, overall uncertainty and uniformity were calculated following the methods described by the different calibration techniques. It was found that protocols based on a conventional film dosimetry technique produced results that were accurate and uniform to within 1%, while some of the unconventional techniques produced much higher uncertainties (> 25% for some techniques). Some of the uncommon methods produced reliable results when irradiated to the standard treatment doses (< 400 cGy), however none could be recommended as an efficient or accurate replacement for a common film analysis technique which uses transmission scanning, red colour channel analysis, netOD and a non-linear calibration curve for measuring doses up to 4000 cGy when using EBT3 film.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

Similar content being viewed by others

References

  1. Niroomand-Rad A, Blackwell CR, Coursey BM, Gall KP, Galvin JM, Galvin WL, McLaughlin WL, Meigooni AS, Nath R, Rodgers JE, Soares CG (1998) Radiochromic film dosimetry: recommendations of the AAPM radiation therapy committee task group 55. Med Phys 25(11):2093–2115

    Article  CAS  PubMed  Google Scholar 

  2. Ashland (2017) Gafchromic dosimetry media—type EBT-3. http://www.gafchromic.com/documents/EBT3_Specifications.pdf Accessed 9 Aug 2016

  3. Micke A, Lewis DF, Yu X (2011) Multichannel film dosimetry with nonuniformity correction. Med Phys 35(5):2523–2534

    Article  Google Scholar 

  4. Van Hoof SJ, Granton PV, Landry G, Podesta M, Verhaegen F (2012) Evaluation of a novel triple-channel radiochromic film analysis procedure using EBT2. Phys Med Biol 57:4353–4368

    Article  PubMed  Google Scholar 

  5. Butson E, Alnawaf H, Yu PKN, Butson M (2011) Scanner uniformity improvements for radiochromic film analysis with matt reflectance backing. Australas Phys Sci Eng Med 34:401–407

    Article  Google Scholar 

  6. Reinhardt S, Hillbrand M, Wilkens JJ, Assmann W (2012) Comparison of Gafchromic EBT2 and EBT3 films for clinical photon and proton beams. Med Phys 39(8):5257–5265

    Article  CAS  PubMed  Google Scholar 

  7. International Speciality Products (2011) Gafchromic EBT3. http://www.ashland.com/Ashland/Static/Documents/ASI/Advanced%20Materials/gafchromic-ebt3.pdf. Accessed 9 Aug 2016

  8. Aland T, Kairn T, Kenny J (2011) Evaluation of a Gafchromic EBT2 film dosimetry system for radiotherapy quality assurance. Australas Phys Sci Eng Med 34:251–260

    Article  CAS  Google Scholar 

  9. Lynch BD, Kozelka J, Ranade MK, Li JG, Simon WE, Dempsey JF (2006) Important considerations for radiochromic film dosimetry with flatbed CCD scanners and EBT Gafchromic film. Med Phys 33(12):4551–4556

    Article  PubMed  Google Scholar 

  10. Devic S, Soares CG, Kirov AS, Ali I, Williamson F, Elizondo A (2004) Dosimetric properties of improved Gafchromic films for seven different digitizers. Med Phys 31(9):2392–2401

    Article  CAS  PubMed  Google Scholar 

  11. Martiskova M, Ackerman B, Jakel O (2008) Analysis of uncertainties in Gafchromic EBT film dosimetry of photon beams. Phys Med Biol 53:7013–7027

    Article  Google Scholar 

  12. Devic S, Seuntjens J, Sham E, Podgorsak EB, Schmidtlein CR, Kirov AS, Soares CG (2005) Precise radiochromic film dosimetry using a flat-bed document scanner. Med Phys 32:2245–2253

    Article  Google Scholar 

  13. Kairn T, Aland T (2012) Calibration of radiochromic film for electron radiotherapy dosimetry. IFMBE Proc 39:1153–1156

    Article  Google Scholar 

  14. Moylan R, Aland T, Kairn T (2013) Dosimetric accuracy of Gafchromic EBT2 and EBT3 film for in vivo dosimetry. Australas Phys Sci Eng Med 36:331–337

    Article  Google Scholar 

  15. Warrener K, Hug B, Liu P, Ralston A, Ebert MA, McKenzie D, Suchowerska N (2014) Small field in-air output factors: the role of miniphantom design and dosimeter type. J Med Phys 41:21723–21728

    Article  Google Scholar 

  16. Wen N, Lu S, Kim J, Qin Y, Huang Y, Zhao B, Liu C, Chetty IJ (2016) Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiation therapy quality assurance using Gafchromic EBT3 films. Radiat Oncol 11(132)

  17. Hu Y, Wang Y, Fogerty G, Liu G (2013) Developing a novel method to analyse Gafchromic EBT2 films in intensity modulated radiation therapy quality assurance. Australas Phys Sci Eng Med 36:487–494

    Article  Google Scholar 

  18. Yao T, Luthjens LH, Gasparini A, Warman JM (2017) A study of four radiochromic films currently used for (2D) radiation dosimetry. Radiat Phys Chem 133:37–44

    Article  CAS  Google Scholar 

  19. Tamponi M, Bona R, Poggiu A, Marini P (2016) A new form of the calibration curve in radiochromic dosimetry, properties and results. Med Phys 43(7):4435–4446

    Article  CAS  PubMed  Google Scholar 

  20. Bennie N, Metcalfe P (2015) Practical IMRT QA dosimetry using Gafchromic film: a quick start guide. Australas Phys Eng Sci Med 39:533–545

    Article  Google Scholar 

  21. Kairn T, Aland T, Kenny J (2010) Local heterogeneities in early batches of EBT2 film: a suggested solution. Phys Med Biol 55:37–42

    Article  Google Scholar 

  22. Devic S, Tomic N, Lewis D (2016) Reference radiochromic film dosimetry: review of technical aspects. Phys Med. https://doi.org/10.1016/j.ejmp.2016.02.008

    Google Scholar 

  23. Peet S, Wilks R, Kairn T, Trapp JV, Crowe SB (2016) Calibrating radiochromic film in beams of uncertain quality. Med Phys 43(10):5647–5652

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to E. Spelleken.

Ethics declarations

Conflict of interest

All authors declare that they have no conflict of interest.

Ethical approval

This article does not contain any studies with human participants or animals performed by any of the authors.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Spelleken, E., Crowe, S.B., Sutherland, B. et al. Accuracy and efficiency of published film dosimetry techniques using a flat-bed scanner and EBT3 film. Australas Phys Eng Sci Med 41, 117–128 (2018). https://doi.org/10.1007/s13246-018-0620-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s13246-018-0620-4

Keywords

Navigation