Abstract
The aim of this work was to investigate the impact of the accuracy of a certain couch model on the overall accuracy of the dose calculation. For this purpose two models of the Varian Exact Couch (Varian Medical Systems, Palo Alto, CA, USA) for use in the planning of volumetric modulated arc therapy (VMAT) were created in Monaco 5.1 (Elekta, Stockholm, Sweden). The first was a simplistic model, with the relative electron densities (ED) of structures based entirely on CT Hounsfield units (HU). The second couch model was obtained by assigning the appropriate EDs to the relevant couch structures. Instead of getting EDs from the CT scan using the Hounsfield units to ED conversion, the second approach was to determine the optimal ED iteratively. The values accepted for the second couch model were the EDs that produced the best agreement between Monaco calculations and the measured couch attenuation. Finally, the performance of the two couch models in clinical prostate VMAT plans was compared.
Similar content being viewed by others
References
Olch AJ, Gerig L, Li H, Mihaylov I, Morgan A (2014) Dosimetric effects caused by couch tops and immobilization devices: Report of AAPM Task Group 176. Med Phys 41(20):061501
ORS CSP12 (1992) Code of safe practice for the use of irradiating apparatus in medical therapy, Office of Radiation Safety, New Zealand Ministry of Health
Pulliam KB, Howell RM, Followill D, Luo D, White RA, Kry SF (2011) The clinical impact of the couch top and rails on IMRT and arc therapy. Phys Med Biol 56(23):7435–7447
Spezi E, Angelini A, Romani F, Guido A, Bunkheila F, Ntreta M, Ferri A (2008) Evaluating the influence of the Siemens IGRT carbon fibre tabletop in head and neck IMRT. Radiother Oncol 89:114–122
Myint WK, Niedbala M, Wilkins D, Gerig LH (2006) Investigating treatment dose error due to beam attenuation by a carbon fibre tabletop, J Appl Clin Med Phys 7(3): 21–27
Van Prooijen M, Kanesalingam T, Islam MK, Heaton RK (2010) Assessment and management of radiotherapy beam intersections with the treatment couch, J Appl Clin Med Phys 11(2):3171
Hu Z, Dai J, Li L, Cao Y, Fu G (2011) Evaluating and modeling of photon beam attenuation by a standard treatment couch, J Appl Clin Med Phys 12(4):3561
Chetty IJ, Curran B, Cygler JE, DeMarco JJ, Ezzell G, Faddegon BA, Kawrakow I, Keall PJ, Liu H, Ma CM, Rogers DW, Seuntjens J, Sheikh-Bagheri D, Siebers JV (2007) Report of the AAPM Task Group no. 105: issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning, Med Phys 34(12):4818–4853
Vanetti E, Nicolini G, Clivio A, Fogliata A, Cozzi L (2009) The impact of treatment couch modelling on Rapidarc. Phys Med Biol 54(9):N157–N166
Teke T, Gill B, Duzenli C, Popescu IA (2011) A Monte Carlo model of the Varian IGRT couch top for RapidArc QA. Phys Med Biol 56(24):N295–N305
IAEA (2000) Absorbed dose determination in external beam radiotherapy. Technical Reports Series No.398, IAEA, Vienna pp. 21–26, 78
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical approval
For this type of study formal consent is not required.
Rights and permissions
About this article
Cite this article
Chyou, Ty., Lorenz, F. Couch modelling for volumetric modulated arc therapy (VMAT). Australas Phys Eng Sci Med 40, 471–480 (2017). https://doi.org/10.1007/s13246-017-0552-4
Received:
Accepted:
Published:
Issue Date: