Skip to main content
Log in

In Vitro Evaluation of Ventricular Cannulation for Rotodynamic Cardiac Assist Devices

  • Published:
Cardiovascular Engineering and Technology Aims and scope Submit manuscript

Abstract

The influence of positioning and geometry of ventricular cannulas for contemporary continuous flow Left Ventricular Assist Devices (LVADs) was evaluated in a non-beating isolated heart preparation with borescopic visualization. Preload and LVAD flow were varied to evaluate degrees of ventricular decompression up to the point of ventricular collapse. The performance of a flanged cannula was compared to a conventional bevel-tipped cannula: quantitatively by the maximal flow attainable, and qualitatively by visualization of fluid tracer particles within the ventricular chamber. Three forms of ventricular suck-down occurred: concentric collapse, gradual entrainment and instantaneous entrainment. In some circumstances, unstable oscillations of the ventricle were observed prior to complete collapse. Under conditions of low preload, the flanged cannula demonstrated less positional sensitivity, provided greater flow, and exhibited fewer areas of stagnation than the beveled cannula. These observations warrant further consideration of a flanged ventricular cannula to mitigate complications encountered with conventional cannulae.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Figure 1
Figure 2
Figure 3
Figure 4
Figure 5
Figure 6
Figure 7

Similar content being viewed by others

References

  1. Amin, D. V., et al. Induction of ventricular collapse by an axial flow blood pump. ASAIO J. 44(5):M685–M690, 1998.

    Article  Google Scholar 

  2. Antaki, J. F., et al. An improved left ventricular cannula for chronic dynamic blood pump support. Artif. Organs 19(7):671–675, 1995.

    Article  Google Scholar 

  3. Bachman, T. Development and evaluation of the quintessential ventricular cannula. In: Biomedical Engineering. Pittsburgh: University of Pittsburgh, 2003, p. 57.

  4. Badiwala, M. V., H. J. Ross, and V. Rao. An unusual complication of support with a continuous-flow cardiac assist device. N. Engl. J. Med. 357(9):936–937, 2007.

    Article  Google Scholar 

  5. Bhama, J. K., et al. Development of an ex vivo ovine ventricular assist device model for intraventricular visualization of the inflow cannula. J. Heart Lung Transplant. 28(8):860–861, 2009.

    Article  Google Scholar 

  6. Choi, S., J. R. Boston, and J. F. Antaki. Hemodynamic controller for left ventricular assist device based on pulsatility ratio. Artif. Organs 31(2):114–125, 2007.

    Article  Google Scholar 

  7. Curtis, A. A multiple modality study of the hemodynamics of left ventricular apical cannulations. In: Bioengineering. Pittsburgh: University of Pittsburgh, 1998.

  8. Curtis, A. S., et al. Novel ventricular apical cannula: in vitro evaluation using transparent, compliant ventricular casts. ASAIO J. 44(5):M691–M695, 1998.

    Article  Google Scholar 

  9. Drummond, A. Biomedical surgical planning for pediatric ventricular assist device (PVAD). In: Biomedical Engineering. Pittsburgh, 2008, p. 189

  10. Ferreira, A., et al. A discriminant-analysis-based suction detection system for rotary blood pumps. Conf. Proc. IEEE Eng. Med. Biol. Soc. 1:5382–5385, 2006.

    Google Scholar 

  11. Griffith, B. P., et al. HeartMate II left ventricular assist system: from concept to first clinical use. Ann. Thorac. Surg. 71(3 Suppl):S116–S120, 2001; discussion S114–S116.

    Google Scholar 

  12. Miyake, Y., et al. Left ventricular mobile thrombus associated with ventricular assist device: diagnosis by transesophageal echocardiography. Circ. J. 68(4):383–384, 2004.

    Article  MathSciNet  Google Scholar 

  13. Reesink, K., et al. Suction due to left ventricular assist: implications for device control and management. Artif. Organs 31(7):542–549, 2007.

    Article  Google Scholar 

  14. Vollkron, M., et al. Development of a suction detection system for axial blood pumps. Artif. Organs 28(8):709–716, 2004.

    Article  Google Scholar 

  15. Votapka, T. V., et al. Left ventricular cannula obstruction in a patient with previous ventricular aneurysmectomy. Ann. Thorac. Surg. 58(4):1182–1184, 1994.

    Article  Google Scholar 

  16. Watanabe, K., et al. Development of a flexible inflow cannula with titanium inflow tip for the NEDO biventricular assist device. ASAIO J. 50(4):381–386, 2004.

    Google Scholar 

Download references

Acknowledgments

The authors would like to thank Arielle Drummond, PhD; and Fangjun Shu, PhD; for their assistance with the development of the isolated heart preparation. The Authors would also like to thank Harvey Borovetz, PhD for his support. Financial support was provided by an STTR Grant from the National Institute of Health 1R41HL082251-01.

Conflict of interest

At the time of the study, Josiah Verkaik was employed by LaunchPoint technologies, Inc. All remaining authors have no disclosures to report.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to James F. Antaki.

Additional information

Associate Editor Keefe B. Manning oversaw the review of this article.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Movie 1- Borescopic and transepicardial echocardiographic long-axis view of concentric ventricular collapse and resulting oscillation generated using aligned flared cannula. (MPG 11581 kb)

Movie 2- Borescopic view of instantaneous ventricular entrainment and resulting oscillation generated using misaligned flared cannula. (MPG 10705 kb)

Movie 3- Borescopic and transepicardial echocardiographic long-axis view of concentric ventricular collapse generated using aligned beveled cannula. (MPG 10705 kb)

Movie 4- Borescopic view of gradual ventricular entrainment and resulting oscillation generated using misaligned beveled cannula. Movie was recorded at 4x normal speed. (MPG 11429 kb)

Rights and permissions

Reprints and permissions

About this article

Cite this article

Bachman, T.N., Bhama, J.K., Verkaik, J. et al. In Vitro Evaluation of Ventricular Cannulation for Rotodynamic Cardiac Assist Devices. Cardiovasc Eng Tech 2, 203–211 (2011). https://doi.org/10.1007/s13239-011-0050-x

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s13239-011-0050-x

Keywords

Navigation