Integrity of stent polymer layer after drug-eluting stent implantation: in vivo comparison of sirolimus-, paclitaxel-, zotarolimus- and everolimus-eluting stents

  • Original Article
  • Published:
Cardiovascular Intervention and Therapeutics Aims and scope Submit manuscript

Abstract

Few data exist with regard to the polymer integrity of drug-eluting stents (DES) in vivo. This study aims to investigate the integrity of the polymer layer of 4 polymer-coated DES in vivo. We assessed the morphology of the polymer layer of sirolimus-eluting stent (SES; Cypher Select™), paclitaxel-eluting stent (PES; Taxus Liberté™), zotarolimus-eluting stent (ZES; Endeavour RX™) and everolimus-eluting stent (EES; Xience V™) by scanning electron microscopy after balloon expansion at nominal and high pressures in the coronary arteries of 3 pigs. Effects of kissing balloon procedure were also explored. The polymer layer of SES, PES and EES were damaged in less than 3 % of the surface area with high pressure procedures, whereas the damaged area reached 38.0 ± 2.6 % in ZES (P < 0.01). The polymer integrity differed greatly among DES after balloon inflation in vivo. This should be taken into account when placing DES in tortuous vessels, calcified, as well as bifurcation lesions because the polymer layer may be easily damaged in these lesions.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5

Similar content being viewed by others

References

  1. Versaci F, Gaspardone A, Tomai F, Crea F, Chiariello L, Gioffre PA. A comparison of coronary-artery stenting with angioplasty for isolated stenosis of the proximal left anterior descending coronary artery. N Engl J Med. 1997;336:817–22.

    Article  CAS  PubMed  Google Scholar 

  2. Grines CL, Cox DA, Stone GW, Garcia E, Mattos LA, Giambartolomei A, Brodie BR, Madonna O, Eijgelshoven M, Lansky AJ, O’Neill WW, Morice MC. Coronary angioplasty with or without stent implantation for acute myocardial infarction. Stent primary angioplasty in myocardial infarction study group. N Engl J Med. 1999;341:1949–56.

    Article  CAS  PubMed  Google Scholar 

  3. Kastrati A, Dibra A, Mehilli J, Mayer S, Pinieck S, Pache J, Dirschinger J, Schomig A. Predictive factors of restenosis after coronary implantation of sirolimus- or paclitaxel-eluting stents. Circulation. 2006;113:2293–300.

    Article  PubMed  Google Scholar 

  4. Serruys PW, Kutryk MJ, Ong AT. Coronary-artery stents. N Engl J Med. 2006;354:483–95.

    Article  CAS  PubMed  Google Scholar 

  5. Stone GW, Ellis SG, Cannon L, Mann JT, Greenberg JD, Spriggs D, O’Shaughnessy CD, DeMaio S, Hall P, Popma JJ, Koglin J, Russell ME. Comparison of a polymer-based paclitaxel-eluting stent with a bare metal stent in patients with complex coronary artery disease: a randomized controlled trial. JAMA. 2005;294:1215–23.

    Article  CAS  PubMed  Google Scholar 

  6. Serruys PW, Morice MC, Kappetein AP, Colombo A, Holmes DR, Mack MJ, Stahle E, Feldman TE, van den Brand M, Bass EJ, Van Dyck N, Leadley K, Dawkins KD, Mohr FW. Percutaneous coronary intervention versus coronary-artery bypass grafting for severe coronary artery disease. N Engl J Med. 2009;360:961–72.

    Article  CAS  PubMed  Google Scholar 

  7. Otsuka Y, Chronos NA, Apkarian RP, Robinson KA. Scanning electron microscopic analysis of defects in polymer coatings of three commercially available stents: comparison of biodivysio, taxus and cypher stents. J Invasive Cardiol. 2007;19:71–6.

    PubMed  Google Scholar 

  8. Daemen J, Serruys PW. Drug-eluting stent update 2007: Part i. A survey of current and future generation drug-eluting stents: Meaningful advances or more of the same? Circulation. 2007;116:316–28.

    Article  PubMed  Google Scholar 

  9. Khattab AA, Otto A, Hochadel M, Toelg R, Geist V, Richardt G. Drug-eluting stents versus bare metal stents following rotational atherectomy for heavily calcified coronary lesions: late angiographic and clinical follow-up results. J Interv Cardiol. 2007;20:100–6.

    Article  PubMed  Google Scholar 

  10. Abdel-Wahab M, Baev R, Dieker P, Kassner G, Khattab AA, Toelg R, Sulimov D, Geist V, Richardt G. Long-term clinical outcome of rotational atherectomy followed by drug-eluting stent implantation in complex calcified coronary lesions. Catheter Cardiovasc Interv. 2013;81:285–91.

    Article  PubMed  Google Scholar 

  11. Fujii K, Mintz GS, Kobayashi Y, Carlier SG, Takebayashi H, Yasuda T, Moussa I, Dangas G, Mehran R, Lansky AJ, Reyes A, Kreps E, Collins M, Colombo A, Stone GW, Teirstein PS, Leon MB, Moses JW. Contribution of stent under expansion to recurrence after sirolimus-eluting stent implantation for in-stent restenosis. Circulation. 2004;109:1085–8.

    Article  PubMed  Google Scholar 

  12. Wiemer M, Butz T, Schmidt W, Schmitz KP, Horstkotte D, Langer C. Scanning electron microscopic analysis of different drug eluting stents after failed implantation: from nearly undamaged to major damaged polymers. Catheter Cardiovasc Interv. 2010;75:905–11.

    PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Tetsuya Watanabe.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Watanabe, T., Fujita, M., Awata, M. et al. Integrity of stent polymer layer after drug-eluting stent implantation: in vivo comparison of sirolimus-, paclitaxel-, zotarolimus- and everolimus-eluting stents. Cardiovasc Interv and Ther 29, 4–10 (2014). https://doi.org/10.1007/s12928-013-0191-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s12928-013-0191-y

Keywords

Navigation