Advertisement

MAPAN

, Volume 33, Issue 2, pp 131–137 | Cite as

Characterization of Capacitance Standards at High Frequency at National Physical Laboratory, India

Original Paper
  • 36 Downloads

Abstract

Four-terminal-pair air dielectric capacitance standards with nominal values of 1000 and 100 pF have been characterized up-to 10 MHz at NPLI. The procedure employed involves the determination of all capacitive and inductive parameters of the simple electrical-equivalent-circuit-model of these capacitance standards. The effective capacitance of each standard has also been computed as a function of frequency from 1 kHz to 10 MHz. The capacitive parameters have been measured at 1 kHz while inductive parameters have been estimated up to 10 MHz using linear regression analysis by employing least-squares-approximation method. The paper highlights the computation procedure of impedance terms which further requires the determination of various capacitive and inductive terms involved in the calculation of effective capacitance. The method employed for the estimation of inductive parameters as a function of frequency is also discussed in detail. The present work will help in the establishment of metrological traceability of capacitance standards at high-frequency at NPLI which will be further used to establish calibration facility for LCR meters and RF impedance analyzers for capacitance parameter up-to 10 MHz.

Keywords

Capacitance standards High frequency Impedance-matrix Electrical equivalent circuit model Inductive Capacitive 

Notes

Acknowledgements

The authors are grateful to The Director NPL, India for permission to publish this work and Dr. V. N. Ojha, Head, Time & Frequency and Electrical & Electronic Metrology division for constant encouragement and motivation. The authors are also thankful to Council of Scientific and Industrial Research (CSIR), India for funding this work under MIST project.

References

  1. [1]
    Babita, Satish, A. Singh and A.K. Saxena, Determination and validation of average value of national standard of DC voltage using different methods at CSIR-NPLI, MAPAN-J. Metrol. Soc. India, 28 (2013) 99–104.Google Scholar
  2. [2]
    Babita, L.K. Narula, T. John, M. Saleem and A.K. Saxena, Preliminary study for the establishment of DC charge calibration facility at CSIR-NPL Using charged capacitance source, MAPAN-J. Metrol. Soc. India, 29 (2014) 269–272.Google Scholar
  3. [3]
    Satish, M.A. Ansari and A.K. Saxena, Determination and comparison of temperature coefficient of standard inductors by measuring change in inductance and resistances, MAPAN-J. Metrol. Soc. India, 29 (2013) 73–76.Google Scholar
  4. [4]
    Joint Committee for Guides in Metrology (JCGM), International vocabulary of metrology—basic and general concepts and associated terms (VIM), Int. Vocab. Metrol., 3 (2008) 104.Google Scholar
  5. [5]
    Satish, S. Kumar, Babita, T. John and A. K. Saxena, Realization of coaxial reference air-lines as high frequency capacitance standard at CSIR-NPL, Measurement, 92 (2016) 166–171.Google Scholar
  6. [6]
    Satish, N. Sawhney, S. Kumar and A. K. Saxena, Evaluation of three terminal capacitance standards at CSIR-NPL, MAPAN-J. Metrol. Soc. India, 30 (2015) 261–265.Google Scholar
  7. [7]
    Satish, Babita, B. Khurana, S. Kumar and A.K. Saxena, Evaluation of four-terminal-pair capacitance standards using electrical equivalent circuit model, Measurement, 73 (2015) 121–126.Google Scholar
  8. [8]
    S. Singh, S. Kumar, Babita and T. John, Realization of four-terminal-pair capacitors as reference standards of impedance at high frequency using impedance-matrix method, IEEE Trans. Instrum. Meas., 66 (2017) 2129–2135.Google Scholar
  9. [9]
    Satish, S. Kumar, Babita, T. John and A.K. Saxena, Evaluation of air dielectric four-terminal-pair capacitance standards using resonance frequency of impedance elements, Measurement, 100 (2017) 176–182.Google Scholar
  10. [10]
    Satish, Babita, B. Khurana, and T. John, Measurement automation to implement evaluation procedure of four-terminal-pair capacitance standards using S-parameters, MAPAN-J. Metrol. Soc. India, 32 (2017) 175–181.Google Scholar
  11. [11]
    S. Singh and T. John, Automated module for characterization of reference standards of capacitance using impedance-matrix method, Turkish J. Electr. Eng. Comput. Sci., (2017).Google Scholar
  12. [12]
    R.N. Jones, Evaluation of three terminal and four terminal pair capacitors at high frequencies. NBS Technical Note 1024 (1980).Google Scholar
  13. [13]
    K. Suzuki, A new universal calibration method for four- terminal-pair admittance standards, IEEE Trans. Instrum. Meas., 40 (1991) 420–422.CrossRefGoogle Scholar
  14. [14]
    T. Aoki, K. Suzuki and K. Yokoi, Calibration method for four-terminal-pair capacitance standards: progress report, In Conference on Precision Electromagnetic Measurements, (1998) 36–37.Google Scholar
  15. [15]
    A.D. Koffman, B.C. Waltrip, N.M. Oldham and S. Avramov-Zamurovic, Capacitance and dissipation factor measurements from 1 kHz to 10 MHz, In Nat. Conf. Standards Laboratories (NCSL) Workshop and Symp. (1998) 63–68.Google Scholar
  16. [16]
    L. Callegaro and F. Durbiano, Four-terminal-pair impedances and scattering parameters, Meas. Sci. Technol., 14 (2003) 523–529.CrossRefADSGoogle Scholar
  17. [17]
    T. Özkan, G. Gülmez, E. Turhan and Y. Gülmez, Four-terminal-pair capacitance characterization at frequencies up to 30 MHz using resonance frequencies, Meas. Sci. Technol., 18 (2007) 3496–3500.CrossRefADSGoogle Scholar
  18. [18]
    S.A. Awan, B.P. Kibble, I.A. Robinson and S.P. Giblin, A new four terminal-pair bridge for traceable impedance measurements at frequencies up to 1 MHz, IEEE Trans. Instrum. Meas., 50 (2001) 282–285.CrossRefGoogle Scholar
  19. [19]
    R.D. Cutkosky, Techniques for comparing four-terminal-pair admittance standards, 74 (1970) 63–78.Google Scholar
  20. [20]
    BIPM, Evaluation of measurement data—guide to the expression of uncertainty in measurement, (2008).Google Scholar
  21. [21]
    S.L. Dahake, R.N. Dhar, A.K. Saxena, V.K. Batra and K. Chandra, Progress in the realization of the units of capacitance, resistance and inductance at the National Physical Laboratory, India, IEEE Trans. Instrum. Meas., IM-32 (1983) 5–8.Google Scholar
  22. [22]
    S. A. Awan and B.P. Kibble, Towards accurate measurement of the frequency dependence of capacitance and resistance standards up to 10 MHz, IEEE Trans. Instrum. Meas. 54, (2005) 516–520.CrossRefGoogle Scholar

Copyright information

© Metrology Society of India 2017

Authors and Affiliations

  1. 1.LF, HF Impedance and DC Metrology, CSIR-NPLNew DelhiIndia
  2. 2.Amity UniversityJaipurIndia

Personalised recommendations