Combined use of an Airtraq® optical laryngoscope, Airtraq video camera, Airtraq wireless monitor, and a fibreoptic bronchoscope after failed tracheal intubation

  • Manuel Ángel Gómez-RíosEmail author
  • Laura Nieto Serradilla
Correspondence

Keywords

Obstructive Sleep Apnea Obstructive Sleep Apnea Syndrome Remifentanil Difficult Airway Fibreoptic Bronchoscope 
These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

To the Editor,

Optical technology and fibreoptic systems have resulted in significant advances in management of the difficult airway. Optical laryngoscopes, such as the Airtraq® (Prodol Meditec S.A., Vizcaya, Spain), provide an improved indirect view of the upper airway, achieving a higher intubation success rate than direct laryngoscopy. However, insertion and advancement of the endotracheal tube (ETT) may occasionally fail. The fibreoptic bronchoscope (FOB) remains the gold standard for the management of difficult intubation. Nevertheless, in rare circumstances, even fibreoptic intubation fails.1 We present the simultaneous use of an Airtraq optical laryngoscope, Airtraq video camera (AVC), Airtraq wireless monitor (AWM), and a FOB in an airway that was grossly distorted due to compression by a cervical tumour.

The patient was a 48-yr-old 150-kg male with morbid obesity (body mass index: 52 kg m−2) and obstructive sleep apnea syndrome. He was scheduled for a total thyroidectomy. Preoperative airway assessment included a Mallampati grade 3, an upper lip bite test2 class II, macroglossia, a thick neck, and a large left cervical mass that compressed and displaced laryngeal structures and the upper tracheal lumen (Figure, Panels A and B). Following application of standard monitors, pre-oxygenation with a facemask, administration of a nasal vasoconstrictor, and topical anesthesia of the naris, oropharynx, and nasopharynx, the patient was sedated with midazolam 1 mg iv and remifentanil at 0.15 μg kg−1 min−1. Attempts were made to insert the FOB through the patient’s nose and mouth, but his vocal cords could not be visualized. The Airtraq was inserted without increasing sedation or administering extra topical analgesia. Guided by bubbling secretions during spontaneous ventilation, this step allowed partial visualization of the glottis in the upper right corner of the viewfinder. Additional maneuvers were ineffective in optimizing visualization and centring the glottis in the middle of the viewfinder, hence, intubation was deemed impossible. Without removing the Airtraq, a FOB was inserted through the size 7.0 ETT mounted in the guiding channel. By viewing the FOB tip in the vicinity of the glottis through the AVC in the AWM, rotating the body of the FOB, and manipulating the tip control lever, the FOB was inserted into the patient’s trachea and the ETT was advanced over the FOB. The procedure was carried out uneventfully.

Cervical masses may lead to serious airway compromise resulting from anatomical distortion secondary to obstruction of the pharynx or deviation of the larynx,1 scenarios where any airway device can be fallible. Passage of a FOB to expose the glottis may be very difficult. On the other hand, the Airtraq optical laryngoscope can be effective in managing difficult airways; however, successful intubation requires optimal positioning of the glottis in the middle of the viewfinder. In patients with a large tongue, obesity, or a cervical mass, combined use of the Airtraq and the FOB may compensate each others’ limitations.3 The Airtraq can provide an unobstructed airway for the FOB, allowing its tip to be sited in the immediate vicinity of the glottic opening. At the same time, the Airtraq provides visualization of the ETT advancing over the FOB4 while the FOB negotiates the sharp angle between the tip of the ETT and the glottis.3

In three published case reports, combined use of the Airtraq and the FOB has been described in similar scenarios in which both devices had failed separately.3-5 Two anesthesiologists were necessary in all cases, one to place the Airtraq while observing the location of the FOB tip by the viewfinder and the other to manage the FOB while manipulating the device according to verbal feedback from the anesthesiologist managing the Airtraq. By using a third device, the AVC, tracheal intubation can be performed by a single operator with the support of an assistant holding the Airtraq in position. This approach makes manipulation of the FOB tip more precise under visualization with an external monitor (Figure, Panel C). The AVC is an optional accessory of Airtraq that can be used to transmit the images to the AWM by radiofrequency. We consider this system, which was designed initially for training or teaching purposes, as a useful complement in difficult airway management.

No single airway instrument is perfect in all circumstances. While failure of the devices described herein is uncommon, operator experience should be considered carefully. As always, appropriate steps must be taken to ensure oxygenation and ventilation, and alternate plans to secure the airway must be immediately available.
Figure

(A) Patient before the surgery. (B) Thyroid TC scan shows a severely enlarged left thyroid lobe (approximate dimensions of 12 × 9 × 15 cm) that compressed larynx structures and the esophagus and displaced them to the right. (C) Procedure of orotracheal intubation of a manniquin using the four devices

Notes

Competing interests

None declared.

References

  1. 1.
    Wulf H, Brinkmann G, Rautenberg M. Management of the difficult airway. A case of failed fiberoptic intubation. Acta Anaesthesiol Scand 1997; 41: 1080-2.PubMedCrossRefGoogle Scholar
  2. 2.
    Khan ZH, Kashfi A, Ebrahimkhani E. A comparison of the upper lip bite test (a simple new technique) with modified Mallampati classification in predicting difficulty in endotracheal intubation: a prospective blinded study. Anesth Analg 2003; 96: 595-9.PubMedCrossRefGoogle Scholar
  3. 3.
    Matioc AA. Use of the Airtraq with a fibreoptic bronchoscope in a difficult intubation outside the operating room. Can J Anesth 2008; 55: 561-2.PubMedCrossRefGoogle Scholar
  4. 4.
    Xue FS, He N, Liu JH, Zhang YM. Use of a fiberoptic bronchoscope to facilitate tracheal intubation after failed intubation using the Airtraq laryngoscope. Acta Anaesthesiol Scand 2010; 54: 256-7.PubMedCrossRefGoogle Scholar
  5. 5.
    Schoettker P, Bathory I, Broome M. Use of the nasotracheal Airtraq to assist difficult nasal fibreoptic intubation. Can J Anesth 2008; 55: 884.PubMedCrossRefGoogle Scholar

Copyright information

© Canadian Anesthesiologists' Society 2011

Authors and Affiliations

  • Manuel Ángel Gómez-Ríos
    • 1
    Email author
  • Laura Nieto Serradilla
    • 2
  1. 1.Department of Anaesthesiology and Perioperative MedicineComplejo Hospitalario Universitario A CoruñaCoruñaSpain
  2. 2.Complejo Hospitalario Universitario de VigoVigoSpain

Personalised recommendations