Case
A 77-year-old female patient with a history of chronic obstructive pulmonary disease and paroxysmal atrial fibrillation presented to our emergency department with collapse and was diagnosed with sick sinus syndrome. Before pacemaker implantation, echocardiography showed normal dimensions of both atria and ventricles and normal systolic left and right ventricle (RV) function. A DDD-R pacemaker was successfully implanted using active-fixation leads without any acute complications. The day after implantation, sensing and pacing parameters were normal and unchanged. Chest radiography then showed a normal position of both leads (Fig. 1a, b). Subsequently, the patient was discharged the same day in good clinical health.
Four days later, however, the patient was readmitted because of sharp chest pain, unrelated to physical activity or posture. Pacemaker data showed a marked switch from bipolar to unipolar lead pacing and malcapture of the RV lead at maximal pacemaker output. Chest radiography revealed an altered RV lead position (Fig. 1c, d). Echocardiography suggested a perforation of the RV lead through the RV apex but without pericardial effusion (Fig. 1e and online video). This raised doubt as to whether the lead tip lay in the pericardial space. Ultimately, thoracic computed tomography (CT) showed that the lead went through the myocardium (Fig. 1f, g). The patient was subsequently transferred to a specialized pacing lead extraction center with surgical backup, where the RV lead was repositioned uneventfully.
Discussion
Acute complications (<24 h) after pacemaker implantation occur in 3–7% of patients, of which ~1% is due to myocardial perforation (MP) [1]. In the subacute phase (1–30 days), MP may also occur, though less frequently, with an incidence of 0.03–0.4% of treated patients/year [2, 3].
Several predictors for development of MP are reported: old age (>80 years), female sex, RV apical lead positioning, and steroid use, while active fixation has been matter of debate [2, 3]. In our female patient, the RV lead was positioned in the RV apex. Although she was not over 80 years, she had nearly reached that age. In retrospect, the RV lead could have been positioned more septally.
MP has a wide variation in clinical presentation, ranging from clinically occult cases to cardiac arrest secondary to pericardial tamponade. Chest pain is the most frequently reported symptom [4].
Diagnosing MP may be challenging. Pacemaker parameter abnormalities are a first indication [4], though normal pacemaker function does not exclude this diagnosis [5].
Although routine chest radiography may be useful for evaluation of lead dislodgement, further imaging investigation is indicated if MP is suspected. Echocardiography can be useful in revealing presence of the RV lead in the pericardial space with or without pericardial effusion, but determining the exact lead tip position is difficult. CT, however, is superior in revealing lead tip position and in detecting MP (accuracy 92.9%, sensitivity 100%, and specificity 85.7%; echocardiography: 62.7%, 41.2%, and 84.2%, respectively) [4].
References
Ellenbogen KA, Wood MA, Shepard RK. Delayed complications following pacemaker implantation. Pacing Clin Electrophysiol. 2002;25:1155–8.
Cano Ó, Andrés A, Alonso P, Osca J, Sancho-Tello MJ, Olagüe J, Martínez-Dolz L. Incidence and predictors of clinically relevant cardiac perforation associated with systematic implantation of active-fixation pacing and defibrillation leads: a single-centre experience with over 3800 implanted leads. Europace. 2017;19:96–102.
Sterlinski M, Przybylski A, Maciag A, Syska P, Pytkowski M, Lewandowski M, Kowalik I, Firek B, Kołsut P, Religa G, Kuśmierczyk M, Walczak F, Szwed H. Subacute cardiac perforations associated with active fixation leads. Europace. 2009;11:206–12.
Rajkumar CA, Claridge S, Jackson T, Behar J, Johnson J, Sohal M, Amraoui S, Nair A, Preston R, Gill J, Rajani R, Aldo Rinaldi C. Diagnosis and management of iatrogenic cardiac perforation caused by pacemaker and defibrillator leads. Europace. 2016;. doi:10.1093/europace/euw074 (Epub Jun 27).
Hirschl DA, Jain VR, Spindola-Franco H, Gross JN, Haramati LB. Prevalence and characterization of asymptomatic pacemaker and ICD lead perforation on CT. Pacing Clin Electrophysiol. 2007;30:28–32.
Author information
Authors and Affiliations
Contributions
Author contributions
RP.J. Boxma: drafting article; M.G.M. Kolff-Kamphuis: critical revision of article; R.M.M. Gevers: critical revision of article; M. Boulaksil: critical revision of article and design of figures.
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Electronic supplementary material
Below is the link to the electronic supplementary material.
12574_2017_337_MOESM1_ESM.avi
Online video: Transthoracic echocardiography (subcostal view) showing the RV lead through the RV apex. Pericardial effusion was absent. See also Fig. 1e. (AVI 7044 kb)
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
About this article
Cite this article
Boxma, R.P.J., Kolff-Kamphuis, M.G.M., Gevers, R.M.M. et al. Subacute right ventricular pacemaker lead perforation: evaluation by echocardiography and cardiac CT. J Echocardiogr 15, 188–190 (2017). https://doi.org/10.1007/s12574-017-0337-5
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12574-017-0337-5