Abstract
The aim was to assess the clinical significance of the interaction of the cardiac implantable electronic device (CIED) of the selected patient and the interfering electromagnetic field (EMF) during practicing of a profession in an industrial environment. The next step was to analyze risk factors using the specific measurement on the phantom and to determine the conditions for minimizing the risk. We present a method of the provocative analysis on the phantom of the stimulation system. The phantom approximates the system in patient’s chest. The measurement was performed using real-time analysis of sensed signals and the pacemaker (PCM) response on the programmer to the source of an electromagnetic interference (EMI) at the workplace. Disturbing signals from an external source can cause an emergency operation of PCM, like asynchronous pacing or pacing inhibition. The interaction was assessed in various exposures, changes of device programming and function responses to EMI. The magnetic field was detected in close proximity (< 5 cm) of CIED from the surface of the EMI source. Induced potentials were continuously sensed at close distances from the source and it activated asynchronous stimulation. The response varied according to the system orientation. At greater distances (< 20 cm above, < 10 cm in front) the intermittent signal sensing and the pacing inhibition occurred during standard PCM programming. Zones for safe operation were determined and the PCM programming was optimized. The interference signal sensing is significantly affected by the CIED programming. This patient can continue to practice the profession safely with high probability by following certain precautions and adjusting PCM settings. We present this analysis on the phantom of the CIED as one of options where the measurement does not pose a risk to the patient with an implant.
Similar content being viewed by others
References
Enticos 4 ProMRI. Enticos 8: Technical manual. Germany: Biotronik; 2016.
IEGM Troubleshooting Guide. Advanced Product Support. Germany: Biotronik; 2017.
ČSN EN 50527-2-1 editor. 2:2017. Postup pro hodnocení vystavení zaměstnanců s aktivními implantabilními zdravotnickými prostředky elektromagnetickým polím — Část 2 – 1: Specifické hodnocení zaměstnanců s kardiostimulátory. Praha: Úřad pro technickou normalizaci, metrologii a státní zkušebnictví; 2017.
Morava J, Richter A, Roubicek T, Soucek T. Vliv polohy kardiostimulační elektrody na detekci elektromagnetické interference. Elektrorevue. 2021;22(5):126.
ISO 14117:2012. Active implantable medical devices - Electromagnetic compatibility - EMC test protocols for implantable cardiac pacemakers, implantable cardioverter defibrillators and cardiac resynchronization devices. Switzerland: International Organization for Standardization; 2012.
Seckler T, Stunder D, Schikowsky C, Joosten S, Zink MD, Kraus T, Marx N, Napp A. Effect of lead position and orientation on electromagnetic interference in patients with bipolar cardiovascular implantable electronic devices. Europace. 2019;12(2):319.
Richter A, Ferkova Z, Bittner V. Impact of External Low-dynamic Magnetic Field on Human Body with Metal Implants, Combination of Energy Conversation Principle and Maxwell Equations. Proceedings of the 19th International Conference on Electrical Drives & Power Electronics, pp. 235–241, 2019.
Ellenbogen KA, Wilkoff BL, Kay GN, Lau CP, Auricchio A. Clinical cardiac pacing, defibrillation, and resynchronization therapy. Philadelphia: Elsevier; 2017.
Morava J, Richter A, Eichler J. Detection of Electromagnetic Interference on Electrodes of Cardiac Implantable Electronic Devices. MEASUREMENT 2019: Proceedings of the 12th International Conference on Measurement, pp.182–185, 2019.
Acknowledgements
This research is a partial result of cooperation between Technical University of Liberec, Faculty of Mechatronics and Faculty of Health Studies, and Regional Hospital Liberec, Department of Cardiology.
Funding
This research was supported by Student Grant Competition SGS-2021-3002 of Technical University of Liberec in 2021.
Author information
Authors and Affiliations
Contributions
All authors have more or less equally contributed in generating/collecting data in their respective fields as well as in the compilation of data and manuscript preparation.
Corresponding author
Ethics declarations
Conflict of interest
Authors declare no conflict of interest.
Ethical statement
Authors state that the research was conducted according to ethical standards. The research (reference number: EK/117/2020) was approved by Ethics Committee of the Regional Hospital in Liberec in 12/2020.
Consent to participate
Informed consent was obtained from the patient participating in this study.
Consent for publication
Not applicable.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Morava, J., Richter, A., Kupec, J. et al. Interaction of work electromagnetic field with implanted cardiostimulation system, analysis by in vitro method at patient practicing of profession. Health Technol. 12, 801–807 (2022). https://doi.org/10.1007/s12553-022-00680-9
Received:
Revised:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12553-022-00680-9