Skip to main content
Log in

An inexpensive method of small photon field dosimetry with EBT3 radiochromic film

  • Original Paper
  • Published:
Health and Technology Aims and scope Submit manuscript

Abstract

The relative output factor for small field photon beams with EBT3 (External beam therapy, Gafchromic ™) Radiochromic film were obtained using indigenously developed a program in MATLAB. MLC and JAWS created square small field sizes for dosimetry. To calibrate Radiochromic film, we have cut the film into several small segments of 6 × 6 cm2. These small pieces of film were irradiated to known doses ranging from 25 to 300 cGy by SAD technique using 6 MV photon beam from Linear accelerator. A Program in MATLAB was written to analyze the scanned film data and all the images of the films were imported into this Program. To validate the accuracy of calibration curve, radiochromic films were irradiated with different field sizes of 10 × 10 cm2, 15 × 1 5 cm2, 20 × 20 cm2 to known dose of 200 cGy each by SAD technique. Field sizes were defined using two methods, in the first method fields were defined using X-Y collimator Jaws with the MLC retracted to their maximum position. In the Second method, MLCs were used to define square field sizes of 1 × 1 cm2 to 5 × 5 cm2, 10 × 10 cm2 and 15 × 15 cm2. Calibration curve created using gafchromic film show percentage variation between measured and delivered dose within 3% of standard field size. Output factors measured by Gafchromic EBT3 film showed close agreement with those measured using the other detector for field sizes of 2.0 × 2.0 cm2 and above. EBT3 film has properties such as linear response, energy, and dose-rate independence. It makes it suitable detector for small field dosimetry. Our Indigenous developed MATLAB Program for film dosimetry gave the desired results. From this study, we can conclude that EBT3 film can be used for relative output factors ranging from small field to large fields.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Subscribe and save

Springer+ Basic
$34.99 /Month
  • Get 10 units per month
  • Download Article/Chapter or eBook
  • 1 Unit = 1 Article or 1 Chapter
  • Cancel anytime
Subscribe now

Buy Now

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Wilcox EE, Daskalov GM. Accuracy of dose measurements and calculations within and beyond heterogeneous tissues for 6MV photon fields smaller than 4cm produced by Cyberknife. Med Phys. 2008;35(6):2259–66.

    Article  Google Scholar 

  2. Sturtewagen E, Fuß M, Paelinck L, De Wagter C, Georg D. Multi-dimensional dosimetric verification of stereotactic radiotherapy for uveal melanoma using radiochromic EBT film. Zeitschrift für Medizinische Physik. 2008;18(1):27–36.

    Article  Google Scholar 

  3. IAEA, TRS. 483. Dosimetry of small static fields used in external beam radiotherapy: an IAEA-AAPM international code of practice for reference and relative dose determination. No. 483. Technical Report Series.

  4. Das IJ, Ding GX, Ahnesjö A. Small fields: nonequilibrium radiation dosimetry. Med Phys. 2008;35(1):206–15.

    Article  Google Scholar 

  5. Almond PR, Biggs PJ, Coursey BM, Hanson WF, Huq MS, Nath R, et al. AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams. Med Phys. 1999;26(9):1847–70.

    Article  Google Scholar 

  6. Andreo P, Huq MS, Westermark M, Song H, Tilikidis A, DeWerd L, et al. Protocols for the dosimetry of high-energy photon and electron beams: a comparison of the IAEA TRS-398 and previous international codes of practice. Phys Med Biol. 2002;47(17):3033–53.

    Article  Google Scholar 

  7. Sharma SD. Challenges of small photon field dosimetry are still challenging. J Med Phys. 2014;39(3):131–2.

    Article  MathSciNet  Google Scholar 

  8. Laub WU, Wong T. The volume effect of detectors in the dosimetry of small fields used in IMRT. Med Phys. 2003;30(3):341–7.

    Article  Google Scholar 

  9. Aspradakis MM, Byrne JP, Palmans H, Duane S, Conway J, Warrington AP, Rosser K. IPEM Report 103: small field MV photon dosimetry; 2010.

  10. Rustgi SN, Frye D. Dosimetric characterization of radiosurgical beams with a diamond detector. Med Phys. 1995;22(12):2117–21.

    Article  Google Scholar 

  11. Lewis D, et al. An efficient protocol for radiochromic film dosimetry combining calibration and measurement in a single scan. Med Phys. 2012;39(10):6339–50.

    Article  Google Scholar 

  12. Brown TAD, Hogstrom KR, Alvarez D, Matthews KL, Ham K, Dugas JP. Dose-response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron-produced monochromatic x-ray beams. Med Phys. 2012;39(12):7412–7.

    Article  Google Scholar 

  13. Butson MJ, Cheung T, Peter KN. Weak energy dependence of EBT gafchromic film dose response in the 50kVp–10MVp X-ray range. Appl Radiat Isot. 2006;64(1):60–2.

    Article  Google Scholar 

  14. Morales JE, Butson M, Crowe SB, Hill R, Trapp JV. An experimental extrapolation technique using the Gafchromic EBT3 film for relative output factor measurements in small x-ray fields. Med Phys. 2016;43:4687–92.

    Article  Google Scholar 

  15. Lárraga-Gutiérrez JM, García-Hernández D, García-Garduño OA, Galván de la Cruz OO, Ballesteros-Zebadúa P, Esparza-Moreno KP. Evaluation of the Gafchromic® EBT2 film for the dosimetry of radiosurgical beams. Med Phys. 2012;39(10):6111–7.

    Article  Google Scholar 

  16. García-Garduño OA, Lárraga-Gutiérrez JM, Rodríguez-Villafuerte M, Martínez-Dávalos A, Celis MA. Small photon beam measurements using radiochromic film and Monte Carlo simulations in a water phantom. Radiother Oncol. 2010;96(2):250–3.

    Article  Google Scholar 

  17. Huet C, Dagois S, Derreumaux S, Trompier F, Chenaf C, Robbes I. Characterization and optimization of EBT2 radiochromic films dosimetry system for precise measurements of output factors in small fields used in radiotherapy. Radiat Meas. 2012;47(1):40–9.

    Article  Google Scholar 

  18. Sankar A, Ayyangar KM, Nehru RM, Kurup PG, Murali V, Enke CA, et al. Comparison of Kodak EDR2 and Gafchromic EBT film for intensity-modulated radiation therapy dose distribution verification. Med Dosim. 2007;31(4):273–82.

    Article  Google Scholar 

  19. Devic S, Tomic N, Soares CG, Podgorsak EB. Optimizing the dynamic range extension of a radiochromic film dosimetry system. Med Phys. 2009;36(2):429–37.

    Article  Google Scholar 

  20. Niroomand-Rad A, Blackwell CR, Coursey BM, Gall KP, Galvin JM, McLaughlin WL, et al. Radiochromic film dosimetry: recommendations of AAPM radiation therapy committee task group 55. Med Phys. 1998;25(11):2093–115.

    Article  Google Scholar 

  21. Devic S, Seuntjens J, Sham E, Podgorsak EB, Schmidtlein CR, Kirov AS, et al. Precise radiochromic film dosimetry using a flat-bed document scanner. Med Phys. 2005;32(7):2245–53.

    Article  Google Scholar 

  22. Alfonso P, Andreo P, Capote R, Huq MS, Kilby W, Kjall P, et al. A new formalism for reference dosimetry of small and nonstandard fields. Med Phys. 2008;35:5179–86.

    Article  Google Scholar 

  23. Francescon P, Beddar S, Satariano N, Das IJ. Variation of kQclin, Qmsrfclin, fmsr for the small-field dosimetric parameters percentage depth dose, tissue-maximum ratio, and off-axis ratio. Med Phys. 2014;41(10):101708.

    Article  Google Scholar 

  24. Das Indra J, Morales J, Francescon P. Small field dosimetry: what have we learnt? In: Guerda Massillon JL, et al., editors. AIP conference proceedings. AIP Publishing; 2016, vol 1747, no 1.

  25. Bassinet C, Huet C, Derreumaux S, Brunet G, Chea M, Baumann M, et al. Small fields output factors measurements and correction factors determination for several detectors for a CyberKnife(R) and linear accelerators equipped with microMLC and circular cones. Med Phys. 2013;40:071725.

    Article  Google Scholar 

  26. Fuss M, Sturtewagen E, De Wagter C, Georg D. Dosimetric characterization of GafChromic EBT film and its implication on film dosimetry quality assurance. Phys Med Biol. 2007;52(14):4211–25.

    Article  Google Scholar 

  27. Martišíková M, Ackermann B, Jäkel O. Analysis of uncertainties in Gafchromic® EBT film dosimetry of photon beams. Phys Med Biol. 2008;53(24):7013–27.

    Article  Google Scholar 

  28. Klein DM, Tailor RC, Archambault L, Wang L, Therriault-Proulx F, Beddar AS. Measuring output factors of small fields formed by collimator jaws and multileaf collimator using plastic scintillation detectors. Med Phys. 2010;37(10):5541–9.

    Article  Google Scholar 

  29. Cheung T, Butson MJ, Peter KN. Independence of calibration curves for EBT Gafchromic films of the size of high-energy X-ray fields. Appl Radiat Isot. 2006;64(9):1027–30.

    Article  Google Scholar 

  30. Haryanto F, Fippel M, Laub W, Dohm O, Nüsslin F. Investigation of photon beam output factors for conformal radiation therapy—Monte Carlo simulations and measurements. Phys Med Biol. 2002;47(11):N133–43.

    Article  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Ankur Mourya.

Ethics declarations

Conflict of interest

No conflict of interest.

Ethical approval

This article does not contain any studies with human participants or animals performed by any of the authors.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Mourya, A., Aggarwal, L.M., Mandal, A. et al. An inexpensive method of small photon field dosimetry with EBT3 radiochromic film. Health Technol. 8, 171–177 (2018). https://doi.org/10.1007/s12553-018-0233-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s12553-018-0233-7

Keywords

Navigation