Initial clinical experience of N13-ammonia myocardial perfusion PET/CT using a compact superconducting production system

  • Brief Report
  • Published:
Journal of Nuclear Cardiology Aims and scope

Abstract

Background

Although N13-ammonia has favorable properties among FDA approved radiotracers, complexity of implementation has limited its use. We describe the initial patient experience of N13-ammonia PET imaging using a compact N13-ammonia production system.

Methods

N13 was produced using the ION-12SC, a 12MeV, 10uA superconducting minimally shielded cyclotron, and reduced to N13-ammonia in an automated multi-use purification unit. Patients were power injected with 9.3 ± 1.1 mCi (344.1 ± 40.7 MBq) of N13-ammonia for rest imaging, and 18.8 ± 0.9 mCi (695.6 ± 33.3 MBq) of N13-ammonia was injected at peak hyperemia for stress testing. Images were interpreted for relative perfusion, left ventricular volumes/function, blood flow quantification, and scored for image quality.

Results

In total 97 patients underwent 98 N13-ammonia PET scans (32 rest only/65 rest-stress/1 stress only). Image quality was 91.8% good or excellent. None were poor/non-diagnostic. Study durations were acceptable. Tracer related radiation dosimetry to patients was 0.7 ± 0.1 mSv (rest only), and 2.1 ± 0.1 mSv (rest-stress).

Conclusion

Clinical N13-ammonia production by the Ionetix ION-12SC delivers high quality myocardial PET perfusion images in a rapid protocol.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Figure 1
Figure 2
Figure 3
Figure 4

References

  1. Danad I, Raijmakers PG, Driessen RS, Leipsic J, Raju R, Naoum C, et al. Comparison of coronary CT angiography, SPECT, PET, and hybrid imaging for diagnosis of ischemic heart disease determined by fractional flow reserve. JAMA Cardiol. 2017;2:1100–7.

    Article  Google Scholar 

  2. Bateman T, Dilsizian V, Beanlands R, Depuey G, Heller G, Wolinsky D. American Society of Nuclear Cardiology and Society of Nuclear Medicine and Molecular Imaging joint position statement on the clinical indications for myocardial perfusion PET. J Nucl Med. 2016;57:1654–6.

    Article  Google Scholar 

  3. Murthy V, Bateman T, Beanlands R, Berman D, Bortes-Neto S, Chareonthaitawee P. Clinical quantification of myocardial blood flow using PET: Joint position paper of the SNMMI Cardiovascular Council and the ASNC. J Nucl Med. 2018;59:273–93.

    Article  CAS  Google Scholar 

  4. Herzog BA, Husmann L, Valenta I, Gaemperli O, Siegrist PT, Tay FM, et al. Long-term prognostic value of 13N-ammonia myocardial perfusion positron emission tomography added value of coronary flow reserve. J Am Coll Cardiol. 2009;54:150–6.

    Article  Google Scholar 

  5. Fiechter M, Ghadri JR, Gebhard C, Fuchs TA, Pazhenkottil AP, Nkoulou RN, et al. Diagnostic value of 13N-ammonia myocardial perfusion PET: added value of myocardial flow reserve. J Nucl Med. 2012;53:1230–4.

    Article  CAS  Google Scholar 

  6. Maddahi J, Packard RR. Cardiac PET perfusion tracers: current status and future directions. Semin Nucl Med. 2014;44:333–43.

    Article  Google Scholar 

  7. International Commission on Radiological Protection. Radiation dose to patients from radiopharmaceuticals. ICRP Publication 80. Ann ICRP 2000;28:113.

  8. Tamaki N, Yonekura Y, Senda M, Kureshi SA, Saji H, Kodama S, et al. Myocardial positron computed tomography with 13N-ammonia at rest and during exercise. Eur J Nucl Med Mol Imaging. 1985;11:246–51.

    Article  CAS  Google Scholar 

  9. Stabin MG. Radiopharmaceuticals for nuclear cardiology: Radiation dosimetry, uncertainties, and risk. J Nucl Med. 2008 Sep;49:1555–63.

    Article  CAS  Google Scholar 

Download references

Disclosures

Justin Pieper, Vaiibhav N. Patel, Sylvia Escolero, Jacob R. Nelson, David Hubers, Jill Rothley, and Richard L. Weinberg have no disclosures or conflicts of interest related to this publication. Edward P. Ficaro and James R. Corbett have financial interest in INVIA Medical Imaging Solutions, which licenses the commercial software used for imaging processing. Alexis Poitrasson-Rivière is employed by INVIA. INVIA Medical Imaging Solutions did not provide direct support to this study. Joseph Oliverio, Christopher K. Shreves, and Nick Freiburger own stock options and are employed by Ionetix Corporation. Venkatesh L. Murthy has received consulting fees and stock options from Ionetix, Inc., owns stock in General Electric and Cardinal Health, has a research Grant from Siemens Medical Imaging, and has provided expert witness testimony on behalf of Jubilant Draximage. Venkatesh L. Murthy is supported by 1R01HL136685 from the National, Heart, Lung, Blood Institute and 1R01AG059729 from the National Institute on Aging.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Justin Pieper MD.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

The authors of this article have provided a PowerPoint file, available for download at SpringerLink, which summarises the contents of the paper and is free for re-use at meetings and presentations. Search for the article DOI on SpringerLink.com.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary material 1 (PPTX 1092 kb)

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Pieper, J., Patel, V.N., Escolero, S. et al. Initial clinical experience of N13-ammonia myocardial perfusion PET/CT using a compact superconducting production system. J. Nucl. Cardiol. 28, 295–299 (2021). https://doi.org/10.1007/s12350-019-01886-7

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s12350-019-01886-7

Keywords

Navigation