A patient may have two contemporary hemoglobin test results; one received from a hematology analyzer and the other from a blood gas analyzer. Such results may pose a dilemma if validated without investigation. A 27-year-old female patient presented to the emergency department because of worsening dyspnea and hemoptysis. Her hemogram revealed Hb 4.5 g/dL; Hct 16.1% while blood gas analysis showed Hb 8.1 g/dL; Hct 25.3%. The results were evaluated to be suspicious. All results were rejected and new requests were made. Hb 9.8 g/dL; Hct 31.9% were seen in repeated hemogram while blood gas analysis showed Hb 10.2 g/dL; Hct 31.5%. It was understood that the nurse used the vein with an intravenous fluid for sampling the first hemogram. A multitude of factors affect the accuracy of hemoglobin concentration measurements. Literature suggests that hemoglobin values of hematology analyzers and blood gas analyzers should be in harmony. Any discordance must be an alarm to stop and review the test results before validation.
References
Whitehead RD Jr, Mei Z, Mapango C, Jefferds MED. Methods and analyzers for hemoglobin measurement in clinical laboratories and field settings. Ann N Y Acad Sci. 2019;1450:147–71.
Berkow L. Factors affecting hemoglobin measurement. J Clin Monit Comput. 2013;27:499–508.
Ray JG, Post JR, Hamielec C. Use of a rapid arterial blood gas analyzer to estimate blood hemoglobin concentration among critically ill adults. Crit Care. 2002;6:72–5.
Zhang JB, Lin J, Zhao XD. Analysis of bias in measurements of potassium, sodium and hemoglobin by an emergency department-based blood gas analyzer relative to hospital laboratory autoanalyzer results. PLoS One. 2015;10:e0122383.
Dolscheid-Pommerich RC, Dolscheid S, Grigutsch D, Stoffel-Wagner B, Graeff I. Comparability of point-of-comparability of point-of-care versus central laboratory hemoglobin determination in emergency patients at a supra-maximal care hospital. PLoS One. 2016;11:e0166521.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interests
The Author states that there are no conflicts of interests regarding the publication of this paper.
Ethical Approval
For the case report study the approval by the ethical committee is not required in our hospital.
Informed Consent
This case report involves information about a human.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
About this article
Cite this article
Aydın, Ö. Discordant Hemoglobin Values Determined by Blood Gas Analyzer and Hematology Analyzer Synchronously Is a Warning for Inappropriate Blood Sampling. Ind J Clin Biochem 37, 250–252 (2022). https://doi.org/10.1007/s12291-020-00938-x
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s12291-020-00938-x