Effects of different fat-suppression methods on T1 values in dynamic contrast-enhanced magnetic resonance imaging: a phantom study

  • Published:
Radiological Physics and Technology Aims and scope Submit manuscript

Abstract

Dynamic contrast-enhanced MRI may yield variable longitudinal-relaxation time (T1) values depending on the precision of the fat-suppression (FS) technique. This study aimed to investigate the influences of FS methods on T1 value measurements on phantoms containing test tubes filled with mixtures of five volumes of fat, six amounts of contrast agent, and water. Volumetric interpolated images were obtained using several FS methods and flip angles. T1 maps were created based on the variable flip angle approach. The T1 values of water obtained by point-resolved single-voxel spectroscopy (SVS) were used as reference values. Notably, FS methods were shown to have substantial effects on the measurement of T1 values. Among the tested FS methods, the Dixon (water) method produced T1 values most similar to SVS, which can be considered as a reference value for clinical practice.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6

References

  1. D’Orsi C, Sickles E, Mendelson E, Morris E. ACR BI-RADS Atlas, breast imaging reporting and data system. 5th ed. Reston: American College of Radiology; 2013.

    Google Scholar 

  2. Tofts P. T1-weighted DCE imaging concepts: modeling, acquisition and analysis. Siemens Magn Flash. 2010;3:31–9.

    Google Scholar 

  3. Profile: DCE MRI Quantification. version 1.6. RSNA 2011: 8-13, 35-46.

  4. Haase A, Frahm J, Hänicke W, Matthaei D. 1H NMR chemical shift selective (CHESS) imaging. Phys Med Biol. 1985;30(4):341–4.

    Article  CAS  PubMed  Google Scholar 

  5. Tannus A, Garwood M. Adiabatic pulses. NMR Biomed. 1997;10:423–43.

    Article  CAS  PubMed  Google Scholar 

  6. Dixon W. Simple proton spectroscopic imaging. Radiology. 1984;153:189–94.

    Article  CAS  PubMed  Google Scholar 

  7. Schick F, Forster J, Machann J, Huppert P, Claussen C. Highly selective water and fat imaging applying multislice sequences without sensitivity to B1 field inhomogeneities. Magn Reson Med. 1997;38:269–74.

    Article  CAS  PubMed  Google Scholar 

  8. Timothy R, Mikulis D. Neuro MR: principles. J Magn Reson. 2007;2(6.4):823–37.

    Google Scholar 

  9. Takezawa K, Goto M, Sakai K, Ikeno H, Nakatsukasa K, Imai H, and Yamada K. Influence of FS to evaluate T1 values in breast cancer: assessing the reliability of pharmacokinetic parameters[Abstract]. Proceedings of the 23th Annual Meeting of the international Society for Magnetic Resonance in Medicine, 30 May – 5 June 2015.

  10. Wang H, Riederer S, Lee J. Optimizing the precision in T1 relaxation estimation using limited flip angles. Magn Reson Med. 1987;5:399–416.

    Article  CAS  PubMed  Google Scholar 

  11. Chung S, Kim D, Breton E, Axel L. Rapid B1 + mapping using a preconditioning RF pulse with TurboFLASH readout. Magn Reson Med. 2010;64:439–46.

    PubMed  PubMed Central  Google Scholar 

  12. Hu HH, Nayak KS. Change in the proton T1 of fat and water in mixture. Magn Reson Med. 2010;63:494–501.

    Article  PubMed  PubMed Central  Google Scholar 

  13. Araki T. Complete explanation of MRI, 2nd Edition, Gakken Medical Shujunsya, Co; 2014. p. 239-247 (in Japanese).

  14. Brookes JA, Redpath TW, Gilbert FJ, Needham G, Murray AD. Measurement of spin-lattice relaxation times with FLASH for dynamic MRI of the breast. Br J Radiol. 1996;69:206–14.

    Article  CAS  PubMed  Google Scholar 

  15. Ren J, Dimitrov I, Sherry AD, Malloy CR. Composition of adipose tissue and marrow fat in humans by 1H NMR at 7 tesla. J Ripid Res. 2008;49:2055–62.

    CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Hiroyasu Ikeno.

Ethics declarations

Ethical approval

This article does not contain any studies or experiments performed by any of the authors on human participants or animals.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Ikeno, H., Sakai, K., Imai, H. et al. Effects of different fat-suppression methods on T1 values in dynamic contrast-enhanced magnetic resonance imaging: a phantom study. Radiol Phys Technol 12, 335–342 (2019). https://doi.org/10.1007/s12194-019-00521-x

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s12194-019-00521-x

Keywords

Navigation