Introduction

Aldehydes are volatile compounds which may indicate the deterioration of food, especially of fat rich foods as a result of processing, storage, or seed pre-treatment. These compounds are the main products of the oxidation of unsaturated fatty acids (Mandić et al. 2013). Hexanal, one of the most common aldehydes in food, was explored in several studies as the indicator of oxidation in oils (Azarbad and Jeleń 2014); however, the hydroperoxides decomposition in the process of autooxidation results in formation of many other aldehydes (alkanals, 2-alkenals, 2,4-alkadienals as the main ones), and also ketones, alcohols, and furanes, to name the most prevailing groups of compounds.

Volatile compounds, among them, volatile lipid oxidation product analysis is a challenging task because of their physical properties, food matrices they are released from and often their trace levels in food. The use of solvents allows for exhaustive extraction of volatile compounds; on the other hand, organic solvents may dilute or mask analytes during GC analysis (Guth and Grosch 1993). Employment of high temperature for the isolation of volatile compounds using distillation methods may result in their degradation, or in formation of artifacts. Headspace techniques such as static or dynamic headspace (SHS or DHS, respectively), headspace sorptive extraction (HSSE), solid-phase microextraction (SPME), and direct thermal desorption (DTD) are robust methods which do not require the use of a solvent or circumstances which impede obtaining a pure volatile compound extract of the analyzed sample (Bylaite and Meyer 2006; Cavalli et al. 2003). Each of these methods enables the isolation of volatile compounds usually without artifacts’ formation. Purge and trap (PT) method is used for volatile compounds’ isolation but, despite its high capacity, may adsorb water and cause artifacts formation during analysis (Manzini et al. 2011). HSSE and SPME are solvent-free methods based on various stationary phases. SPME invented in the beginning of 90-ties (Arthur and Pawliszyn 1990) became quickly one of most often sample preparation method used for volatile and semi-volatile compounds. Its simplicity of use and high range of available sorbents makes it a useful tool for volatiles and flavor compounds analysis in food (Jeleń et al. 2012). Nowadays, SPME has been used in many studies into volatiles in oils (Benincasa et al. 2003; Cecchi and Alfei 2013; Wei et al. 2015). HSSE is used for extracting volatile compounds from gaseous and aqueous phases (known as stir-bar sorptive extraction, SBSE) with the use of a stir-bar, usually coated with adsorbents such as polydimethylsiloxane (PDMS) (Sánchez-Rojas et al. 2009). The static and dynamic headspace extraction techniques are simple and robust methods, but—compared with other solventless methods—SHS has lower sensitivity (Wenzl and Lankmayr 2002). Despite its relatively low sensitivity, SHS with mass spectrometer has been used with satisfying results for differentiation between olive oils or detecting adulteration in virgin olive oils (Lorenzo et al. 2002; Marcos Lorenzo et al. 2002). Recent review by Sghaier and coworkers summarize different approaches to analysis of plant oil volatile compounds (Sghaier et al. 2016). Combining simplicity and robustness of static headspace analysis in its automated form with trapping volatiles on polymers in a single sampling unit may be the solution to overcome low sensitivity issues of SHS.

The aim of the study was to compare the extraction of volatile lipid oxidation products (VLOP) from rapeseed oil using static headspace method, in which typical loop transfer for compound injection was used, with a combination of headspace method in which compounds can be extracted multiple times from a headspace vial into a Tenax trap before transferring them into gas chromatograph.

Materials and Methods

Reagents and Standards

The following compounds were used to represent different classes of VLOP: alkanals—pentanal, hexanal, octanal, nonanal, decanal; 2-alkenals—2-pentenal, 2-hexenal, 2-heptenal, 2-nonenal; 2,4-alkadienals – 2,4-heptadienal and 2,4-decadienal; unsaturated alcohols—2-pentene-1-ol, 3-hexene-1-ol, 2-hexene-1-ol, 1-octene-3-ol; ketones—1-pentene-3-one, 6-methyl-5-hepten-2-one, 2-pentanone, 2-heptanone; and esters—3-hexenyl acetate, 3-hexyl acetate and also 2-pentylfurane. All standards were purchased at Sigma-Aldrich (Poznań, Poland) in the highest available purity (usually 98–99%, GC).

Sample Preparation and GC/MS Analysis

For method development, a mixture of volatile lipid oxidation products (VLOP) was prepared in refined deodorized rapeseed oil. Standards were prepared gravimetrically. To obtain working solutions, the stock solution was dissolved in fresh refined rapeseed oil to obtain 1 mg/L solution, which was used for subsequent analyses, when method parameters were evaluated and appropriate solutions were prepared for linearity and LOD testing.

Analyses were performed on GC/MS-TQ8030 (Shimadzu, Japan) triple quadrupole mass spectrometer system. The gas chromatograph (GC-2010 Plus) was equipped with a single split/splitless injection port and a Zebron ZB-5 column (30 m × 0.25 mm × 0.25 μm, Phenomenex). The GC/MS system was equipped with AOC-20i autoinjector. The GC/MS system was used in a single quadrupole mode with the first quadrupole set to transmission of all ions. All headspace samples were introduced into GC/MS by direct interface, using HS-20 Trap (Shimadzu, Japan) static headspace autosampler with adsorbent trap. The absorbent trap dimension was 2 mm (ID) × 100 mm and the trap was filled with Tenax TA (amount not specified). Parameters for the extraction of volatiles using HS-20 Trap were as follows: for injections in “loop” mode (static headspace)—sample line temp. 140 °C; transfer line 140 °C; pressure equilibrium time 0.1 min; load time 0.5 min; load equilibrium 0.1 min; injection time 0.5 min; GC cycle time 40 min. For “trap” mode, additionally, trap parameters were set as follows: trap cooling temp. −25 °C; trap equilibrium temp. −25 °C. Trap desorption temp. varied depending on the experiment (tested in a 200–300 °C range). Multi injection count for trap varied (1–10, depending on the experiment). For both modes, the following parameters were evaluated in method development: HS-20 Trap oven temperature, shaking level (1–5), equilibrium time (10–60 min), and injection time (0.5 and 1 min).

Results and Discussion

To develop and efficient extraction method using static headspace analysis basic parameters, such as extraction temperature, incubation time, shaking intensity, and injection parameters, were optimized to achieve maximal sensitivity. Static headspace autosampler used in these experiments used injection by loop, so the volatile compounds after equilibrium was established were transferred to the thermostated loop, from which they were swept by the carrier gas into gas chromatograph. Basic extraction parameters tested are shown in Fig. 1. In the graph, total peak areas of standards used in method elaboration are shown. Temperatures ranging from 40 to 90 °C were evaluated and the threefold increase was obtained when extraction at the lowest and the highest temperatures were compared. Because of the instability of the matrix and some standards, especially in regard to possible oxidation processes, for further experiments, extraction temperature of 70 °C was chosen. For this temperature, extraction time ranging from 10 to 60 min was tested. As shown in Fig. 1b, the increase was noted from 10 to 40-min extraction and the reproducibility was the highest at 40-min extraction, which was used for further analyses. Interestingly, the repeatability was the lowest at the shortest, but also the longest extraction times. As the migration of volatiles from the matrix is influenced by mixing, to facilitate fasted transfer to the headspace, shaking levels (alternative of mixing speed in autosampler) were tested. With the highest degree of shaking, the repeatability of peak areas was the highest. Therefore, the highest shaking level was used for subsequent analyses. Additionally, different combinations of pressurizing time and injection time were compared, to optimize injection process (results not shown). The highest total peak areas were obtained for pressurizing time 1 min and injection time 0.5 min; however, the values differed for particular compounds, especially with regard to pressurizing time.

Fig. 1
figure 1

Development of optimal parameters for HS extraction of VLOP using loop system. a Influence of temperature on the extraction of volatile compounds from oil (extraction time 30 min). b Influence of extraction time on the extraction of volatile compounds from oil (extraction temperature 90 °C). c Influence of shaking intensity on the extraction of volatile compounds from oil

The headspace autosampler was equipped with a Tenax A filled trap that can be cooled and multiple extractions can be performed from a single vial one, after one, without achieving equilibrium, which differs this method from multiple headspace extraction (MHE). Cooling the trap improves adsorption on the polymer and up to ten extractions can be performed for a single sample. Figure 2 illustrates the increase in peak areas of compounds when different extraction numbers were applied. In Fig. 2a, total peak areas are shown, whereas in Fig. 2b, peak areas of (E)-2-nonenal are presented as an example. Dramatic increase in peak areas can be noted when one extraction is compared to ten extractions. However, the high number of extractions despite trap cooling and high desorption temperatures may result in column overloading and decrease in peak symmetry and quality. Manipulating with split ratio improves peak shapes, but results in lower sensitivity, so the balance between these two factors has to be elaborated depending on the type of sample and compounds analyzed. For our set of compounds, the optimal number of extractions that did not degrade significantly chromatographic separation was seven, and it was used for method elaboration. Higher number of extractions ended in serious peak overlapping. For all extractions using trap system, trap temperature was kept at −25 °C and different desorption temperatures were tested (200, 250, 275, and 300 °C). The highest total peak area was observed for desorption temperature of 275 and 300 °C (detailed results not shown). The peak areas of desorbed compounds at different temperatures differed substantially. For example, hexanal peak area after desorption at 300 °C was 133% of that at temperature of 200 °C, but the peak area of 2,4-decadienal desorbed at 300 °C was 433% of the peak area of that compound desorbed at 200 °C. The carry-over for 2,4-decadienal was the highest among all tested compounds. To minimize carry-over desorption temperature of 300 °C was chosen for the analyses. Table 1 shows basic performance parameters for both methods. For “loop” injection mode, both the reproducibility and carry-over are higher than for “trap” method for majority (16/20) compounds. Linearity of both methods which were tested in a range of 0.1–20 or 0.1–50 mg/L was comparable. The highest difference was noted in the LOQ values and sensitivity of both methods. Gain in sensitivity can be seen in Fig. 3, where standard curves for hexanal and (E)-2-nonenal are shown. Dramatic differences indicate the potential of trapping volatiles in the headspace for determination of these volatiles. Similar differences were noted for the remaining compounds. When differences between LOQ for “loop” and “trap” methods are compared (Table 1), the latter one provides from 10 to 33 times lower LOQs depending on a compound. The limits of quantitation for lipid oxidation products are very low, regarding their presence in oils, and can compete with LOQs achieved using SPME. A potential advantage of “trap” method is the high linear range, which was for the majority of compounds in a 0.1–20 mg/L range, whereas in SPME sampling usually the linear range is often limited, especially when fibers based on polymers are used (carboxene/divinylbenzene or PDMS, carboxene/PDMS), which are very often selected to obtain low limits of detection. Linear range for “trap” method was lower than “loop” for majority of compounds (for 12 out of 20 the linear range was 0.1–20 mg/L), but this was related to adsorption of the analytes on limited sites (pores) of Tenax surface. Nevertheless, the Tenax surface is much higher than for microextraction methods; therefore, the linear range was satisfactory.

Fig. 2
figure 2

Changes in peak areas as a result of number of extraction performer using headspace sampler in a “trap” mode. a Total peak area for all compounds used for method development. b Peak areas of (E)-2-nonenal

Table 1 Basic characteristics for methods using static headspace extraction (loop) and static headspace extraction with compounds trapping on Tenax trap (trap)
Fig. 3
figure 3

Calibration curves for static headspace using “loop” and “trap” approaches (with seven extractions from a single vial) for two aldehydes. a Hexanal. b (E)-2-nonenal

Static headspace analysis of volatile compounds was used as a method for monitoring oxidative stability of various oils in which total volatiles peak areas were measured (Warner et al. 1989). Automated static headspace is a common extraction technique in oil production plants, where it is used for both the residue extraction solvent analysis and volatile compounds in oils. In comparative studies on volatile compounds in oils, static headspace, along with dynamic headspace and direct injection, was used when inter-laboratory collaborative studies were performed to monitor, among other parameters, pentane and hexanal in oils (Warner and Nelsen 1996). Authors noted lower area counts for static headspace compared to dynamic headspace methods, but stated that static headspace can be a method capable of determination of oxidation level. Snyder et al. 1988 compared static headspace and purge and trap noting the much higher proportions of 2,4-decadienal and 2,4-heptadienal in trapping technique compared to static headspace. However, the static headspace produced higher proportions of low molecular compounds. Indeed, the advantage of static headspace technique is its high sensitivity for low-boiling compounds. This is due to the relation of vapor pressure of compounds in the homolog series to the number of carbon atoms in the molecule. The use of SPME for volatile compounds of plant oils has been verified in many works and it proved to be a sensitive method by which tens of compounds could be monitored in fresh and oxidized oils. Limits of detection in low micrograms per liter can be achieved (Jeleń et al. 2000). SPME has been proposed as a method for analysis of potential markers for oxidation status (nonanal proposed as indicator) (Vichi et al. 2003), or quality—freshness markers of olive oil ((E)-hex-2-enal, C6 alcohols and C5 ketones) (Cavalli et al. 2004), although in many papers describing its use in flavor analysis, it is based mainly on area count comparison, not a quantitative method (Jeleń et al. 2012). A comparative study was performed in which static headspace was compared to headspace solid-phase microextraction, headspace sorptive extraction and direct thermal desorption (Cavalli et al. 2003). Compared to sorptive methods, static headspace showed very low sensitivity and was concluded to be not suitable for characterization of olive oil volatile compounds for that reason.

Presented in this paper, combination of headspace with compounds trapping can dramatically increase the sensitivity and therefore contribute to its use also for low levels of volatile compounds to be quantified.

Conclusions

Presented comparison shows that combining static automated headspace analysis, where a certain volume of headspace is transferred to the gas chromatograph without sample enrichment, with multiple extraction and trapping volatiles before their injection results in significant improvement of method sensitivity. As the trapping system is located in the automated static headspace sampler, it provides a relatively simple from technical point of view and serious from method performance point of view enhancement of its analytical abilities. Combining static headspace with compounds trapping overcomes the most severe static headspace drawback—low sensitivity, maintaining high linearity and its robustness and sample preparation simplicity. To minimize the risk of column overloading and carry-over issues, the method needs compromise in number of extractions and desorption temperatures, which depend highly on the analyte character and the matrix.