Introduction

To achieve normal kinetic function of the knee, balanced flexion gap, and appropriate patellar tracking, rotation of the femoral component during TKA has been emphasized [30, 3639]. There are two methods to determine femoral component rotation in the axial plane: the measured resection technique using the bony landmark and the gap balancing technique [8, 10, 11, 16]. The transepicondylar axis (TEA), AP axis, and posterior condylar axis (PCA), which are references used during the measured resection technique, are widely accepted and used [16, 19]. Advocates of the gap balancing technique are concerned with low reproducibility and reliability of the bony landmarks for the measured resection [11, 20, 22].

However, a surgeon using the gap balancing technique-based TKA occasionally might encounter excessive internal rotation of the femoral component, which can cause postoperative patellofemoral complications, and should be avoided [14, 36]. With the gap balancing technique, excessive medial release and varus proximal tibial resection can be associated with internal rotation of the femoral component (Fig. 1) [6, 8, 11, 14, 18, 25, 38]. Although a rectangular flexion gap can be achieved in these circumstances, resultant kinematic alignment of the distal femur may be abnormal, especially in the patellofemoral joint. Previous studies have evaluated the factors associated with femoral component rotational alignment with a separate factor analysis using unadjusted statistical methods [6, 11, 14, 25]. However, without controlling for relevant confounding variables such as extent of medial release, varus proximal tibial resection, and mechanical varus alignment of the knee, it is difficult to identify the strengths of the associations between the factors of interest and femoral component rotation, or whether some of the observed associations were the result of confounding variables.

Fig. 1A–C
figure 1

The possible causes of an internally rotated femoral component are shown. (A) In the varus knee, some medial release frequently is required. Adequate medial release results in a rectangular mediolateral flexion gap, which is parallel to the transepicondylar axis. (B) An overly performed medial release will result in an internally rotated femoral component relative to the transepicondylar axis. (C) Inadvertent varus resection of the proximal tibia will cause internal rotation of the femoral component.

In the current study we asked: (1) What pre- and intraoperative factors are independently associated with internal rotation of the femoral component in TKA using the gap balancing technique? (2) To what degree does femoral component rotation as defined by the navigation system differ from rotation as measured by a postoperative CT?

Patients and Methods

Between October 2007 and June 2014, we performed 712 primary TKAs. During that time, 655 knees underwent primary TKA with navigation (92.0%) (Fig. 2). In all cases, primary TKAs were performed with navigation, except when the navigation system was malfunctioning or unavailable. Inclusion criteria were as follows: the knees with symptomatic degenerative osteoarthritis receiving primary TKA using gap balancing technique-based navigation; varus or mild valgus aligned knees that require medial soft tissue release; and preoperative radiographs including CT. Preoperative CT was performed to identify the original rotational alignment of the femur and tibia. For 61 knees, preoperative CT was declined by the patients, therefore CT was not performed in those patients’ knees. In 10 knees, preoperative CT was insufficient for evaluation and these were excluded from the study (Fig. 2). All procedures were performed by the senior author (H-CL). The patients’ medical records and radiographs of their knees were reviewed retrospectively. Patients without data regarding computer-assisted surgery, patients who underwent revision TKA, those with inadequate preoperative evaluation (radiography and CT), and those with valgus knees not requiring medial release were excluded. Primary TKAs for inflammatory arthritis such as rheumatoid arthritis also were excluded (Fig. 2). In cases of bilateral TKAs, only the first surgically treated knee was included. After application of these criteria, 377 knees were included for our first research question regarding univariable and multivariable analysis of femoral component internal rotation (Table 1).

Fig. 2
figure 2

The study flow diagram is shown. RA = rheumatoid arthritis.

Table 1 Case summary (n = 377)

According to femoral component rotation directed by the navigation system to achieve a rectangular flexion gap, the entire cohort was divided in two groups: one group with an internally rotated femoral component (Group IR, cases) and one group with a neutrally or an externally rotated femoral component (Group non-IR, controls). To analyze what degree femoral component rotation, as defined by the navigation system, differs from rotation as measured by postoperative CT, a total of 88 knees were included. Postoperative CTs were performed in 88 patients (131 knees) with unexplained pain or stiffness after surgery to identify malrotation of the components. In bilateral TKAs, the first 43 knees that underwent TKA were selected. The 88 knees included seemed representative of the larger population in that the variables we analyzed (Table 1) were not different in those of the entire population (377 knees) (Appendix 1. Supplemental material is available with the online version of CORR ®).

The current study was approved by the ethics committee of Korea University Guro Hospital (institutional review board approval number: KUGH14148).

The standard medial parapatellar approach was used. After exposing the knee, medial osteophytes were removed and tibial insertions of the deep medial collateral ligament (MCL) and posterior oblique ligament were released along the medial meniscus, 3 to 5 mm below the medial joint line. Further steps of medial release, if performed, are described later. A computer navigation system (OrthoPilot® navigation system; B. Braun Aesculap®, Tuttlingen, Germany) was used. Gap balancing-based computer navigation was used for primary TKA. Resection of the tibia was performed before femoral resection. Detailed surgical steps and manipulation of the navigation system were performed as recommended by the manufacturer. The prostheses implanted, e.motion® (233 knees), Columbus® (81 knees), and e.motion® Pro (63 knees) (B. Braun Aesculap®), varied according to time when the TKA was performed. The first implant was used for the first 47 months of the study, and the second implant was used for the next 19 months. The last implant was used for the final 14 months of the study. The types of implants were determined according to the condition of the PCL and the feasibility of a balanced extension-flexion gap. For example, we used a PCL-substituting implant when the PCL showed severe degeneration or laxity, or when the flexion gap was too tight compared with the extension gap. A cruciate retaining type was used in 231 knees (61%), and posterior substituting type was used in 146 knees (39%).

After registration of the bony landmarks, the proximal tibial resection was performed (Fig. 3). The varus-valgus angle of the resected tibia was checked (Fig. 3A). After proximal tibial resection, extension and flexion gaps were measured using a gap tensioning device. If it was decided that the PCL should be resected because of its consistency and the chosen implant, then resection was performed before gap measurement. With these gap data, the femoral resection planning step was performed (Fig. 3B). To obtain a rectangular space for the extension and flexion gaps, greater than 2 mm mediolateral asymmetry was not permitted according to our navigation data. When further medial release (mediolateral asymmetry > 2 mm) was needed according to the navigation gap data, stepwise medial release was performed as follows: Step 1, removal of medial osteophytes and release of tibial insertions of the deep MCL and posterior oblique ligament (this step was performed before the initial femoral planning step); Step 2, partial release of the tibial insertion of the superficial MCL above the level of the pes anserine tendon and direct arm of the semimembranosus tendon; and Step 3, near complete release of the tibial insertion of the superficial MCL below the level of the pes anserine tendon (Fig. 4) [3, 23, 24, 31, 39]. In the case of multiple medial release steps, gap checking and femoral planning were repeated before the femoral AP resection.

Fig. 3A–C
figure 3

Important navigation procedures are shown. (A) After registration of bony landmarks and hip, knee, and ankle centers, the proximal tibial resection was performed. Varus-valgus angle of the resected tibia was checked (asterisk). (B) In the femoral planning step, to achieve a rectangular mediolateral gap in knee extension and flexion, varus-valgus angle and rotation (dagger) of the femoral component are adjusted with navigation dictation. The numbers displayed next to femurs and tibias represent expected amounts of bone resection. (C) After distal femoral resection, the rotational position of the AP femoral cutting jig on the distal femur (double dagger), which is positioned manually, is displayed in real time. This represents actual femoral component rotation. In this case, the femoral component is in 2º external rotation relative to the posterior condylar axis. Green = this value is as planned; Ext. = extension.

Fig. 4
figure 4

The medial release steps are shown. Removal of medial osteophytes and release of tibial insertions of the deep MCL and posterior oblique ligament are performed in Step 1. If further medial release was required, the proximal insertion of the superficial MCL (above the pes anserine tendon) and anterior arm of the semimembranosus were released (Step 2). In the final step, Step 3, the distal insertion of the superficial MCL (below pes anserine tendon) was released. MCL = medial collateral ligament.

In addition to medial and lateral gaps of extension and flexion, femoral component rotation directed by the computer was recorded when the symmetric mediolateral space was achieved (Fig. 3B). The navigation system used in the current study measures rotation of the femoral component based on the PCA. According to the planned femoral component rotation, the entire cohort was divided into cases and controls. As planned in this step, distal, anterior, and posterior femoral resections were performed. However, there can be a slight difference between the planned femoral component rotation and actual rotational position of the femoral cutting jig owing to possible error during manual placement of the jig. Therefore, jig rotation measured just before AP resection of the femur was recorded (actual femoral component rotation) (Fig. 3C).

Using standing AP and lateral radiographs, pre- and postoperative hip-knee-ankle angle and preoperative medial proximal tibial angle were measured. A positive pes anserine tendon value indicates a varus knee. Using the preoperative CT of the knee, the amount of rotation of the clinical TEA relative to the PCA (TEA–PCA) was measured (Fig. 5A) [38]. A positive TEA–PCA value indicates external rotation of the TEA relative to the PCA. Postoperative CT was performed in 88 knees (Fig. 5B). All postoperative radiologic evaluations were performed 1 to 2 weeks after surgery. In these knees, postoperative TEA–PCA, which was measured using CT, was compared with the theoretical postoperative TEA–PCA. Theoretical postoperative TEA–PCA was calculated as: preoperative TEA–PCA minus actual femoral component rotation (Fig. 5C). For example, with 5º external rotation for preoperative TEA–PCA and 2º external rotation for actual femoral component rotation, the theoretical postoperative TEA–PCA is 3º of external rotation. If the theoretical postoperative TEA–PCA is significantly different from the postoperative TEA–PCA, which was measured using postoperative CT, then this indicates an error during the navigation procedure. Compared with postoperative TEA–PCA, theoretical postoperative TEA–PCA not within a tolerance level of ± 3º was considered an outlier [13, 26, 29, 41]. All radiologic measurements were performed twice during a 2-week period by two orthopaedic surgeons (SYL and KMJ).

Fig. 5A–C
figure 5

Comparisons between the theoretical postoperative TEA–PCA and postoperative TEA–PCA measured using postoperative CT are shown. (A) Preoperative TEA–PCA is measured using preoperative CT, and (B) postoperative TEA–PCA is measured using postoperative CT. (C) Theoretical postoperative TEA–PCA is calculated by subtracting the preoperative TEA–PCA from the rotational position of the AP femoral cutting jig on the distal femur (actual femoral component rotation [F-Rot]; Fig. 3C). For example, with 5º external rotation for preoperative TEA–PCA and 2º external rotation for actual femoral component rotation, the theoretical postoperative TEA–PCA is 3º external rotation. If the theoretical postoperative TEA–PCA is significantly different using the postoperative TEA–PCA, which was measured using postoperative CT, then this indicates an error during the navigation procedure. TEA = transepicondylar axis; PCA = posterior condylar axis; Ext. = extension.

Statistical Analysis

The reliability of radiologic measurements was evaluated using intraclass correlation coefficients (ICCs) [1]. Comparisons of gaps and their changes according to each compartment and extension-flexion were performed using one-way ANOVA testing followed by Tukey’s post hoc test. The differences in these analyses were considered significant at a probability of 0.05 or less.

The variables of Group IR and Group non-IR were compared using an independent t-test, Fisher’s exact test, and linear-by-linear association analysis. Cases were defined as having internal rotation of the planned femoral component rotation. Univariable analysis of the variables (age, sex, BMI, surgically treated side, preoperative hip-knee-ankle angle, preoperative medial proximal tibial angle, preoperative TEA–PCA, PCL resection, type of prosthesis, coronal angle of the resected tibia, and extent of medial release) of cases (Group IR) and controls (Group non-IR) was performed. Only covariates with a resulting p value of 0.15 or less were included in the backward stepwise multivariable logistic regression analysis. The number of cases (Group IR) in the current study was 39 (prevalence, 10.3%); two variables were included in the multivariable logistic regression analysis. The required sample size was approximately 194 (2 × 10/0.103) [35]. Regarding multicollinearity between the variables included in the multivariable analysis, a variance inflation factor greater than 5 was considered indicative of collinearity [15]. To analyze the navigation error in the axial plane, theoretical postoperative TEA–PCA and postoperative TEA–PCA measured using postoperative CT were compared using the paired t-test and Pearson’s correlation analysis. A probability of 0.05 or less indicated significance. Statistical analyses were performed using SPSS, Version 20.0 (IBM Corp, Armonk, NY, USA).

All inter- and intraobserver ICCs showed excellent agreement regarding radiographic measurement reliability (Table 2). As progressively larger releases were performed, we found the medial gaps in the flexed knee increased more than the medial gap in the extended knee, which was greater than the lateral gaps in the flexed or extended knee (Table 3). In the demographic comparison between two groups, the mean planned femoral component rotation of Group IR and Group non-IR were −2.0° ± 1.1º and 1. 4° ± 1. 6º, respectively. The preoperative hip-knee-ankle angle of Group IR was greater than that of Group non-IR. The proportions of the larger medial release steps (Steps 2 and 3) were higher in Group IR than Group non-IR (Table 4). Univariable factor analysis showed that the preoperative hip-knee-ankle angle (odds ratio [OR], 1.1; 95% CI, 1.0–1.1; p = 0.037) and extent of medial release (Step 1 as reference; Step 2: OR, 5.9, [95% CI, 2.3–15]; Step 3: OR, 23, [95% CI, 8.4–66], p < 0.001) were selected for multivariable analysis (p ≤ 0.15) (Table 5).

Table 2 Intraclass correlation coefficients for radiologic evaluations
Table 3 Comparisons of gaps and their changes according to each compartment and extension-flexion (n = 377)
Table 4 Comparison according to femoral component rotation directed by the navigation system to obtain a rectangular mediolateral flexion gap
Table 5 Univariable analysis* of factors associated with femoral component internal rotation directed by the navigation system to obtain a rectangular mediolateral flexion gap (n = 377)

Results

After controlling for relevant confounding variables such as the preoperative hip-knee-ankle angle, preoperative TEA–PCA, and coronal angle of the resected tibia, we found that the extent of medial release (Step 1 as reference; Step 2: OR, 5.7, [95% CI, 2.2–15); Step 3: OR, 22, [95% CI, 7.8–62], p < 0.001) was the only factor we identified that was associated with internal rotation of the femoral component (Table 6). All variance inflation factors of these covariates showed values less than 5, which indicates that there is no serious multicollinearity between the covariates.

Table 6 Multivariable analysis* of factors associated with femoral component internal rotation

With the numbers available, we found no difference between the mean theoretical postoperative TEA–PCA and the postoperative TEA–PCA measured using postoperative CT (4.8° ± 2.7º versus 5.0° ± 2.3º; mean difference, 0.2° ± 1.5º; p = 0.160). There was a strong positive correlation between the theoretical and postoperative values of TEA–PCA (r = 0.838; p < 0.001), meaning that as theoretical TEA–PCA increased, so did postoperative TEA–PCA. Among the 88 knees, only three (3.4%) showed a difference exceeding ± 3º for the theoretical and actual postoperative TEA–PCA values.

Discussion

Because excessive internal rotation of the femoral component can be associated with patellofemoral tilt, subluxation, and dislocation, it is important to achieve correct rotational alignment of the femoral component [14, 36]. Although studies identified several factors, such as excessive medial release and varus proximal tibial resection, that were associated with femoral component internal rotation during gap balancing technique-based TKA, the studies evaluated the factors using unadjusted statistical methods [6, 11, 14, 25]. Because this approach does not take into account the influence of possible confounding variables, simple comparisons without adjustment of this sort are potentially misleading. To overcome this problem, we performed multivariable analysis, controlling for relevant confounding variables such as extent of medial release, varus proximal tibial resection, and mechanical varus alignment of the knee. The most-important finding of the current study is that the extent of medial release is the only factor we identified that was associated with internal rotation of the femoral component in TKAs performed using the gap balancing technique. We found that a near-complete release of the superficial MCL tibial insertion below the level of the pes anserine tendon was consistently associated with internal rotation of the femoral component. Once complete release of the superficial MCL has been performed, it cannot be undone to overcome internal rotation of the femoral component, therefore a carefully subdivided medial-releasing technique, particularly for the superficial MCL seems important to prevent internal rotation of the femoral component. Calculating with femoral rotation profiles of navigation records, preoperative and postoperative CT, errors during navigation referencing were not suspected.

Our study has several limitations. The first is patient selection. Although we tried to use navigation whenever possible except when the system was not working or available, there may be selection bias owing to the patients who did not receive this technology. For the second question of the current study, only a limited number of patients underwent postoperative CT. Although these patients were not substantially different from the entire population, patients who had postoperative CT underwent additional imaging because of unexplained pain or stiffness after the arthroplasty. We could not ethically perform postoperative CT in all patients, therefore we limited it to patients with a notable problem, and an accurate rotation profile was needed. With this limited number of cases, we could not integrate the navigation accuracy to the covariate of multivariable analysis. Second, there could be some overlap among the medial release steps because there is no definite boundary between medial soft tissue structures. However, this possibility seems to be minimal because the medial release technique used in our study was simplified to three steps compared with that used in previous studies [17, 31]. Although there are numerous medial release techniques that differ in sequence and structures released, Hunt et al. [17] noted that there is a lack of evidence supporting current methods. Furthermore, it is difficult to distinguish each medial fiber structure during surgery, therefore alternative medial release methods other than the traditional medial subperiosteal release technique have been proposed [2, 33]. In addition, preoperative medial ligamentous laxity or contracture, which can affect medial gaps, was not evaluated. Third, there could be another selection bias in type of implant and in PCL resection. Although both covariates showed little association with femoral component internal rotation in our analysis, the decision regarding PCL resection and implant type were made by the surgeon subjectively, and by time sequence (the implants were used sequentially during a period of years), respectively. Future randomized studies of these covariates may be required to decrease the issue of selection bias.

Numerous studies have shown a relationship between medial release and increment of the medial tibiofemoral gap [5, 6, 21, 27, 31, 42]. In the current study, gap increment according to each compartment showed that the medial flexion gap was the most-increased compartment, followed by the medial extension gap, as a progressively larger release was done. This finding was consistent with those of previous studies [5, 28, 43]. Excessive internal or external rotation of the femoral component could happen during the gap balancing technique, which uses the tibial cut surface as a reference for femoral component rotation [11, 14, 25, 32, 34, 38]. Too much release of the medial structures may result in symptomatic instability, use of very thick polyethylene bearings, or even constrained implants to address the excessive medial gap [40]. Thicker polyethylene bearings and constrained implants are associated with a higher risk of loosening and joint-line elevation [4, 39, 40]. In addition, uneven load distribution and early wear on the bearing could result [12]. Heesterbeek et al. [14] indicated that excessive medial release may cause less external rotation of the femoral component. However, they used a gap tensioning device and goniometer, which may be less precise than the navigation system we used, and their study included only 32 knees with medial releases. Although Christensen et al. [6] showed more medial release results with less external rotation of the femoral component, no detailed gap data corresponding to each release step were provided in their report, and in their study, soft tissue release was the only variable considered to affect the femoral component rotation. Dennis et al. [11] noted that excessive varus tibial resection greater than 3º is one factor associated with an internally implanted femoral component in TKA with the gap balancing technique. In the current study, the coronal angle (varus–valgus) of the resected tibia did not seem to be associated with internal rotation of the femoral component. The mean coronal angle of the resected tibia was almost neutral, and there were no outliers exceeding ± 3º [13, 26, 29, 41]. This may be because of the small variability in the coronal angle of the resected tibia. We suspect that excessive varus tibial resections are not common, with or without computer assistance [8, 31, 39, 42].

In the current study, navigation data regarding femoral component rotational alignment showed high accuracy compared with that of expected rotational alignment of the femoral component measured by pre- and postoperative CT scans. This finding is consistent with those of previous studies [13, 41]. However, those studies were based on the comparison between conventional and navigated TKAs. In those studies, although numerous outliers were consistently smaller in the navigated group than in the conventional group, it is difficult to conclude that intraoperative navigation data were accurate when compared with expected rotational profile measure with pre- and postoperative CT scans. Two recent studies compared the rotational profile of navigation data with that of postoperative CT, which was similar to our method [7, 44]. However, Dahabreh et al. [7] reported that mismatch exceeding 3° between postoperative CT and navigation measurements was 22%, which was larger than ours (3.4%). We suspect that difference between the two studies occurred because they used the measured resection-based navigation TKA (medial and lateral epicondylar referencing), which was different from ours (gap balancing-based technique, the most anterior and posterior points of the femur referencing). Dahabreh et al. [7] stated that landmark (medial and lateral epicondyles) registration errors are possible. This is indirect evidence of low reproducibility and reliability of bony landmarks for the measured resection, which others have commented on [9, 11, 20, 22]. In addition, the number of cases in the study of Zambianchi et al. [44] was relatively small.

After considering preoperative coronal and axial alignment, PCL resection, coronal angle of the resected tibia, and extent of the medial release as possible associated factors, the extent of the medial release was the only factor we found that was associated with internal rotation of the femoral component in a gap balancing TKA. To avoid internal rotation of the femoral component, we recommend a carefully subdivided medial-releasing technique, especially for the superficial MCL, because once the superficial MCL has been completely released it cannot easily be restored. Further studies on safe and reproducible medial release techniques are needed.