Skip to main content

Advertisement

Log in

Reduced Articular Surface of One-piece Cups: A Cause of Runaway Wear and Early Failure

  • Symposium: Complications of Hip Arthroplasty
  • Published:
Clinical Orthopaedics and Related Research®

Abstract

Background

Despite the clinical success of modern metal-on-metal articulations, concerns with wear-related release of metal ions persist. Evidence suggests metal ion release is related to the effective coverage of the head in the metal shell (the cup’s functional articular arc). A recent study suggests a reduced functional articular arc is associated with increased ion release and the arc is a function of component design, size, and the abduction angle.

Questions/purposes

The purposes of this study were to (1) measure the functional articular arc in different sizes of currently available one-piece metal shells from several different manufacturers; and (2) compare the functional articular arc of these one-piece metal shells with the 180º arc of conventional hip arthroplasty acetabular components.

Methods

We calculated the available articular surface arc for 33 one-piece metal cups using measurements of cup depth and internal cup radius.

Results

The arc of the articular surface varied among manufacturers and generally decreased with decreasing shell diameter. The mean functional articular arc was 160.5° ± 3.6° (range, 151.8°–165.8°), which was less than the 180° arc of a conventional acetabular component.

Conclusions

Our data show certain cup designs are at higher risk for failure as a result of the decreased articular surface arc. This, along with analysis of abduction angles, supports the recent findings of bearing failure with vertically placed implants. Care must be taken when implanting these shells to ensure they are placed in less abduction to avoid edge loading and the potential for early bearing failure.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Subscribe and save

Springer+ Basic
$34.99 /Month
  • Get 10 units per month
  • Download Article/Chapter or eBook
  • 1 Unit = 1 Article or 1 Chapter
  • Cancel anytime
Subscribe now

Buy Now

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2

Similar content being viewed by others

References

  1. Bhamra M, Case C. Biological effects of metal-on-metal hip replacements. Proc Inst Mech Eng H. 2006;220:379–384.

    CAS  PubMed  Google Scholar 

  2. Brown C, Fisher J, Ingham E. Biological effects of clinically relevant wear particles from metal-on-metal hip prostheses. Proc Inst Mech Eng H. 2006;220:355–369.

    CAS  PubMed  Google Scholar 

  3. Buergi ML, Walter WL. Hip resurfacing arthroplasty: the Australian experience. J Arthroplasty. 2007;22(Suppl 3):61–65.

    Article  PubMed  Google Scholar 

  4. De Haan R, Campbell PA, Su E, De Smet K. Revision of metal-on-metal resurfacing arthroplasty of the hip. J Bone Joint Surg Br. 2008;90:1158–1163.

    Article  PubMed  Google Scholar 

  5. De Haan R, Pattyn C, Gill HS, Murray DW, Campbell PA, De Smet K. Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement. J Bone Joint Surg Br. 2008;90:1291–1297.

    Article  PubMed  Google Scholar 

  6. Dowson D. Tribological principles in metal-on-metal hip joint design. Proc Inst Mech Eng H. 2006;220:161–171.

    CAS  PubMed  Google Scholar 

  7. Dowson D, Jin Z-M. Metal-on-metal hip joint tribology. Proc Inst Mech Eng H. 2006;220:107–118.

    Article  CAS  PubMed  Google Scholar 

  8. Grammatopolous G, Pandit H, Kwon YM, Gundle R, McLardy-Smith P, Beard DJ, Murray DW, Gill HS. Hip resurfacings revised for inflammatory pseudotumour have a poor outcome. J Bone Joint Surg Br. 2009;91:1019–1024.

    Article  CAS  PubMed  Google Scholar 

  9. Isaac G, Thompson J, Williams S, Fisher J. Metal-on-metal bearing surfaces: materials, manufacture, design, optimization, and alternatives. Proc Inst Mech Eng H. 2006;220:119–133.

    CAS  PubMed  Google Scholar 

  10. Jacobs JJ, Skipor AK, Doorn PF, Campbell P, Schmalzried TP, Black J, Amstutz HC. Cobalt and chromium concentrations in patients with metal on metal total hip replacements. Clin Orthop Relat Res. 1996;329(Suppl):S256–S263.

    Article  PubMed  Google Scholar 

  11. Khan M, Kuiper J-H, Richardson J. The exercise related rise in plasma cobalt levels after metal-on-metal hip resurfacing arthroplasty. J Bone Joint Surg Br. 2008;90:1152–1157.

    Article  CAS  PubMed  Google Scholar 

  12. Langton DJ, Jameson SS, Joyce TJ, Hallab NJ, Natu S, Nargol AV. Early failure of metal-on-metal bearings in hip resurfacing and large-diameter total hip replacement: a consequence of excess wear. J Bone Joint Surg Br. 2010;92:38–46.

    Article  CAS  PubMed  Google Scholar 

  13. Langton DJ, Jameson SS, Joyce TJ, Webb J, Nargol AV. The effect of component size and orientation on the concentrations of metal ions after resurfacing arthroplasty of the hip. J Bone Joint Surg Br. 2008;90:1143–1151.

    CAS  PubMed  Google Scholar 

  14. Lee R, Essner A, Wang A. Tribological considerations in primary and revision metal-on-metal arthroplasty. J Bone Joint Surg Am. 2008;90(Suppl 3):118–124.

    Article  PubMed  Google Scholar 

  15. Morlock M, Bishop N, Zustin J, Hahn M, Ruther W, Amling M. Modes of implant failure after hip resurfacing: morphological and wear analysis of 267 retrieval specimens. J Bone Joint Surg Am. 2008;90(Suppl 3):89–95.

    Article  PubMed  Google Scholar 

  16. Ollivere B, Darrah C, Barker T, Nolan J, Porteous MJ. Early clinical failure of the Birmingham metal-on-metal hip resurfacing is associated with metallosis and soft-tissue necrosis. J Bone Joint Surg Br. 2009;91:1025–1030.

    Article  CAS  PubMed  Google Scholar 

  17. Pandit H, Glyn-Jones S, McLardy-Smith P, Gundle R, Whitwell D, Gibbons C, Ostlere S, Athanasou N, Gill H, Murray D. Pseudotumors associated with metal-on-metal hip resurfacings. J Bone Joint Surg Br. 2008;90:847–851.

    Article  CAS  PubMed  Google Scholar 

  18. Schmalzreid TP. Why total hip resurfacing. J Arthroplasty. 2007;22(Suppl 3):57–60.

    Article  Google Scholar 

  19. Williams S, Leslie I, Isaac G, Jin Z, Ingham E, Fisher J. Tribology and wear of metal-on-metal hip prostheses: influence of cup angle and head position. J Bone Joint Surg Am. 2008;90(Suppl 3):111–117.

    Article  PubMed  Google Scholar 

Download references

Acknowledgments

We thank Christi Sychterz Terefenko, MS, for her assistance with the manuscript.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to William L. Griffin MD.

Additional information

One or more of the authors (WG, BS, TF) received funding from research endowments from The Winkler Fund and The Smith Arthritis Fund.

This work was performed at The OrthoCarolina Hip and Knee Center in conjunction with The OrthoCarolina Research Institute and the Department of Mechanical Engineering, University of North Carolina–Charlotte, Charlotte, NC, USA.

About this article

Cite this article

Griffin, W.L., Nanson, C.J., Springer, B.D. et al. Reduced Articular Surface of One-piece Cups: A Cause of Runaway Wear and Early Failure. Clin Orthop Relat Res 468, 2328–2332 (2010). https://doi.org/10.1007/s11999-010-1383-8

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11999-010-1383-8

Keywords

Navigation