Abstract
An 80-year-old woman had undergone initial mitral valve replacement using a Björk-Shiley mechanical valve owing to mitral stenosis 25 years earlier. Suddenly, she had anemia and an increased lactic dehydrogenase (LDH) level. Transesophageal echography (TEE) showed perivalvular leakage. In a redo operation, two side-by-side stitches of the valve on the posterior annulus were loosened without cutting and the sewing cuff at that site was floated over the annulus, leading to the perivalvular leakage. The valve was easily removed; and round, hard, degenerative calcified tissue composed of remnant mitral valve in the suture site during the initial operation was found just under the sewing cuff. After resection of this calcified round tissue, a 25-mm bioprosthesis was put in place. Her postoperative recovery was uneventful, and 47 days after surgery she was discharged without perivalvular leakage or anemia.
Similar content being viewed by others
References
Czer LS, Chaux A, Matloff JM, DeRobertis MA, Nessim SA, Scarlata D, et al. Ten-year experience with the St. Jude Medical valve for primary valve replacement. J Thorac Cardiovasc Surg 1990;100:44–54; discussion 54-5.
Sayed HM, Dacie JV, Handley DA, Lewis SM, Cleland WP. Haemolytic anaemia of mechanical origin after open heart surgery. Thorax 1961;16:356–360.
Horstkotte D, Korfer R, Seipel L, Bircks W, Loogen F. Late complications in patients with Bjork-Shiley and St. Jude Medical heart valve replacement. Circulation 1983;68(Pt 2):II175–II184.
Fernandez G, Deville C, Horovitz S, Thibaud D, Fontan F. [Long-term results (6 to 17 years) of 372 mitral valve replacements using the Bjork-Shiley prosthesis.] Ann Chir 1989;43:94–98.
Hammermeister KE, Sethi GK, Henderson WG, Oprian C, Kim T, Rahimtoola S. A comparison of outcomes in men 11 years after heart-valve replacement with a mechanical valve or bioprosthesis: Veterans Affairs Cooperative Study on Valvular Heart Disease. N Engl J Med 1993;328:1289–1296.
Lindblom D. Long-term clinical results after aortic valve replacement with the Bjork-Shiley prosthesis. J Thorac Cardiovasc Surg 1988;95:658–667.
Sugimoto T, Miyashita M, Ota T, Yamashita C, Okada M, Nakamura K. [Surgical treatment of valvular heart disease associated with Behcet’s disease.] Nippon Kyobu Geka Gakkai Zasshi 1990;38:2320–2325.
Minami H, Wakita N, Kawanishi Y, Kitano I, Sakata M, Shida T. Aortic aneurysm and aortic regurgitation following aortic valve replacement due to Takayasu’s arteritis. Jpn J Thorac Cardiovasc Surg 2002;50:43–45.
Antunes MJ, Wessels A, Sadowski RG, Schutz JG, Vanderdonck KM, Oliveira JM, et al. Medtronic Hall valve replacement in a third-world population group: a review of the performance of 1000 prostheses. J Thorac Cardiovasc Surg 1988;95:980–993.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Minami, H., Asada, T. & Gan, K. Perivalvular leakage 25 years after initial mitral valve replacement with a Björk-Shiley prosthesis. Gen Thorac Cardiovasc Surg 56, 462–464 (2008). https://doi.org/10.1007/s11748-008-0275-2
Received:
Accepted:
Published: