Abstract
Ultrasonic cleaning is often used as part of the manufacturing process of small medical devices such as guide wires and vascular implants. Ultrasonic cleaning at frequencies close to the natural frequency of the device can result in resonance, resulting in significant mechanical damage and possibly premature failure. This paper provides case studies of ultrasonic cleaning-induced fatigue and corresponding failures in small medical devices. Preventative measures, including analytical tools such as finite element analysis (FEA), to ensure that ultrasonic cleaning frequencies do not result in resonance and stresses sufficient to cause fatigue damage are also discussed.
Similar content being viewed by others
References
Mallette, L.A., Chen, I., Johnson, T.W.: A case study in ultrasonic cleaning damage. IEEE Aerospace Conf. 3(21–28), 77–80 (1998)
Yuqi, J., Dong, T., Xianzhong, S., Fai, L.: The effect of ultrasonic cleaning on the bond wires, Proceedings of the 12th IPFA, 2005, Singapore, IEEE, pp. 237–241
Marrey, R.V., Burgermeister, R., Grishaber, R.B., Ritchie, R.O.: Fatigue life and prediction for cobalt-chromium stents: a fracture mechanics analysis. Biomaterials 27, 1988–2000 (2006)
Timoshenko, S., Young, D.H., Weaver, W.: Vibration Problems in Engineering, 4th edn. Wiley, New York (1974)
Handbook of Silicon Wafer Cleaning Technology, 2nd edn. In: Reinhardt, K.A., Kern, W. (eds.). William Andrews Publishing (2008)
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
James, B.A., McVeigh, C., Rosenbloom, S.N. et al. Ultrasonic Cleaning-Induced Failures in Medical Devices. J Fail. Anal. and Preven. 10, 223–227 (2010). https://doi.org/10.1007/s11668-010-9339-6
Received:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11668-010-9339-6