Abstract
Recently, computed tomography with photon-counting detector (PCD-CT) has been developed to enable high-resolution imaging at a lower radiation dose. PCD-CT employs a photon-counting detector that can measure the number of incident X-ray photons and their energy. The newly released PCD-CT (NAEOTOM Alpha, Siemens Healthineers, Forchheim, Germany) has been in clinical use at our institution since December 2022. The PCD-CT offers several advantages over current state-of-the-art energy-integrating detector CT (EID-CT). The PCD-CT does not require septa to create a detector channel, while EID-CT does. Therefore, downsizing the anode to achieve higher resolution does not affect the dose efficiency of the PCD-CT. CT is an indispensable modality for evaluating ear ossicles. The ear ossicles and joints are clearly depicted by PCD-CT. In particular, the anterior and posterior legs of the stapes, which are sometimes unclear on conventional CT scans, can be clearly visualized. We present cases of congenital anomalies of the ossicular chain, ossicular chain dislocation, tympanosclerosis, and cholesteatoma in which PCD-CT was useful. This short article reports the usefulness of PCD-CT in the 3D visualization of the ear ossicles.
Similar content being viewed by others
Avoid common mistakes on your manuscript.
Introduction
X-ray computed tomography (CT) is an indispensable modality for evaluating ear ossicles. CT with a photon-counting detector (PCD-CT) was recently developed to enable high-resolution imaging at lower radiation doses. The usefulness of this tool in evaluating the ear ossicles has been reported in studies in which investigational PCD-CT [1] and cadavers [2] were used. The newly released PCD-CT system (NAEOTOM Alpha; Siemens Healthineers, Forchheim, Germany) has been used clinically at our institution since December 2022. This short article further reports the usefulness of PCD-CT in the 3D visualization of the ear ossicles.
PCD-CT employs a photon-counting detector (PCD) that measures the number of incident X-ray photons and their energy.
PCD-CT offers several advantages over the current state-of-the-art CTs that use energy-integrating detectors (EID). The PCD-CT does not require septa to create a detector channel for the EID; what serves the same purpose as a detector channel in EID are the smaller pixelated anodes (0.2 mm × 0.2 mm) that are placed on the bottom of a single crystal of cadmium–telluride (Fig. 1). Therefore, the dose efficiency of the PCD does not decrease with smaller anodes to achieve a higher spatial resolution. Compared with EID-CT with the same detector cell size (0.25 × 0.25 mm), PCD-CT shows a 19% reduction in image noise [3] (Fig. 2).
Schematic of the direct-converting photon-counting detector. The X-ray absorbed in a semiconductor—a single crystal of cadmium–telluride (CdTe)—produces electron–hole pairs that are separated by a strong electric field E between the cathode and pixelated anodes. No septa were observed between the anodes. ASIC application-specific integrated circuit
PCD is also unique in that it converts incident X-ray photons into pulses, the heights of which reflect the energy value. The energy of each X-ray photon was measured using an energy discriminator with a preset PCD energy threshold. The PCD was equipped with multiple energy discriminators allowing simultaneous energy measurements. The lowest energy threshold can eliminate electrical noise in the measurement signal, which reduces the signal-to-noise ratio in high-resolution and low-dose imaging. This mechanism reduces image noise by approximately 46% [4].
Acquisition and imaging
The scanning and imaging parameter settings used for this study are shown in Table 1. The automatic tube current modulation was used for patient dose optimization. Filtered back-projection and iterative reconstruction techniques are available for PCD-CT image reconstruction.
Demonstration of the normal ear ossicles
The ear ossicles comprise three bones: the malleus, incus, and stapes, joined by the incudomalleolar and incudostapedial joints. The ear ossicles and joints are depicted clearly by PCD-CT (Fig. 3). In particular, the anterior and posterior legs of the stapes, which are sometimes unclear on conventional CT scans, can be clearly visualized.
Congenital anomalies of the ossicular chain
Congenital middle ear anomalies can be classified into four categories according to the Cremers and Teunissen classification [5]. We present a case of category 3, "Ossicular chain anomaly, mobile stapes footplate," which is the most frequent one (Fig. 4).
Ossicular chain dislocation
Ossicular chain dislocation is relatively rare and may be associated with trauma to the temporal bone, although its exact incidence is unknown [6].
It is estimated that up to 50% of temporal bone fractures result in damage to the ear ossicles [7, 8]. However, ossicular chain dislocations can occur even without temporal bone fractures [9].
Radiologically, it is usually challenging to find an incudomalleolar joint dislocation [10]. However, PCD-CT can identify minute dislocations on high-resolution images. Furthermore, 3D reconstruction using PCD-CT allowed us to examine the incudomalleolar and incudostapedial joints in multiple planes and angles, thereby increasing the possibility of identifying traumatic injury structures in the temporal bone (Fig. 5).
A case of traumatic ossicular chain dislocation in a woman in her 50 s with a chief complaint of left hearing loss. A; 3D image of the normal right ossicles. B; 3D image of left ossicular chain dislocation. C; Axial image of the normal right ossicle. D; Axial image of left ossicular chain dislocation. E, F; intraoperative images. PCD-CT showing dislocation of the left incudomalleolar joint in the 3D image (B, arrow) and axial image (D, arrow). The head of malleus is anteriorly dislocated in the left tympanum, unlike the normal incudomalleolar joint in the right (A, C). During the surgery, anterior deviation of the malleus was observed. Dislocation of the incudomalleolar joint (E, arrow). There was no evidence of incudostapedial joint disarticulation (F, arrow). Preoperative CT findings were consistent with intraoperative findings.
Tympanosclerosis
Tympanosclerosis was observed as a high-density area on the CT scans (Fig. 6). Chronic suppurative otitis media is the most common etiological factor in tympanosclerosis. Tympanosclerosis of the middle ear cavity is most often observed around the malleus handle [11].
Cholesteatoma
Bone destruction is a feature of cholesteatoma, and the coexistence of a soft tissue-dense lesion in the tympanum and destruction of the ear ossicles are specific findings of cholesteatoma (Fig. 7).
Conclusion
The ear ossicles could be clearly visualized using PCD-CT, particularly in 3D images.
References
Benson JC, Rajendran K, Lane JI, Diehn FE, Weber NM, Thorne JE, et al. A new frontier in temporal bone imaging: photon-counting detector CT demonstrates superior visualization of critical anatomic structures at reduced radiation dose. AJNR Am J Neuroradiol. 2022;43:579–84.
Zhou W, Lane JI, Carlson ML, Bruesewitz MR, Witte RJ, Koeller KK, et al. Comparison of a photon-counting-detector CT with an energy-integrating-detector CT for temporal bone imaging: a cadaveric study. AJNR Am J Neuroradiol. 2018;39:1733–8.
Pourmorteza A, Symons R, Henning A, Ulzheimer S, Bluemke DA. Dose efficiency of quarter-millimeter photon-counting computed tomography: first-in-human results. Invest Radiol. 2018;53:365–72.
Yu Z, Leng S, Kappler S, Hahn K, Li Z, Halaweish AF, et al. Noise performance of low-dose CT: comparison between an energy integrating detector and a photon counting detector using a whole-body research photon counting CT scanner. J Med Imaging (Bellingham). 2016;3: 043503.
Teunissen EB, Cremers WR. Classification of congenital middle ear anomalies. Report on 144 ears. Ann Otol Rhinol Laryngol. 1993;102:606–12.
Schubiger O, Valavanis A, Stuckmann G, Antonucci F. Temporal bone fractures and their complications. Examination with high resolution CT. Neuroradiology. 1986;28:93–9.
Yetiser S, Hidir Y, Birkent H, Satar B, Durmaz A. Traumatic ossicular dislocations: etiology and management. Am J Otolaryngol. 2008;29:31–6.
Meriot P, Veillon F, Garcia JF, Nonent M, Jezequel J, Bourjat P, et al. CT appearances of ossicular injuries. Radiographics. 1997;17:1445–54.
Delrue S, Verhaert N, Dinther JV, Zarowski A, Somers T, Desloovere C, et al. Surgical management and hearing outcome of traumatic ossicular injuries. J Int Adv Otol. 2016;12:231–6.
Maillot O, Attyé A, Boyer E, Heck O, Kastler A, Grand S, et al. Post-traumatic deafness: a pictorial review of CT and MRI findings. Insights Imaging. 2016;7:341–50.
Ho KY, Tsai SM, Chai CY, Wang HM. Clinical analysis of intratympanic tympanosclerosis: etiology, ossicular chain findings, and hearing results of surgery. Acta Otolaryngol. 2010;130:370–4.
Funding
The author(s) received no financial support for the research, authorship, or publication of this article.
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
Toshihide Itoh is an employee of Siemens Healthineers. The authors declare no conflict of interest.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
This article is published under an open access license. Please check the 'Copyright Information' section either on this page or in the PDF for details of this license and what re-use is permitted. If your intended use exceeds what is permitted by the license or if you are unable to locate the licence and re-use information, please contact the Rights and Permissions team.
About this article
Cite this article
Takahashi, Y., Higaki, F., Sugaya, A. et al. Evaluation of the ear ossicles with photon-counting detector CT. Jpn J Radiol 42, 158–164 (2024). https://doi.org/10.1007/s11604-023-01485-0
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11604-023-01485-0