Abstract

Purpose

To conduct computed tomography (CT)-guided puncture exactly and safely, we newly developed a laser guiding puncture system that can be used in a commercially available CT scanner.

Materials and methods

The laser-guided CT puncture system is built on the CT table with an aluminum frame. Preliminary simulation tests were conducted using two models representing the body and nodular lesions, and puncture procedures were carried out for 15 patients using this system.

Results

The mean distance and standard deviation from the center in simulation experiments conducted using this puncture system were 2.95 ± 1.20 mm for operator A and 3.52 ± 1.12 mm for operator B. There was no statistically significant difference between the operators (P = 0.40) or the angles (P = 0.32). For five lung biopsy patients, the distance from the target point planned before biopsy to the actual last puncture point was 0–8 mm. For 10 percutaneous vertebroplasty (PVP) patients (two performed in Th11, one in Th12, five in L1, two in L2), the plan before the puncture procedure was to pass the needle through the vertebral pedicle in all cases. The distance between the planned target point and the actual last puncture point was 0–5 mm.

Conclusion

This system has the potential to accomplish the CT-guided puncture procedure safely and accurately.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

References

  1. N Nitta M Mori K Murata M Takahashi A Mishina H Matsuo et al. (1997) ArticleTitleA new mechanically manipulated unit for CT-guided biopsy Nippon Igaku Houshasen Gakkai Zasshi 57 675–7 Occurrence Handle1:STN:280:DyaK1c%2FjtFClsQ%3D%3D

    CAS  Google Scholar 

  2. JM Meyer T Schmitz-Rode G Krombach JE Wildberger J Pfeffer RW Gunther (2001) ArticleTitleNavi-ball: a new guidance device for CT-directed punctures Invest Radiol 36 299–302 Occurrence Handle11323517 Occurrence Handle10.1097/00004424-200105000-00008 Occurrence Handle1:STN:280:DC%2BD3MnjsV2rug%3D%3D

    Article  PubMed  CAS  Google Scholar 

  3. T Tanaka A Furukawa N Nitta M Yamasaki M Takahashi K Murata (2001) ArticleTitleDevelopment of a new device for CT guided percutaneous biopsy Jpn J Intervent Radiol 16 365–7

    Google Scholar 

  4. J Kettenbach G Kronreif M Furst W Birkfellner R Hanel W Ptacek (2005) ArticleTitleRobot-assisted biopsy using computed tomography-guidance: initial results from in vitro tests Invest Radiol 40 219–28 Occurrence Handle15770140 Occurrence Handle10.1097/01.rli.0000155285.05672.cf

    Article  PubMed  Google Scholar 

  5. A Magnusson E Radecka M Lonnemark H Raland (2005) ArticleTitleComputed-tomography-guided procedures using a new guidance device Acta Radiol 46 505–9 Occurrence Handle16224926 Occurrence Handle10.1080/02841850510021508 Occurrence Handle1:STN:280:DC%2BD2MrkvVGqug%3D%3D

    Article  PubMed  CAS  Google Scholar 

  6. JR Haaga RJ Alfidi (1976) ArticleTitlePrecise biopsy localization by computed tomography Radiology 118 603–7 Occurrence Handle1251009 Occurrence Handle1:STN:280:CSmC38flslI%3D

    PubMed  CAS  Google Scholar 

  7. K Katada R Kato H Anno Y Ogura S Koga Y Ida et al. (1996) ArticleTitleGuidance with real-time CT fluoroscopy: early clinical experience Radiology 200 851–6 Occurrence Handle8756943 Occurrence Handle1:STN:280:BymA2cjos1c%3D

    PubMed  CAS  Google Scholar 

  8. PR Frederick TH Brown MH Miller AL Bahr KH Taylor (1985) ArticleTitleA light-guidance system to be used for CT-guided biopsy Radiology 154 535–6 Occurrence Handle3966142 Occurrence Handle1:STN:280:BiqD1cbjvVU%3D

    PubMed  CAS  Google Scholar 

  9. Y Nagata T Nishida M Abe M Hiraoka M Takahashi K Fujiwara et al. (1990) ArticleTitleLaser projection system for radiotherapy and CT-guided biopsy J Comput Assist Tomogr 14 1046–8 Occurrence Handle2229558 Occurrence Handle10.1097/00004728-199011000-00042 Occurrence Handle1:STN:280:By6D3snis1Q%3D

    Article  PubMed  CAS  Google Scholar 

  10. Y Miaux A Guermazi D Gossot P Bourrier D Angoulvant A Khairoune et al. (1995) ArticleTitleLaser guidance system for CT-guided procedures Radiology 194 282–4 Occurrence Handle7997570 Occurrence Handle1:STN:280:ByqD1M%2Fht10%3D

    PubMed  CAS  Google Scholar 

  11. FS Pereles HT Ozgur PJ Lund EC Unger (1997) ArticleTitlePotentials of laser guidance system for percutaneous musculoskeletal procedures Skeletal Radiol 26 650–3 Occurrence Handle9428072 Occurrence Handle10.1007/s002560050304 Occurrence Handle1:STN:280:DyaK1c%2FovFGnsw%3D%3D

    Article  PubMed  CAS  Google Scholar 

  12. H Ishizaki T Katsuya Y Koyama H Ishijima T Moteki JC Aoki (1998) ArticleTitleT-guided percutaneous intervention using a simple laser director device AJR Am J Roentgenol 170 745–6

    Google Scholar 

  13. K Brabrand TM Aalokken GA Krombach RW Gunther R Tariq A Magnusson et al. (2004) ArticleTitleMulticenter evaluation a new laser guidance system for computed tomography intervention Acta Radiol 45 308–12 Occurrence Handle15239427 Occurrence Handle10.1080/02841850410005039 Occurrence Handle1:STN:280:DC%2BD2czjsleiuw%3D%3D

    Article  PubMed  CAS  Google Scholar 

  14. FS Pereles M Baker R Baldwin E Krupinski EC Unger (1998) ArticleTitleAccuracy of CT biopsy: laser guidance versus conventional freehand techniques Acad Radiol 5 766–70 Occurrence Handle9809074 Occurrence Handle10.1016/S1076-6332(98)80260-2 Occurrence Handle1:STN:280:DyaK1M%2Fis1KjsA%3D%3D

    Article  PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Norihisa Nitta.

About this article

Cite this article

Nitta, N., Takahashi, M., Tanaka, T. et al. Laser-guided computed tomography puncture system: simulation experiments using artificial phantom lesions and preliminary clinical experience. Radiat Med 25, 187–193 (2007). https://doi.org/10.1007/s11604-006-0116-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11604-006-0116-0

Key words

Navigation