5 Conclusions
The calibration of tiltmeter’s scale value is very important. Without a correct calibration, the observation is of little application significance. At present, a ‘direct calibration’ method (ZHU, 1993) is used, which gives a known displacement value on the instrument and then measures the output signals of the instrument under such a condition. However, the known displacement value must be very accurate.
We believe that the calibration by bulging plate can meet the accuracy, but the following factors should be noted.
-
1)
It should be very strict in selecting materials and techniques for bulging-plate making.
-
2)
Regular recalibration is needed for bulging plates that have been used for many years in order to ensure the credible scale values. Moreover, it is necessary to develop a portable equipment for itinerate calibration at seismic stations.
-
3)
Mercury is toxiferous as a kind of pressure-transmitting medium in the bulging plates. By accident, the vapor of leaking mercury is liable to cause hydrargyrism to the observer. Therefore, it’s better to replace mercury with innocuous high-density liquid in the bulging plates.
Similar content being viewed by others
References
LI Ping, CHEN De-fu, FENG Liu-yi. 1989. Preliminary study on scale constants calibration of tiltmeter [A]. In: Proceedings on the Earthquake Monitoring and Forecast Method [C]. Beijing: Seismological Press, 145–151 (in Chinese).
Melchior P. 1966. The Earth Tides [M]. Paliar: Pregamon Press, 126–189.
Melchior P. 1978. The Tides of the Planet Earth [M]. Paliar: Pregamon Press, 215–256.
First Medical College of Shanghai (eds). 1972. Pests Elimination and Occupational Diseases Prevention [M]. Shanghai: Shanghai Renmin Publishing House, 157–158 (in Chinese).
Verbaandert J. 1962. L’etalonage des pandules horigontaux [J]. Bollettno Di Geofisica, 4(16): 419–446.
YUAN Xi-guang. 1986. A Handbook of Sensor [M]. Beijing: National Defense Industry Press, 102–108 (in Chinese).
ZHANG Zhi-zhong and ZHU Hu. 2002. Research on without-mercury bulging plate calibration device for horizontal [J]. Annual of the Chinese Geophysical Society, 203 (in Chinese).
ZHU Hu and FENG Hai-ying. 1980. Calibration of quartz horizontal pendulum type tiltmeter [J]. Acta Seismologica Sinica, 2(3): 332–341 (in Chinese).
ZHU Hu. 1993. Calibration of fixed-point deformation [J]. Seismological and Geomagnetic Observation and Research, 14(1): 33–39 (in Chinese).
Author information
Authors and Affiliations
Additional information
Contribution No. 05FE3022, Institute of Geophysics, China Earthquake Administration.
About this article
Cite this article
Zhang, Zz., Zhu, H. & Wu, J. Research on bulging plate of scaler for horizontal pendulum tiltmeter. Acta Seimol. Sin. 18, 618–623 (2005). https://doi.org/10.1007/s11589-005-0041-9
Received:
Revised:
Accepted:
Issue Date:
DOI: https://doi.org/10.1007/s11589-005-0041-9