PET/CT-guided biopsy with respiratory motion correction

  • Original Article
  • Published:
International Journal of Computer Assisted Radiology and Surgery Aims and scope Submit manuscript

Abstract

Purpose

Given the ability of positron emission tomography (PET) imaging to localize malignancies in heterogeneous tumors and tumors that lack an X-ray computed tomography (CT) correlate, combined PET/CT-guided biopsy may improve the diagnostic yield of biopsies. However, PET and CT images are naturally susceptible to problems due to respiratory motion, leading to imprecise tumor localization and shape distortion. To facilitate PET/CT-guided needle biopsy, we developed and investigated the feasibility of a workflow that allows to bring PET image guidance into interventional CT suite while accounting for respiratory motion.

Methods

The performance of PET/CT respiratory motion correction using registered and summed phases method was evaluated through computer simulations using the mathematical 4D extended cardiac-torso phantom, with motion simulated from real respiratory traces. The performance of PET/CT-guided biopsy procedure was evaluated through operation on a physical anthropomorphic phantom. Vials containing radiolabeled 18F-fluorodeoxyglucose were placed within the physical phantom thorax as biopsy targets. We measured the average distance between target center and the simulated biopsy location among multiple trials to evaluate the biopsy localization accuracy.

Results

The computer simulation results showed that the RASP method generated PET images with a significantly reduced noise of 0.10 ± 0.01 standardized uptake value (SUV) as compared to an end-of-expiration image noise of 0.34 ± 0.04 SUV. The respiratory motion increased the apparent liver lesion size from 5.4 ± 1.1 to 35.3 ± 3.0 cc. The RASP algorithm reduced this to 15.7 ± 3.7 cc. The distances between the centroids for the static image lesion and two moving lesions in the liver and lung, when reconstructed with the RASP algorithm, were 0.83 ± 0.72 mm and 0.42 ± 0.72 mm. For the ungated imaging, these values increased to 3.48 ± 1.45 mm and 2.5 ± 0.12 mm, respectively. For the ungated imaging, this increased to 1.99 ± 1.72 mm. In addition, the lesion activity estimation (e.g., SUV) was accurate and constant for images reconstructed using the RASP algorithm, whereas large activity bias and variations (± 50%) were observed for lesions in the ungated images. The physical phantom studies demonstrated a biopsy needle localization error of 2.9 ± 0.9 mm from CT. Combined with the localization errors due to respiration for the PET images from simulations, the overall estimated lesion localization error would be 3.08 mm for PET-guided biopsies images using RASP and 3.64 mm when using ungated PET images. In other words, RASP reduced the localization error by approximately 0.6 mm. The combined error analysis showed that replacing the standard end-of-expiration images with the proposed RASP method in PET/CT-guided biopsy workflow yields comparable lesion localization accuracy and reduced image noise.

Conclusion

The RASP method can produce PET images with reduced noise, attenuation artifacts and respiratory motion, resulting in more accurate lesion localization. Testing the PET/CT-guided biopsy workflow using computer simulation and physical phantoms with respiratory motion, we demonstrated that guided biopsy procedure with the RASP method can benefit from improved PET image quality due to noise reduction, without compromising the accuracy of lesion localization.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8

Similar content being viewed by others

References

  1. Bar-Shalom R, Yefremov N, Guralnik L, Gaitini D, Frenkel A, Kuten A, Altman H, Keidar Z, Israel O (2003) Clinical performance of PET/CT in evaluation of cancer: additional value for diagnostic imaging and patient management. J Nucl Med 44(8):1200–1209

    PubMed  Google Scholar 

  2. Hillner BE, Siegel BA, Liu D, Shields AF, Gareen IF, Hanna L, Stine SH, Coleman RE (2008) Impact of positron emission tomography/computed tomography and positron emission tomography (PET) alone on expected management of patients with cancer: initial results from the national oncologic PET registry. J Clin Oncol 26(13):2155–2161

    Article  Google Scholar 

  3. Farwell MD, Pryma DA, Mankoff DA (2014) PET/CT imaging in cancer: current applications and future directions. Cancer 120(22):3433–3445

    Article  CAS  Google Scholar 

  4. Lardinois D, Weder W, Hany TF, Kamel EM, Korom S, Seifert B, von Schulthess GK, Steinert HC (2003) Staging of non-small-cell lung cancer with integrated positron-emission tomography and computed tomography. N Engl J Med 348(25):2500–2507

    Article  Google Scholar 

  5. Cerfolio RJ, Ojha B, Bryant AS, Raghuveer V, Mountz JM, Bartolucci AA (2004) The accuracy of integrated PET-CT compared with dedicated PET alone for the staging of patients with nonsmall cell lung cancer. Ann Thorac Surg 78(3):1017–1023

    Article  Google Scholar 

  6. von Schulthess GK, Steinert HC, Hany TF (2006) Integrated PET/CT: current applications and future directions. Radiology 238(2):405–422

    Article  Google Scholar 

  7. Metser U, Miller E, Lerman H, Even-Sapir E (2007) Benign nonphysiologic lesions with increased 18F-FDG uptake on PET/CT: characterization and incidence. Am J Roentgenol 189(5):1203–1210

    Article  Google Scholar 

  8. Zhuang H, Alavi A (2002) 18-Fluorodeoxyglucose positron emission tomographic imaging in the detection and monitoring of infection and inflammation. Semin Nucl Med 32(1):47–59

    Article  Google Scholar 

  9. Tatli S, Gerbaudo VH, Mamede M, Tuncali K, Shyn PB, Silverman SG (2010) Abdominal masses sampled at PET/CT-guided percutaneous biopsy: initial experience with registration of prior PET/CT images. Radiology 256(1):305–311

    Article  Google Scholar 

  10. O’Sullivan PJ, Rohren EM, Madewell JE (2008) Positron emission tomography–CT imaging in guiding musculoskeletal biopsy. Radiol Clin 46(3):475–486

    Article  Google Scholar 

  11. Mallarajapatna GJ, Kallur KG, Ramanna NK, Susheela SP, Ramachandra PG (2009) PET/CT-guided percutaneous biopsy of isolated intramuscular metastases from postcricoid cancer. J Nucl Med Technol 37(4):220–222

    Article  Google Scholar 

  12. Govindarajan MJ, Nagaraj KR, Kallur KG, Sridhar PS (2009) PET/CT guidance for percutaneous fine needle aspiration cytology/biopsy. Indian J Radiol 19(3):208–209

    Article  CAS  Google Scholar 

  13. Cohade C, Osman M, Marshall LT, Wahl RL (2003) PET-CT: accuracy of PET and CT spatial registration of lung lesions. Eur J Nucl Med Mol Imaging 30(5):721–726

    Article  Google Scholar 

  14. Goerres GW, Burger C, Schwitter MR, Heidelberg TN, Seifert B, von Schulthess GK (2003) PET/CT of the abdomen: optimizing the patient breathing pattern. Eur Radiol 13(4):734–739

    PubMed  Google Scholar 

  15. Beyer T, Rosenbaum S, Veit P, Stattaus J, Müller SP, DiFilippo FP, Schöder H, Mawlawi O, Roberts F, Bockisch A, Kühl H (2005) Respiration artifacts in whole-body 18 F-FDG PET/CT studies with combined PET/CT tomographs employing spiral CT technology with 1 to 16 detector rows. Eur J Nucl Med Mol Imaging 32(12):1429–1439

    Article  Google Scholar 

  16. Allen-Auerbach M, Yeom K, Park J, Phelps M, Czernin J (2006) Standard PET/CT of the chest during shallow breathing is inadequate for comprehensive staging of lung cancer. J Nucl Med 47(2):298–301

    PubMed  Google Scholar 

  17. Liu C, Pierce LA, Alessio AM, Kinahan PE (2009) Impact of respiratory motion on tumor quantification and delineation in static PET/CT imaging. Phys Med Biol 54(24):7345

    Article  Google Scholar 

  18. Kinahan P, MacDonald L, Ng L, Alessio A, Segars P, Tsui B, Pathak S (2006) Compensating for patient respiration in PET/CT imaging with the registered and summed phases (RASP) procedure. In: 3rd IEEE international symposium on biomedical imaging: nano to macro, pp 1104–1107

  19. Khare R, Sala G, Kinahan P, Esposito G, Banovac F, Cleary K, Enquobahrie A (2013) Experimental evaluation of a deformable registration algorithm for motion correction in PET-CT guided biopsy. In: Nuclear science symposium and medical imaging conference (NSS/MIC), pp 1–5

  20. Johnson H, Harris G, Williams K (2007) BRAINSFit: mutual information registrations of whole-brain 3D images, using the insight toolkit. Insight J. http://hdl.handle.net/1926/1291

  21. Yaniv Z, Wilson E, Lindisch D, Cleary K (2009) Electromagnetic tracking in the clinical environment. Med Phys 36(3):876–892

    Article  Google Scholar 

  22. Fedorov A, Beichel R, Kalpathy-Cramer J, Finet J, Fillion-Robin JC, Pujol S, Bauer C, Jennings D, Fennessy F, Sonka M, Buatti J (2012) 3D Slicer as an image computing platform for the Quantitative Imaging Network. Magn Reson Imaging 30(9):1323–1341

    Article  Google Scholar 

  23. PET-CT project wiki page. http://public.kitware.com/Wiki/PET-CT. Accessed on 8 Jan 2015

  24. Zukić D, Finet J, Wilson E, Banovac F, Esposito G, Cleary K, Enquobahrie A (2015) SlicerPET: a workflow based software module for PET/CT guided needle biopsy. Comput Aided Radiol Surg 10:S186–S187. https://doi.org/10.1007/s11548-015-1213-2

    Article  Google Scholar 

  25. Enquobahrie A, Gobbi D, Turek MW, Cheng P, Yaniv Z, Lindseth F, Cleary K (2008) Designing tracking software for image-guided surgery applications: IGSTK experience. Int J Comput Assist Radiol Surg 3(5):395–403

    Article  Google Scholar 

  26. Segars WP, Tsui BM (2009) MCAT to XCAT: the evolution of 4-D computerized phantoms for imaging research. Proc IEEE 97(12):1954–1968

    Article  Google Scholar 

  27. Zhang R, Alessio AM, Pierce LA, Byrd DW, Lee TC, De Man B, Kinahan PE (2017) Improved attenuation correction for respiratory gated PET/CT with extended-duration cine CT: a simulation study. In: Medical imaging 2017: Physics of medical imaging 10132:101321I

  28. Lin R, Wilson E, Tang J, Stoianovici D, Cleary K (2007) A computer-controlled pump and realistic anthropomorphic respiratory phantom for validating image-guided systems. In: Proceedings of the SPIE medical imaging conference 2007: visualization and image-guided procedures, vol 6509, p 65090E1-9

  29. Yaniv Z, Cheng P, Wilson E, Popa T, Lindisch D, Campos-Nanez E, Abeledo H, Watson V, Cleary K, Banovac F (2010) Needle-based interventions with the image-guided surgery toolkit (IGSTK): from phantoms to clinical trials. IEEE Trans Biomed Eng 57(4):922–933

    Article  Google Scholar 

  30. Tatli S, Gerbaudo VH, Mamede M, Tuncali K, Shyn PB, Silverman SG (2010) Abdominal masses sampled at PET/CT-guided percutaneous biopsy: initial experience with registration of prior PET/CT images. Radiology 256(1):305–311

    Article  Google Scholar 

  31. Venkatesan AM, Kadoury S, Abi-Jaoudeh N, Levy EB, Maass-Moreno R, Krücker J, Dalal S, Xu S, Glossop N, Wood BJ (2011) Real-time FDG PET guidance during biopsies and radiofrequency ablation using multimodality fusion with electromagnetic navigation. Radiology 260(3):848–856

    Article  Google Scholar 

  32. Chang G, Chang T, Pan T, Clark JW, Mawlawi OR (2010) Implementation of an automated respiratory amplitude gating technique for PET/CT: clinical evaluation. J Nucl Med 51(1):16–24

    Article  Google Scholar 

Download references

Funding

This work was supported in part by the National Institute of Health (NIH) Grants R42 CA153488, R01 CA160253 and R42 CA167907.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Paul E. Kinahan.

Ethics declarations

Conflict of interest

The authors declare that they have no conflict of interest.

Ethical approval

This article does not contain any studies with human participants performed by any of the authors.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Zhang, R., Zukić, D., Byrd, D.W. et al. PET/CT-guided biopsy with respiratory motion correction. Int J CARS 14, 2187–2198 (2019). https://doi.org/10.1007/s11548-019-02047-4

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11548-019-02047-4

Keywords

Navigation