Temporal enhancement of 2D color Doppler echocardiography sequences by fragment-based frame reordering and refinement

  • Original Article
  • Published:
International Journal of Computer Assisted Radiology and Surgery Aims and scope Submit manuscript

Abstract

Purpose

The goal of this study was to develop an algorithm that enhances the temporal resolution of two-dimensional color Doppler echocardiography (2D CDE) by reordering all the acquired frames and filtering out the frames corrupted by out-of-plane motion and arrhythmia.

Methods

The algorithm splits original frame sequence into the fragments based on the correlation with a reference frame. Then, the fragments are aligned temporally and merged into a resulting sequence that has higher temporal resolution. We evaluated the algorithm with 10 animal epicardial 2D CDE datasets of the right ventricle and compared it with the existing approaches in terms of resulting frame rate, image stability and execution time.

Results

We identified the optimal combination of alternatives for each step, which resulted in an increase in frame rate from 14 ± 0.87 to 238 ± 93 Hz. The average execution time was 7.23 ± 0.48 s in comparison with 0.009 ± 0.001 s for ECG gating and 1167.37 ± 587.85 s for flow reordering. Our approach demonstrated a significant (p < 0.01) increase in image stability compared with ECG gating and flow reordering.

Conclusion

This work presents an offline algorithm for temporal enhancement of 2D CDE. Unlike previous frame reordering approaches, it can filter out-of-plane or corrupted frames, increasing the quality of the results, which substantially increases diagnostic value of 2D CDE. It can be used for high-frame-rate intraoperative imaging of intraventricular and valve regurgitant flows and is potentially modifiable for real-time use on ultrasound machines.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. Bercoff J, Montaldo G, Loupas T, Savery D, Meziere F, Fink M, Tanter M (2011) Ultrafast compound Doppler imaging: providing full blood flow characterization. IEEE Trans Ultrason Ferroelectr Freq Control 58(1):134–147

    Article  PubMed  Google Scholar 

  2. Osmanski BF, Pernot M, Fink M, Tanter M (2013) In vivo transthoracic ultrafast Doppler imaging of left intraventricular blood flow pattern. In: 2013 IEEE International Ultrasonics Symposium, pp 1741–1744

  3. Tong L, Ramalli A, Jasaityte R, Tortoli P, D’hooge J (2014) Multi-transmit beam forming for fast cardiac imaging; experimental validation and in vivo application. IEEE Trans Med Imaging 33(6):1205–1219

    Article  PubMed  Google Scholar 

  4. Cikes M, Tong L, Sutherland RG, D’hooge J (2014) Ultrafast cardiac ultrasound imaging technical principles, applications, and clinical benefits. J Am Coll Cadiol Imaging 7:812–823

    Article  Google Scholar 

  5. Chang LW, Hsu KH, Li PC (2009) Graphics processing unit-based high-frame-rate color Doppler ultrasound processing. IEEE Trans Ultrason Ferroelectr Freq Control 56(9):1856–1860

    Article  PubMed  Google Scholar 

  6. Perrin DP, Vasilyev NV, Marx GR, del Nido PJ (2012) Temporal enhancement of three dimensional echocardiography by frame reordering. J Am Coll Cadiol Imaging 5(3):300–304

    Article  Google Scholar 

  7. Schneider RJ, Perrin DP, Vasilyev NV, Marx GR, del Nido PJ (2012) Howe RD Real-time image-based rigid registration of three-dimensional ultrasound. Med Image Anal 16(2):402–414

    Article  PubMed  Google Scholar 

  8. Terentjev AB, Settlemier SH, Perrin DP, del Nido PJ, Shturts IV, Vasilyev NV (2016) Temporal enhancement of two-dimensional color Doppler echocardiography. Proc SPIE 9784:7

    Google Scholar 

  9. Ta CN, Eghtedari M, Mattrey FR, Kono Y, Kummel AC (2014) 2-Tier in-plane motion correction and out-of-plane motion filtering for contrast-enhanced ultrasound. Invest Radiol 49(11):707–719

    Article  PubMed  PubMed Central  Google Scholar 

  10. Danudibroto A, Bersvendsen J, Mirea O, Gerard O, D’hooge J, Samset E (2016) Image-based temporal alignment of echocardiographic sequences. Proc SPIE 9790:7

    Google Scholar 

  11. Zhang W, Noble JA, Brady JM (2007) Spatio-temporal registration of real time 3D ultrasound to cardiovascular MR sequences. In: Medical image computing and computer-assisted intervention. Springer, Berlin, pp 343–350

  12. Dmitrienko A, D’Agostino RB (2018) Multiplicity considerations in clinical trials. N Engl J Med 378(22):2115–2122

    Article  PubMed  Google Scholar 

  13. Sengupta PP, Pedrizzetti G, Kilner PJ, Kheradvar A, Ebbers T, Tonti G, Fraser AG, Narula J (2012) Emerging trends in CV flow visualization. J Am Coll Cardiol Imaging 5(3):305–316

    Article  Google Scholar 

  14. Munoz DR, Markl M, Mur JML, Barker A, Farnandez-Golfin C, Lancellotti P, Zamorano Gomez JL (2013) Intracardiac flow visualization: current status and future directions. Eur Heart J Cardiovasc Imaging. 14(11):1029–1038

    Article  Google Scholar 

  15. Siciliano M, Migliore F, Badano L, Bartaglia E, Pedrizzetti G, Gavedon S, Zorzi A, Corrado D, Iliceto S, Muraru D (2017) Cardiac resynchronization therapy by multipoint pacing improves response of left ventricular mechanics and fluid dynamics: a three-dimensional and particle image velocimetry echo study. EP Europace 19(11):1833–1840

    Article  Google Scholar 

  16. Janaswamy P, Walters TE, Nazer B, Lee RJ (2016) Current treatment strategies for heart failure: role of device therapy and LV reconstruction. Curr Treat Options Cardiovasc Med 18(9):57

    Article  PubMed  Google Scholar 

  17. Payne CJ, Wamala I, Bautista-Salinas D, Saeed M, Van Story D, Thalhofer T, Hovarth MA, Abah C, del Nido PJ, Walsh CJ, Vasilyev NV (2017) Soft robotic ventricular assist device with septal bracing for therapy of heart failure. Sci Robot 2(12):eaan6736

  18. Horvath MA, Wamala I, Rytkin E, Doyle E, Payne CJ, Thalhofter T, Berr I, Solovyeva A, Saeed M, Hendren S, Roche ET, del Nido PJ, Walsh CJ, Vasilyev NV (2017) An intracardiac soft robotic device for augmentation of blood ejection from the failing right ventricle. Ann Biomed Eng 45(9):2222–2233

    Article  PubMed  PubMed Central  Google Scholar 

  19. Schleder S, Dendl LM, Ernstberger A, Nerlich M, Hoffstetter P, Jung EM, Heiss P, Stroszczynski C, Schreyer AG (2013) Diagnostic value of a hand-carried ultrasound device for free intra-abdominal fluid and organ lacerations in major trauma patients. Emerg Med J 30(3):e20–e20

    Article  PubMed  Google Scholar 

  20. Rubenstein JJ, Pohost GM, Dinsmore RE, Harthorne JW (1975) The echocardiographic determination of mitral valve opening and closure. Correlation with hemodynamic studies in man. Circulation 51(1):98–103

    Article  CAS  PubMed  Google Scholar 

Download references

Acknowledgements

We thank Zurab Machaidze, MD, from Boston Children’s Hospital for assisting with image acquisitions and Oleg Talalov from Peter the Great St. Petersburg Polytechnic University for consultation on data representation.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Nikolay V. Vasilyev.

Ethics declarations

Conflict of interest

Scott Settlemier is an employee of Philips Healthcare. All other authors have reported that they have no conflict of interest.

Ethical approval

All applicable international, national and/or institutional guidelines for the care and use of animals were followed.

Informed consent

Statement of informed consent was not applicable since the manuscript does not contain any patient data.

Electronic supplementary material

Below is the link to the electronic supplementary material.

Supplementary material 1 (MP4 5056 kb)

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Terentjev, A.B., Perrin, D.P., Settlemier, S.H. et al. Temporal enhancement of 2D color Doppler echocardiography sequences by fragment-based frame reordering and refinement. Int J CARS 14, 577–586 (2019). https://doi.org/10.1007/s11548-019-01926-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s11548-019-01926-0

Keywords

Navigation