User-friendly freehand ultrasound calibration using Lego bricks and automatic registration

  • Yiming Xiao
  • Charles Xiao Bo Yan
  • Simon Drouin
  • Dante De Nigris
  • Anna Kochanowska
  • D. Louis Collins
Original Article

Abstract

Purpose

As an inexpensive, noninvasive, and portable clinical imaging modality, ultrasound (US) has been widely employed in many interventional procedures for monitoring potential tissue deformation, surgical tool placement, and locating surgical targets. The application requires the spatial mapping between 2D US images and 3D coordinates of the patient. Although positions of the devices (i.e., ultrasound transducer) and the patient can be easily recorded by a motion tracking system, the spatial relationship between the US image and the tracker attached to the US transducer needs to be estimated through an US calibration procedure. Previously, various calibration techniques have been proposed, where a spatial transformation is computed to match the coordinates of corresponding features in a physical phantom and those seen in the US scans. However, most of these methods are difficult to use for novel users.

Methods

We proposed an ultrasound calibration method by constructing a phantom from simple Lego bricks and applying an automated multi-slice 2D–3D registration scheme without volumetric reconstruction. The method was validated for its calibration accuracy and reproducibility.

Results

Our method yields a calibration accuracy of \(1.23\pm 0.26\) mm and a calibration reproducibility of 1.29 mm.

Conclusion

We have proposed a robust, inexpensive, and easy-to-use ultrasound calibration method.

Keywords

Ultrasound Calibration Phantom  Image registration Lego 2D–3D 

Notes

Compliance with ethical standards

Conflict of interest

Yiming Xiao, Charles Xiao Bo Yan, Simon Drouin, Dante De Nigris, Anna Kochanowska, and D. Louis Collins declare that they have no conflict of interest.

References

  1. 1.
    Machi J, Oishi AJ, Furumoto NL, Oishi RH (2004) Intraoperative ultrasound. Surg Clin N Am 84:1085–1111CrossRefPubMedGoogle Scholar
  2. 2.
    Boctor E, deOliveira M, Choti M, Ghanem R, Taylor R, Hager G, Fichtinger G (2006) Ultrasound monitoring of tissue ablation via deformation model and shape priors. Med Image Comput Comput Assist Interv 9:405–412PubMedGoogle Scholar
  3. 3.
    Hammoud MA, Ligon BL, ElSouki R, Shi WM, Schomer DF, Sawaya R (1996) Use of intraoperative ultrasound for localizing tumors and determining the extent of resection: a comparative study with magnetic resonance imaging. J Neurosurg 84:737–741CrossRefPubMedGoogle Scholar
  4. 4.
    Charboneau JW, Reading CC, Welch TJ (1990) Ct and sonographically guided needle-biopsy—current techniques and new innovations. Am J Roentgenol 154:1–10CrossRefGoogle Scholar
  5. 5.
    Comeau RM, Sadikot AF, Fenster A, Peters TM (2000) Intraoperative ultrasound for guidance and tissue shift correction in image-guided neurosurgery. Med Phys 27:787–800CrossRefPubMedGoogle Scholar
  6. 6.
    Keles GE, Lamborn KR, Berger MS (2003) Coregistration accuracy and detection of brain shift using intraoperative sononavigation during resection of hemispheric tumors. Neurosurgery 53:556–562CrossRefPubMedGoogle Scholar
  7. 7.
    Mercier L, Lango T, Lindseth F, Collins DL (2005) A review of calibration techniques for freehand 3-D ultrasound systems. Ultrasound Med Biol 31:449–471CrossRefPubMedGoogle Scholar
  8. 8.
    Detmer PR, Bashein G, Hodges T, Beach KW, Filer EP, Burns DH, Jr DE (1994) Strandness, 3D ultrasonic image feature localization based on magnetic scanhead tracking—in-vitro calibration and validation. Ultrasound Med Biol 20:923–936CrossRefPubMedGoogle Scholar
  9. 9.
    Trobaugh JW, Trobaugh DJ, Richard WD (1994) Three-dimensional imaging with stereotactic ultrasonography. Comput Med Imaging Gr 18:315–323CrossRefGoogle Scholar
  10. 10.
    Prager RW, Rohling RN, Gee AH, Berman L (1998) Rapid calibration for 3D freehand ultrasound. Ultrasound Med Biol 24:855–869CrossRefPubMedGoogle Scholar
  11. 11.
    Najafi M, Afsham N, Abolmaesumi P, Rohling R (2015) A closed-form differential formulation for ultrasound spatial calibration: single wall phantom. Ultrasound Med Biol 41:1079–1094CrossRefPubMedGoogle Scholar
  12. 12.
    Najafi M, Afsham N, Abolmaesumi P, Rohling R (2014) A closed-form differential formulation for ultrasound spatial calibration: multi-wedge phantom. Ultrasound Med Biol 40:2231–2243CrossRefPubMedGoogle Scholar
  13. 13.
    Brown RA (1979) Stereotactic head frame for use with Ct body scanners. Investig Radiol 14:401–401CrossRefGoogle Scholar
  14. 14.
    Comeau RM, Fenster A, Peters TM, (1998) Integrated MR and ultrasound imaging for improved image guidance in neurosurgery. In: Proceedings of SPIE, The International Society for Optical Engineering, pp 747–754Google Scholar
  15. 15.
    Lindseth F, Tangen GA, Lango T, Bang J (2003) Probe calibration for freehand 3D ultrasound. Ultrasound Med Biol 29:1607–1623CrossRefPubMedGoogle Scholar
  16. 16.
    Pagoulatos N, Haynor DR, Kim Y (2001) A fast calibration method for 3D tracking of ultrasound images using a spatial localizer. Ultrasound Med Biol 27:1219–1229CrossRefPubMedGoogle Scholar
  17. 17.
    Boctor E, Viswanathan A, Choti M, Taylor RH, Fichtinger G, Hager G (2004) A novel closed form solution for ultrasound calibration. In: 2nd IEEE international symposium on biomedical imaging: macro to nano, Vols 1 and 2. pp 527–530, 2004Google Scholar
  18. 18.
    Chen TK, Thurston AD, Ellis RE, Abolmaesumi P (2009) A real-time freehand ultrasound calibration system with automatic accuracy feedback and control. Ultrasound Med Biol 35:79–93CrossRefPubMedGoogle Scholar
  19. 19.
    Hsu PW, Prager RW, Gee AH, Treece GM (2008) Real-time freehand 3D ultrasound calibration. Ultrasound Med Biol 34:239–251CrossRefPubMedGoogle Scholar
  20. 20.
    Hsu PW, Treece GM, Prager RW, Houghton NE, Gee AH (2008) Comparison of freehand 3D ultrasound calibration techniques using a stylus. Ultrasound Med Biol 34:1610–1621CrossRefPubMedGoogle Scholar
  21. 21.
    Melvaer EL, Morken K, Samset E (2012) A motion constrained cross-wire phantom for tracked 2D ultrasound calibration. Int J Comput Assist Radiol Surg 7:611–620CrossRefPubMedGoogle Scholar
  22. 22.
    Barratt DC, Penney GP, Chan CS, Slomczykowski M, Carter TJ, Edwards PJ, Hawkes DJ (2006) Self-calibrating 3D-ultrasound-based bone registration for minimally invasive orthopedic surgery. IEEE Trans Med Imaging 25:312–323CrossRefPubMedGoogle Scholar
  23. 23.
    Blackall JM, Rueckert D, Maurer CR, Penney GP, Hill DLG, Hawkes DJ (2000) An image registration approach to automated calibration for freehand 3D ultrasound. Med Image Comput Comput-Assist Interv—Miccai 1935:462–471Google Scholar
  24. 24.
    Mercier L, Del Maestro RF, Petrecca K, Kochanowska A, Drouin S, Yan CX, Janke AL, Chen SJ, Collins DL (2011) New prototype neuronavigation system based on preoperative imaging and intraoperative freehand ultrasound: system description and validation. Int J Comput Assist Radiol Surg 6:507–522CrossRefPubMedGoogle Scholar
  25. 25.
    Poon TC, Rohling RN (2005) Comparison of calibration methods for spatial tracking of a 3D ultrasound probe. Ultrasound Med Biol 31:1095–1108CrossRefPubMedGoogle Scholar
  26. 26.
    Caramanos Z, Fonov VS, Francis SJ, Narayanan S, Pike GB, Collins DL, Arnold DL (2010) Gradient distortions in MRI: characterizing and correcting for their effects on SIENA-generated measures of brain volume change. Neuroimage 49:1601–1611CrossRefPubMedGoogle Scholar
  27. 27.
    Soehl M, Walsh R, Rankin A, Lasso A, Fichtinger G (2014) Tracked ultrasound calibration studies with a phantom made of LEGO bricks. Med Imaging, Image-Guided Proced, Robot Interv, Modeling 9036:2014Google Scholar
  28. 28.
    Yan CXB, Goulet B, Tampieri D, Collins DL (2012) Ultrasound-CT registration of vertebrae without reconstruction. Int J Comput Assist Radiol Surg 7:901–909CrossRefPubMedGoogle Scholar
  29. 29.
    Cleary K, Anderson J, Brazaitis M, Devey G, DiGioia A, Freedman M, Gronemeyer D, Lathan C, Lemke H, Long D, Mun SK, Taylor R (2000) Final report of the technical requirements for image-guided spine procedures workshop, April 17–20, 1999, Ellicott City, MA, USA. Comput Aided Surg 5:180–215PubMedGoogle Scholar

Copyright information

© CARS 2016

Authors and Affiliations

  1. 1.McConnell Brain Imaging Center, Montreal Neurological Institute and HospitalMontrealCanada
  2. 2.Department of RadiologyMcGill University Health CentreMontrealCanada
  3. 3.NeuroRx ResearchMontrealCanada

Personalised recommendations