Abstract
Purpose
Identifying key steps and barriers within complex and simple surgical procedures can be accomplished in a structured and rigorous manner using surgical process modeling. For lower extremity long bone fracture stabilization, the current standard of care is closed intramedullary (IM) nailing, which, despite its widespread use, is associated with challenges that greatly impact operative time and lead to the frustration of medical staff. The aim of this study was to identify challenging surgical steps in IM nailing and understand their underlying causation.
Methods
Eight semi-structured interviews with staff orthopedic surgeons and eight detailed surgical observations were conducted to understand the surgical steps, challenges and adapted techniques used in IM nailing. Hierarchical decomposition was then utilized to structure the IM nailing surgical procedure into phases, steps and activities.
Results
In the developed IM nailing surgical process model, the most challenging steps were identified as fracture reduction (75 %) and entry point selection (25 %), both of which were associated with high levels of frustration in the observed surgeries. Both of these steps utilize 2D fluoroscopic imaging to guide 3D alignment. Challenges arise when the alignment in one plane is lost while adjusting the alignment in the perpendicular plane. This leads to unpredictable repetition of activities which can be time-consuming and frustrating.
Conclusion
Identifying the causation of surgical challenges in IM nailing through surgical process modeling forms a knowledge base that can be used to guide future improvements to techniques and surgical instrumentation.
Similar content being viewed by others
References
Hazan EJ, Joskowicz L (2003) Computer-assisted image-guided intramedullary nailing of femoral shaft fractures. Tech Orthop 18(2):191–200. doi:10.1097/00013611-200306000-00008
Crookshank MC, Edwards MR, Sellan M, Whyne CM, Schemitsch EH (2014) Can fluoroscopy-based computer navigation improve entry point selection for intramedullary nailing of femur fractures? Clin Orthop Relat Res 472:2720–2727. doi:10.1007/s11999-013-2878-x
Liebmann P, Bohn S, Neumuth T (2011) Design and validation of a robust surgical workflow management system. M2Cai2011
Lalys F, Jannin P (2014) Surgical process modelling: a review. Int J Comput Assist Radiol Surg 9:495–511. doi:10.1007/s11548-013-0940-5
Ahmadi S-A, Sielhorst T, Stauder R, Horn M, Feussner H, Navab N (2006) Recovery of surgical workflow without explicit models. Med Image Comput Comput Assist Interv 9:420–428. doi:10.1007/11866565_52
Raimbault M, Jannin P, Morandi X, Riffaud L, Gibaud B (2003) Models of surgical procedures for multimodal image-guided neurosurgery. Stud Health Technol Inform 95(October 2001):50–55. doi:10.3233/978-1-60750-939-4-50
Neumuth D, Loebe F, Herre H, Neumuth T (2011) Modeling surgical processes: a four-level translational approach. Artif Intell Med 51(3):147–161. doi:10.1016/j.artmed.2010.12.003
Neumuth T, Jannin P, Schlomberg J, Meixensberger J, Wiedemann P, Burgert O (2011) Analysis of surgical intervention populations using generic surgical process models. Int J Comput Assist Radiol Surg 6:59–71. doi:10.1007/s11548-010-0475-y
Wang L, Landes J, Weidert S, Blum T, Heide A, Euler E, Navab N (2010) First animal cadaver study for interlocking of intramedullary nails under camera augmented mobile C-arm: a surgical workflow based preclinical evaluation. Lect Notes Comput Sci (including Subser Lect Notes Artif Intell Lect Notes Bioinform). LNCS 6135:56–66. doi:10.1007/978-3-642-13711-2_6
Riffaud L, Neumuth T, Morandi X, Trantakis C, Meixensberger J, Burgert O, Trelhu B, Jannin P (2010) Recording of surgical processes: a study comparing senior and junior neurosurgeons during lumbar disc herniation surgery. Neurosurgery 67. doi:10.1227/NEU.0b013e3181f741d7
Forestier G, Lalys F, Riffaud L, Trelhu B, Jannin P (2012) Classification of surgical processes using dynamic time warping. J Biomed Inform 45(2):255–264. doi:10.1016/j.jbi.2011.11.002
Casaletto JA, Rajaratnam V (2004) Surgical process re-engineering: carpal tunnel decompression—a model. Hand Surg 09(1):19–27. doi:10.1142/S0218810404002066
Sandelowski M (1995) Focus on qualitative methods sample size in qualitative. Res Nurs Heal 18:179–183. doi:10.1002/nur.4770180211
Coyne IT (1997) Sampling in qualitative research. Purposeful and theoretical sampling; merging or clear boundaries? J Adv Nurs 26:623–630. doi:10.1046/j.1365-2648.1997.t01-25-00999.x
Miller G, Galanter E, Pribram K (1960) Plans and the structure of behavior. Henry Holt, New York
Muller ME, Allgower M, Schneider R, Willenegger H (1991) Manual of internal fixation: techniques recommended by the AO-ASIF Group, 3rd edn. Springer, Berlin
Roberts JW, Libet LA, Wolinsky PR (2012) Who is in danger? Impingement and penetration of the anterior cortex of the distal femur during intramedullary nailing of proximal femur fractures. J Trauma Acute Care Surg 73:249–254. doi:10.1097/TA.0b013e318256a0b6
Kanawati AJ, Jang B, McGee R, Sungaran J (2014) The influence of entry point and radius of curvature on femoral intramedullary nail position in the distal femur. J Orthop 11:68–71. doi:10.1016/j.jor.2014.04.010
Giannoudis PV, Pape HC, Schutz M (2007) Femur shaft. https://www2.aofoundation.org/wps/portal/surgery?showPage=approach&contentUrl=srg/32/04-Approaches/32-A10-ante_nail_piri.jsp&bone=Femur&segment=Shaft&approach=Antegradenailing%28piriformis%29&Language=en
Weil YA, Liebergall M, Mosheiff R, Helfet DL, Pearle AD (2007) Long bone fracture reduction using a fluoroscopy-based navigation system? A feasibility and accuracy study. Comput Aided Surg 12:295–302. doi:10.1080/10929080701657974
Diotte B, Fallavollita P, Wang L, Weidert S, Euler E, Thaller P (2015) Multi-modal intra-operative navigation during distal locking of intramedullary nails. IEEE Trans Med Imaging 34(2):487–495
Westphal R, Winkelbach S, Wahl F, Gosling T, Oszwald M, Hufner T, Krettek C (2009) Robot-assisted long bone fracture reduction. Int J Robot Res 28(10):1259–1278. doi:10.1177/0278364909101189
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflicts of interest
The authors declare that they have no conflict of interest.
Ethical standard
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.
Informed consent
The informed consent to participate in this study is attached as a supplementary material.
Rights and permissions
About this article
Cite this article
Ebrahimi, H., Yee, A. & Whyne, C. Surgical process analysis identifies lack of connectivity between sequential fluoroscopic 2D alignment as a critical impediment in femoral intramedullary nailing. Int J CARS 11, 297–305 (2016). https://doi.org/10.1007/s11548-015-1262-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11548-015-1262-6