Abstract
Purpose
To investigate if the routine use of an aortic balloon within 15–30 min after Ovation stent graft ring inflation would resolve any inflow stenosis, which may reach 60 %, at the level of the sealing rings. Moreover, we estimated the potential hemodynamic compromise in these patients during rest and exercise.
Methods
Following 3-dimensional reconstruction of AAA models, cross-sectional area of the infrarenal aorta just proximal the sealing mechanism (A aort, R aort, respectively) and internal area at the site of stenosis (A int, R int, respectively) were measured for 83. Forty-nine patients were managed without and 34 with an aortic balloon use. Pressure drop during rest and exercise was estimated.
Results
Technical success was 98 % and there were no perioperative deaths, one type-I endoleak, and 12 (14.5 %) type-II endoleaks. Median A int and R int were significantly reduced compared to A aort [55 % reduction, 143 (range 28–380) mm2 vs 314 (range 177–531) mm2, P value <0.001] and R aort [42 % reduction, 6.75 (range 3–11) mm vs 10 (range 7.5–13) mm, P value <0.001]. The observed stenosis was significantly less for patients in whom an aortic balloon was used intraoperatively (area reduction 36 vs 59 %, P value = 0.009). This stenosis caused a statistically significant, but clinically insignificant ΔP in both groups during rest (0.13 vs 0.06 mmHg, P value = 0.02) and exercise (1 vs 0.5 mmHg, P value = 0.02).
Conclusion
The advantages of the unique sealing mechanism of the Ovation device seem to be accompanied by an inflow stenosis which is significantly reduced when neck molding with an aortic balloon is used. Overall, the hemodynamic impact of this abnormality seems to be clinically insignificant at 1-month follow-up.
Similar content being viewed by others
References
Parodi JC, Palmaz JC, Barone HD (1991) Transfemoral intraluminal graft implantation for abdominal aortic aneurysms. Ann Vasc Surg 5:491–499
Sachs T, Schermerhorn M, Pomposelli F, Cotterill P, O’Malley J, Landon B (2011) Resident and fellow experiences after the introduction of endovascular aneurysm repair for abdominal aortic aneurysm. J Vasc Surg 54:881–888
Ioannou CV, Kontopodis N, Kehagias E, Papaioannou A, Kafetzakis A, Papadopoulos G, Pantidis D, Tsetis D (1051) Endovascular aneurysm repair with the Ovation TriVascular Stent Graft System utilizing a predominantly percutaneous approach under local anaesthesia. Br J Radiol 2015(88):20140735
The United Kingdom EVAR Trial Investigators, Greenhalgh RM, Brown LC, Powell JT, Thompson SG, Epstein D, Sculpher MJ (2010) Endovascular versus open repair of abdominal aortic aneurysm. N Engl J Med 362:1863–1871
Sweet MP, Fillinger MF, Morrison TM, Morrison TM, Abel D (2011) The influence of gender and aortic aneurysm size on eligibility for endovascular abdominal aortic aneurysm repair. J Vasc Surg 54:931–937
Moulakakis KG, Dalainas I, Kakisis J, Giannakopoulos TG, Liapis CD (2012) Current knowledge on EVAR with the ultra-low profile Ovation Abdominal Stent-graft System. J Cardiovasc Surg (Torino) 53:427–432
de Donato G, Setacci F, Sirignano P, Galzerano G, Borrelli MP, di Marzo L, Setacci C (2014) Ultra-low profile Ovation device: is it the definitive solution for EVAR? J Cardiovasc Surg (Torino) 55:33–40
Mangialardi N, Ronchey S, Kasemi H, Alberti V, Fazzini S, Serrao E (2013) Percutaneous endovascular aneurysm repair with the ultra-low profile Ovation abdominal stent-graft system. J Cardiovasc Surg (Torino) 54:581–587
Mehta M, Valdιs FE, Nolte T, Mishkel GJ, Jordan WD, Gray B, Eskandari MK, Botti C, A Pivotal Clinical Study to Evaluate the Safety and Effectiveness of the Ovation Abdominal Stent Graft System Investigators (2014) One-year outcomes from an international study of the Ovation Abdominal Stent Graft System for endovascular aneurysm repair. J Vasc Surg 59:65–73
Nano G, Mazzaccaro D, Stegher S, Occhiuto MT, Malacrida G, Tealdi DG, Alberti A, Volpe P (2014) Early experience with Ovation endograft system in abdominal aortic disease. J Cardiothorac Surg 9:48
Ioannou CV, Kontopodis N, Metaxa E, Papaharilaou Y, Georgakarakos E, Kafetzakis A, Kehagias E, Tsetis D (2014) Graft inflow stenosis induced by the inflatable ring fixation mechanism of the ovation stent-graft system: hemodynamic and clinical implications. J Endovasc Ther 21:829–838
www.trivascular.com/images/files/810-0012-01-01rD_IFU_PMA_Ovation_Prime.pdf
Moll FL, Powell JT, Fraedrich G, Verzini F, Haulon S, Waltham M, van Herwaarden JA, Holt PJ, van Keulen JW, Rantner B, Schlösser FJ, Setacci F, Ricco JB (2011) European Society for Vascular Surgery. Management of abdominal aortic aneurysms clinical practice guidelines of the European society for vascular surgery. Eur J Vasc Endovasc Surg 41(Suppl 1):S1–S58
Chaicof EL, Brewster DC, Dalman RL, Makaroun MS, Illig KA, Sicard GA, Timaran CH, Upchurch Jr GR, Veith FJ, Society for Vascular Surgery (2009) The care of patients with an abdominal aortic aneurysm: the Society for Vascular Surgery practice guidelines. J Vasc Surg 50(Suppl 4):S2–S49. doi:10.1016/j.jvs.2009.07.002
Kontopodis N, Tsetis D, Kehagias E, Daskalakis N, Galanakis N, Ioannou CV (2015) Totally percutaneous endovascular aneurysm repair using the preclosing technique: towards the least invasive therapeutic alternative. Surg Laparosc Endosc Percutan Tech 25:354–357
Yushkevich PA, Piven J, Hazlett HC, Smith RG, Ho S, Gee JC, Gerig G (2006) User-guided 3D active contour segmentation of anatomical structures: significantly improved efficiency and reliability. Neuroimage 31:1116–1128
Antiga L, Steinman DA (2004) Robust and objective decomposition and mapping of bifurcating vessels. IEEE Trans Med Imaging 23:704–713
Wentland AL, Grist TM, Wieben O (2013) Repeatability and internal consistency of abdominal 2D and 4D phase contrast MR flow measurements. Acad Radiol 20:699–704
Pedersen EM, Kozerke S, Ringgaard S, Scheidegger MB, Boesiger P (1999) Quantitative abdominal aortic flow measurements at controlled levels of ergometer exercise. Magn Reson Imaging 17:489–494
O’Carroll C, Sorbie KS (1993) Generalization of the Poiseuille law for one- and two-phase flow in a random capillary network. Phys Rev E Stat Phys Plasmas Fluids Relat Interdiscip Topics 47:3467–3476
Moore WS, Malone JM (1979) Effect of flow rate and vessel caliber on critical arterial stenosis. J Surg Res 26:1
Brun JF, Varlet-Marie E, Romain AJ, Guiraudou M, Raynaud de Mauverger E (2013) Exercise hemorheology: moving from old simplistic paradigms to a more complex picture. Clin Hemorheol Microcirc 55:15–27. doi:10.3233/CH-131686
de Donato G, Setacci F, Bresadola L, Castelli P, Chiesa R, Mangialardi N, Nano G, Setacci C, Trivascular ovation italian study (2016) Aortic neck evolution after endovascular repair with TriVascular Ovation stent graft. J Vasc Surg 63:8–15. doi:10.1016/j.jvs.2015.07.099
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
None to declare.
Ethical approval
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki declaration and its later amendments or comparable ethical standards.
Informed consent
Informed consent was obtained from all individual participants included in the study.
Funding
None.
Rights and permissions
About this article
Cite this article
Ioannou, C.V., Kontopodis, N., Georgakarakos, E. et al. Routine use of an aortic balloon to resolve possible inflow stenosis induced by the inflatable ring fixation mechanism of the Ovation endograft. Radiol med 121, 882–889 (2016). https://doi.org/10.1007/s11547-016-0671-0
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11547-016-0671-0