Abstract
Digital breast tomosynthesis (DBT) is a recently developed system for three-dimensional imaging that offers the potential to reduce the false positives of mammography by preventing tissue overlap. Many qualitative evaluations of digital breast tomosynthesis were previously performed by using a phantom with an unrealistic model and with heterogeneous background and noise, which is not representative of real breasts. The purpose of the present work was to compare reconstruction algorithms for DBT by using various breast phantoms; validation was also performed by using patient images. DBT was performed by using a prototype unit that was optimized for very low exposures and rapid readout. Three algorithms were compared: a back-projection (BP) algorithm, a filtered BP (FBP) algorithm, and an iterative expectation maximization (EM) algorithm. To compare the algorithms, three types of breast phantoms (homogeneous background phantom, heterogeneous background phantom, and anthropomorphic breast phantom) were evaluated, and clinical images were also reconstructed by using the different reconstruction algorithms. The in-plane image quality was evaluated based on the line profile and the contrast-to-noise ratio (CNR), and out-of-plane artifacts were evaluated by means of the artifact spread function (ASF). Parenchymal texture features of contrast and homogeneity were computed based on reconstructed images of an anthropomorphic breast phantom. The clinical images were studied to validate the effect of reconstruction algorithms. The results showed that the CNRs of masses reconstructed by using the EM algorithm were slightly higher than those obtained by using the BP algorithm, whereas the FBP algorithm yielded much lower CNR due to its high fluctuations of background noise. The FBP algorithm provides the best conspicuity for larger calcifications by enhancing their contrast and sharpness more than the other algorithms; however, in the case of small-size and low-contrast microcalcifications, the FBP reduced detectability due to its increased noise. The EM algorithm yielded high conspicuity for both microcalcifications and masses and yielded better ASFs in terms of the full width at half maximum. The higher contrast and lower homogeneity in terms of texture analysis were shown in FBP algorithm than in other algorithms. The patient images using the EM algorithm resulted in high visibility of low-contrast mass with clear border. In this study, we compared three reconstruction algorithms by using various kinds of breast phantoms and patient cases. Future work using these algorithms and considering the type of the breast and the acquisition techniques used (e.g., angular range, dose distribution) should include the use of actual patients or patient-like phantoms to increase the potential for practical applications.
Similar content being viewed by others
Explore related subjects
Discover the latest articles, news and stories from top researchers in related subjects.References
American Cancer Society (2014) Cancer facts and figures 2014. American Cancer Society, Atlanta
Sundresh NJ, Narendran S, Gokul C, Ashwini T (2014) An epidemological study and management of early breast cancer. World J Pharm Pharm Sci 3(8):2036–2053
Lee CH, Dershaw D, Kopans D et al (2010) Breast cancer screening with imaging: recommendations from the society of breast imaging and the ACR on the use of mammography, breast MRI, breast ultrasound, and other technologies for the detection of clinically occult breast cancer. J Am Coll Radiol 7:18–27. doi:10.1016/j.jacr.2009.09.022
Azar FS (2000) Imaging techniques for detecting breast cancer: survey and perspectives. Tech Rep (CIS) Pap 115. http://repository.upenn.edu/cis_reports/115
David G, Gordon SA, Denise MC et al (2009) Digital breast tomosynthesis: observer performance study. Am J Roentgenol 193:586–591. doi:10.2214/AJR.08.2031
Steven PP, Tor DT, Christine AK, Helene MN (2007) Digital breast tomosynthesis: initial experience in 98 women with abnormal digital screening mammography. Am J Roentgenol 189:616–623. doi:10.2214/AJR.07.2231
Park JM et al (2007) Breast tomosynthesis: present considerations and future applications. RadioGraphics 27:S231–S240. doi:10.1148/rg.27si075511
Dobbins JT III, Godfrey DJ (2003) Digital x-ray tomosynthesis: current state of the art and clinical potential. Phys Med Biol 48:R65–R106
Sechopoulos I (2013) A review of breast tomosynthesis. Part I. The image acquisition process. Med Phys 40:014301. doi:10.1118/1.4770279
Sechopoulos I (2013) A review of breast tomosynthesis. Part II. Image reconstruction, processing and analysis, and advanced applications. Med Phys 40(1):014302. doi:10.1118/1.4770281
Wu T, Stewart A, Stanton M et al (2003) Tomographic mammography using a limited number of low-dose cone-beam projection images. Med Phys 30:365–380. doi:10.1118/1.1543934
Zhang Y, Chan HP, Sahiner B (2006) A comparative study of limited-angle cone-beam reconstruction methods for breast tomosynthesis. Med Phys 33(10):3781–3795. doi:10.1118/1.223754
Wu T, Moore RH, Rafferty EA, Kopans DB (2004) A comparison of reconstruction algorithms for breast tomosynthesis. Med Phys 31:2636–2647. doi:10.1118/1.1786692
Reiser I, Bian J, Nishikawa RM, Sidky EY, Pan X (2009) Comparison of reconstruction algorithms for digital breast tomosynthesis. Physics.med-ph. arXiv:0908.2610v1
Dance DR, Young KC, Van Engen RE (2011) Estimation of mean glandular dose for breast tomosynthesis: factors for use with the UK, European and IAEA breast dosimetry protocols. Phys Med Biol 56:453–471. doi:10.1088/0031-9155/56/2/011
Turbell H (2001) Cone-Beam reconstruction using filtered backprojection. Dissertation, Linkopings universitet
Lauritsch G and Harer W (2011) A theoretical framework for filtered backprojection in tomosynthesis. SPIE 3338:1127–1137. http://spiedl.org/terms
Mertelmeier T, Orman J, Haerer W, Dudam MK (2006) Optimizing filtered backprojection reconstruction for a breast tomosynthesis prototype device. SPIE 6142:61420F. doi:10.1117/12.651380
McLachlan GJ, Krishnan T (2008) The EM Algorithm and Extensions. Wiley, New Jersey
Kontos D, Bakic PR et al (2009) Parenchymal texture analysis in digital breast tomosynthesis for breast cancer risk estimation: a preliminary study. Acad Radiol 16(3):283–298. doi:10.1016/j.acra.2008.08.014
Kontos D, Ikejimba LC et al (2011) Analysis of parenchymal texture with digital breast tomosynthesis: comparison with digital mammography and implications for cancer risk assessment. Radiology 261:80–91
Park HS, Kim YS, Kim HJ et al (2013) Optimization of configuration parameters in a newly developed digital breast tomosynthesis system. J Radiat Res. doi:10.1093/jrr/rrt130
Park HS, Kim YS et al (2014) Experimental investigation for determination of optimal X-ray beam tube voltages in a newly developed digital breast tomosynthesis system. Nucl Instrum Methods Phys Re A 763:163–169
Young S, Park S, Anderson SK, Badano A, Myers KJ, Bakic P (2009) Estimating breast tomosynthesis performance in detection tasks with variable-background phantoms. Proc SPIE 7258:72580O–72589O. doi:10.1117/12.813900
Byng JW, Mainprize JG, Yaffe MJ (1998) X-ray characterization of breast phantom materials. Phys Med Biol 43:1367–1377
Acknowledgments
We would like to acknowledge financial support from the R&D Convergence Program of the National Research Council of Science and Technology of the Republic of Korea (Grant CAP-13-3-KERI). This research was supported by Basic Science Research Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Science, ICT and Future Planning (NRF-2010-0018504).
Author information
Authors and Affiliations
Corresponding author
Ethics declarations
Conflict of interest
The authors declare that they have no conflict of interest.
Ethical standards
All procedures performed in studies involving human participants were in accordance with the ethical standards of the institutional and/or national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.
Informed consent
Informed consent was obtained from all individual participants included in the study.
Rights and permissions
About this article
Cite this article
Kim, Ys., Park, Hs., Lee, HH. et al. Comparison study of reconstruction algorithms for prototype digital breast tomosynthesis using various breast phantoms. Radiol med 121, 81–92 (2016). https://doi.org/10.1007/s11547-015-0583-4
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s11547-015-0583-4